Data Quality Reports for Session: 182550 User: zhengg1 Completed: 12/30/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160818.2ENA/OZONE/C1 - Failing pumpenaaosozoneC1.01Does not affect quality
D160907.11ENA/TRACEGAS/C1 - Flow meters not operatingenaaoscoC1.01Does not affect quality
D160909.17ENA/OZONE/C1 - Water condensing in the lineenaaosozoneC1.01Incorrect


DQRID : D160818.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/22/2016140009/20/20161100Does not affect quality
Subject:
ENA/OZONE/C1 - Failing pump
DataStreams:enaaosozoneC1.01
Description:
The pump was failing, and the operator attempted a repair.  Eventually, the flow bypassed 
the pump internally and was replaced with an external HTDMA vacuum pump.  Data appear 
valid since the flow only affects the internal calibration source and not the ambient 
measurement.

Internal pump was repaired on 2016-09-20 11:00.  The flow returned to the nominal value 
decreased slightly).  The response to calibration increased proportionally (as expected 
given the lower flow rate and longer residence time in the O3 generator cell).  The 
instrument response to ambient O3 should not be a function of the flow rate.  HOWEVER, a slight 
discontinuity was noted in the ambient O3 values prior to this DQR and following this DQR 
end date.  The amount of uncertainty was judged as <5 ppbv.
Measurements:enaaosozoneC1.01:
  • Ozone pressure(ozone_pressure)
  • ozone_bench_temp
  • Ozone concentration(ozone)


Back To Table of Contents

DQRID : D160907.11
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/25/2016114109/07/20161900Does not affect quality
Subject:
ENA/TRACEGAS/C1 - Flow meters not operating
DataStreams:enaaoscoC1.01
Description:
Flow meters were not operating so that there were no standard adds for calibration.
Measurements:enaaoscoC1.01:
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(co_dry)
  • N2O mole fraction of during batch measurement(n2o)
  • Calculated CO concentration(co)


Back To Table of Contents

DQRID : D160909.17
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/17/2016103508/18/20160538Incorrect
08/15/2016020208/15/20160939Incorrect
08/20/2016134808/21/20161726Incorrect
08/24/2016084008/24/20161105Incorrect
08/01/2016000008/04/20161832Incorrect
Subject:
ENA/OZONE/C1 - Water condensing in the line
DataStreams:enaaosozoneC1.01
Description:
Instrument suffered major oscillations in the signal.  Working with Operator it was 
discovered that the Split System Air Conditioner was blowing directly on the Ozone detector.  
Water was condensing in the line.  The air was redirected and the temperature setpoint 
raised.  This reduced but didn't entirely eliminate the oscillations.
Measurements:enaaosozoneC1.01:
  • Ozone pressure(ozone_pressure)
  • ozone_bench_temp
  • Ozone concentration(ozone)


Back To Table of Contents