Data Quality Reports for Session: 181631 User: cwilliams Completed: 11/23/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D131120.4SGP/KAZR/C1 - Reprocessing Notice (RID 1117)sgpkazrspeccmaskgecopolC1.a0, sgpkazrspeccmaskgexpolC1.a0Does not affect quality


DQRID : D131120.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/18/2011000003/15/20142359Does not affect quality
Subject:
SGP/KAZR/C1 - Reprocessing Notice (RID 1117)
DataStreams:sgpkazrspeccmaskgecopolC1.a0, sgpkazrspeccmaskgexpolC1.a0
Description:
See below for details of reprocessing that was completed on 20150626.  Original 
DQR category: 4 (Red, Incorrect).

-----------------------------------------------------------

The wrong calibrations were used due to a problem with the software. Mentor discovered the 
reflectivity between long and short pulse did not agree with each other after analyzing 
data. Data will be reprocessed.
Measurements:sgpkazrspeccmaskgexpolC1.a0:
  • Reflectivity(reflectivity)

sgpkazrspeccmaskgecopolC1.a0:
  • Reflectivity(reflectivity)


Back To Table of Contents



END OF DATA