Data Quality Reports for Session: 181505 User: cwilliams Completed: 11/21/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D110809.1SGP/KAZR/C1 - Transmitter turned OFFsgpkazrspeccmaskgexpolC1.a0, sgpkazrspeccmaskgecopolC1.a0Missing
D131120.4SGP/KAZR/C1 - Reprocessing Notice (RID 1117)sgpkazrspeccmaskgecopolC1.a0, sgpkazrspeccmaskgexpolC1.a0Does not affect quality


DQRID : D110809.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/02/2011160008/08/20111400Missing
Subject:
SGP/KAZR/C1 - Transmitter turned OFF
DataStreams:sgpkazrspeccmaskgexpolC1.a0, sgpkazrspeccmaskgecopolC1.a0
Description:
The transmitter shutdown and the software that monitors this did not restart the 
transmitter.
Measurements:sgpkazrspeccmaskgexpolC1.a0:
  • locator_mask
  • Spectrum of raw counts from spectrometer(spectra)
  • East longitude(lon)
  • signal_to_noise_ratio
  • Mean Doppler Velocity(mean_doppler_velocity)
  • North latitude(lat)
  • Reflectivity(reflectivity)
  • time_offset
  • range
  • time
  • base_time
  • velocity_bins
  • Altitude above mean sea level(alt)
  • Spectral Width(spectral_width)

sgpkazrspeccmaskgecopolC1.a0:
  • Altitude above mean sea level(alt)
  • velocity_bins
  • locator_mask
  • Mean Doppler Velocity(mean_doppler_velocity)
  • North latitude(lat)
  • base_time
  • Spectral Width(spectral_width)
  • Spectrum of raw counts from spectrometer(spectra)
  • time
  • signal_to_noise_ratio
  • Reflectivity(reflectivity)
  • East longitude(lon)
  • time_offset
  • range


Back To Table of Contents

DQRID : D131120.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/18/2011000003/15/20142359Does not affect quality
Subject:
SGP/KAZR/C1 - Reprocessing Notice (RID 1117)
DataStreams:sgpkazrspeccmaskgecopolC1.a0, sgpkazrspeccmaskgexpolC1.a0
Description:
See below for details of reprocessing that was completed on 20150626.  Original 
DQR category: 4 (Red, Incorrect).

-----------------------------------------------------------

The wrong calibrations were used due to a problem with the software. Mentor discovered the 
reflectivity between long and short pulse did not agree with each other after analyzing 
data. Data will be reprocessed.
Measurements:sgpkazrspeccmaskgexpolC1.a0:
  • Reflectivity(reflectivity)

sgpkazrspeccmaskgecopolC1.a0:
  • Reflectivity(reflectivity)


Back To Table of Contents



END OF DATA