Data Quality Reports for Session: 181335 User: koontz Completed: 11/16/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D030515.3NSA/MWR/C1 - no air temperature signalnsamwrC1.00Missing


DQRID : D030515.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/13/2003190601/15/20050054Missing
Subject:
NSA/MWR/C1  - no air temperature signal
DataStreams:nsamwrC1.00
Description:
When the new blower was upgraded by Radiometrics and reinstalled on the MWR, the air 
temperature sensor failed to properly report. It was determined that the wires carrying the 
signal to the analog board did not conform to the standard expected by the upgraded blower. 
The problem was corrected by changing the wiring and modifying the MWR software to read 
the signal from the appropriate corresponding channel.
Measurements:nsamwrC1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents



END OF DATA