Data Quality Reports for Session: 182253 User: sherman Completed: 12/09/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160314.2OLI/MASC/M1 - False triggering of camerasolimascM1.00Suspect
D160323.3OLI/MASC/M1 - Missing instrument data possibleolimascM1.00Suspect


DQRID : D160314.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/25/2015180006/16/20152200Suspect
11/13/2014200001/29/20152110Suspect
Subject:
OLI/MASC/M1 - False triggering of cameras
DataStreams:olimascM1.00
Description:
Radar RF pollution causing a high frequency false triggering (black images) of the MASC 
cameras, power outages at the site, and heat pad power problems as well as IR sensor 
sensitivity problems caused an abundance of MASC imagery without any information.
Measurements:olimascM1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D160323.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/25/2015000008/25/20162359Suspect
Subject:
OLI/MASC/M1 - Missing instrument data possible
DataStreams:olimascM1.00
Description:
The MASC instrument has been installed on a pole on the corner of a connex container roof. 
The location is preliminary for this season, it causes aerodynamic problems. During 
windy situations (almost all the time during snowfall up in Oliktok) there is probably a kind 
of swirl forming and the snowflakes are concentrated around the margins of the inlet 
ring. Triggering only occurs when snowflakes pass close to the center of the ring. The 
location atop the container is obstructed and results in a loss of data. Double fence wind 
shielding is proposed for the following seasons. A move of the MASC to an unobstructed 
location away from the containers and installation of wind shielding is planned once the 
weather conditions are more favorable the latest in fall 2016.
Measurements:olimascM1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents