D140530.2MAO/SMPS/S1 - clogged impactor at the SMPS aerosol sample inletmaoaossmpsS1.a1Suspect
D140815.1MAO/AOS/S1/M1 - Instrument maintenancemaoaossmpsS1.a1Incorrect
D141203.3MAO/ALL/S1 - Inlet down/off for cleaning and maintenance.maoaossmpsS1.a1Incorrect
D150311.17MAO/SMPS/S1 - Instrument maintenancemaoaossmpsS1.a1Incorrect
D150429.6MAO/AOS/S1 - Flow rate changemaoaossmpsS1.a1Does not affect quality
D151203.4MAO/SMPS/S1 - Reprocess: SMPS datamaoaossmpsS1.a1Does not affect quality


DQRID : D140530.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/07/2014143005/13/20142043Suspect
Subject:
MAO/SMPS/S1 - clogged impactor at the SMPS aerosol sample inlet
DataStreams:maoaossmpsS1.a1
Description:
The clogging of the impactor leads to a decrease in the aerosol sample flow rate of the 
SMPS, which is nominally 1 lpm.  This decrease in the SMPS sample flow rate results in an 
underestimation of the ambient size distribution and integrated number concentration.
Measurements:maoaossmpsS1.a1:
  • aerosol_flow
  • CPC_inlet_flow
  • Total concentration(total_concentration)
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • CPC_sample_flow


Back To Table of Contents

DQRID : D140815.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/09/2014160008/15/20141500Incorrect
Subject:
MAO/AOS/S1/M1 - Instrument maintenance
DataStreams:maoaossmpsS1.a1
Description:
Instrument maintenance and cleaning inlets affecting data.
Measurements:maoaossmpsS1.a1:
  • gas_viscosity
  • median
  • low_voltage
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • mean
  • sheath_flow
  • aerosol_flow
  • d50
  • CPC_inlet_flow
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • sample_pressure
  • geometric_mean_std
  • status_flag
  • mode
  • CPC_sample_flow


Back To Table of Contents

DQRID : D141203.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/10/2014132508/10/20142115Incorrect
Subject:
MAO/ALL/S1 - Inlet down/off for cleaning and maintenance.
DataStreams:maoaossmpsS1.a1
Description:
Inlet stack of MAOS A lowered. All instruments on internal cabin air. No atmospheric 
validity on HTDMA, PSAP, f(RH) dry neph, wet neph, aethalometer, PASS3, SP2, ACSM, CCN-200, 
Impactor, CPC, UCPC, SMPS.
Measurements:maoaossmpsS1.a1:
  • gas_viscosity
  • median
  • low_voltage
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • mean
  • sheath_flow
  • aerosol_flow
  • d50
  • CPC_inlet_flow
  • Total concentration(total_concentration)
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • geometric_mean_std
  • sample_pressure
  • status_flag
  • mode
  • CPC_sample_flow


Back To Table of Contents

DQRID : D150311.17
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/06/2015155803/06/20151612Incorrect
03/05/2015145803/05/20151522Incorrect
03/04/2015134503/04/20151348Incorrect
03/05/2015145803/06/20151848Incorrect
03/05/2015135103/05/20151414Incorrect
03/04/2015130203/04/20151317Incorrect
03/05/2015143503/05/20151458Incorrect
Subject:
MAO/SMPS/S1 - Instrument maintenance
DataStreams:maoaossmpsS1.a1
Description:
A number of instrument maintenance tasks were performed.  The flow rate of the system from 
the impactor was tested.  Inlet flow rates and nozzle pressures were examined on the CPC 
3772.  The accuracy of the classifier voltage and sheath flow rates were evaluated.  A 
HEPA filter zero test was performed.  During an approximately 24 hour period, the voltage 
cable was left unplugged from the classifier.
Measurements:maoaossmpsS1.a1:
  • median
  • low_voltage
  • geometric_mean
  • lower_size
  • high_voltage
  • mean
  • d50
  • Total concentration(total_concentration)
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • geometric_mean_std
  • mode


Back To Table of Contents

DQRID : D150429.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/17/2014130004/20/20151845Does not affect quality
Subject:
MAO/AOS/S1 - Flow rate change
DataStreams:maoaossmpsS1.a1
Description:
At begining of period, the bypass blower on MAOS A failed.  This was replaced by the 
instrument operator, but operator was unable to reprogram the PID flow controller.  The blower 
ran approximately 50% faster than designed flow rate.  It is unlikely this significantly 
affected the flow dynamics or transmission efficiencies in the MAOS A sampling stack.  
The data is colored transparent, but the DQR is filed to alert the user.  No specific 
degradation of signals were noted at this time.
Measurements:maoaossmpsS1.a1:
  • gas_viscosity
  • median
  • low_voltage
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • mean
  • sheath_flow
  • aerosol_flow
  • d50
  • CPC_inlet_flow
  • Total concentration(total_concentration)
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • geometric_mean_std
  • sample_pressure
  • status_flag
  • mode
  • CPC_sample_flow


Back To Table of Contents

DQRID : D151203.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/29/2014000012/28/20150000Does not affect quality
Subject:
MAO/SMPS/S1 - Reprocess: SMPS data
DataStreams:maoaossmpsS1.a1
Description:
No affect on data quality
Measurements:maoaossmpsS1.a1:
  • status_flag


Back To Table of Contents