Data Quality Reports for Session: 181935 User: devarakond Completed: 11/29/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160225.1AWR/GVRP/S1 - Biased brightness temperaturesawrgvrpS1.b1Incorrect
D160225.2AWR/GVRP/S1 - Reprocessing Notice (RID 1438)awrgvrpS1.b1Does not affect quality


DQRID : D160225.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/24/2015050001/17/20162200Incorrect
Subject:
AWR/GVRP/S1 - Biased brightness temperatures
DataStreams:awrgvrpS1.b1
Description:
The calibration of the instrument suddenly changed affecting the brightness temperatures
Measurements:awrgvrpS1.b1:
  • Microwave brightness temperature(brightnessTemperature)


Back To Table of Contents

DQRID : D160225.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/20/2015000001/17/20162200Does not affect quality
Subject:
AWR/GVRP/S1 - Reprocessing Notice (RID 1438)
DataStreams:awrgvrpS1.b1
Description:
DATA REPROCESSED.  Data was reprocessed on 20160324 to fix issues noted in the original 
DQR copied below.

--------------------------------------------------------------------------------

The altitude of the site was outside of the range of the pressure sensor quality control 
limits. The lowest data quality limit (valid_min) should be set to 750 mb.
Measurements:awrgvrpS1.b1:
  • qc_surfacePressure


Back To Table of Contents



END OF DATA