Data Quality Reports for Session: 180996 User: takahashih1 Completed: 11/02/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D070628.1FKB/MWRP/M1 - Reprocessed: Incorrect calibrationfkbmwrpM1.b1Does not affect quality
D071009.1FKB/MWRP/M1 - LN2 Calibration-fkbmwrpM1.b1Incorrect
D071016.1FKB/MWRP/M1 - Reprocessed: Updated 30-GHz calibration coefficientfkbmwrpM1.b1Does not affect quality
D071207.1FKB/MWRP/M1 - Heavy rain and fog at the sitefkbmwrpM1.b1Does not affect quality
D080116.1FKB/MWRP/M1 - Possible slight drift of V-band calibrationfkbmwrpM1.b1Suspect
D080116.2FKB/MWRP/M1 - Periodic decrease in the 30-GHz brightness temperaturefkbmwrpM1.b1Incorrect


DQRID : D070628.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/20/2007052006/21/20072000Does not affect quality
Subject:
FKB/MWRP/M1 - Reprocessed: Incorrect calibration
DataStreams:fkbmwrpM1.b1
Description:
The software was started with the wrong calibration data.

Data reprocessed 2071227.
Measurements:fkbmwrpM1.b1:
  • Liquid water path, from brightness temperatures using statistical retrieval(liquidWaterPath)
  • Liquid water content, cloud, vert. prof. sfc.-10km, from Tbright stat. retrvl.(liquidWaterContent)
  • Water vapor, tot. precipitable, from Tbright, 23.835 & 30.0 GHz, stat. retrvl.(totalPrecipitableWater2)
  • MET air temperature(temperature)
  • Mixing ratio, water vap., vert. prof. sfc-10km, frm retr. water vap. dens. prof(waterVaporMixingRatio)
  • Height, lifting condensation level(liftingCondensationLevel)
  • Humidity, relative, prof. from sfc.-10km, from water vap. dens. and temp. prof.(relativeHumidity)
  • Water vapor, total precipitable, derived from Tbright using stat. retrvl.(totalPrecipitableWater)
  • Pressure, lifting condensation level(liftingCondensationLevelPres)
  • Temperature, virtual, vert. prof. sfc-10km, fr. wtr. vap. dens. and temp. prof.(virtualTemperature)
  • Height, level of free convection(levelFreeConvection)
  • Microwave brightness temperature(brightnessTemperature)
  • liquidWaterPath2RmsError
  • Liquid water path, from Tbright at 23.835 and 30.0 GHz only, using stat. retrvl.(liquidWaterPath2)
  • Density, water vapor, prof. from sfc.-10 km, der. from Tbright stat. retrvl.(waterVaporDensity)
  • Pressure, level of free convection(levelFreeConvectionPres)
  • liquidWaterContentRmsError
  • liquidWaterPathRmsError


Back To Table of Contents

DQRID : D071009.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/17/2007110009/17/20071500Incorrect
Subject:
FKB/MWRP/M1 - LN2 Calibration-
DataStreams:fkbmwrpM1.b1
Description:
Between 1100 and 1500 there were two calibration attempts.
Measurements:fkbmwrpM1.b1:
  • Humidity, relative, at instrument height (1 or 2 m)(surfaceRelativeHumidity)
  • Liquid water path, from brightness temperatures using statistical retrieval(liquidWaterPath)
  • Temperature, dew point, vert. prof. sfc-10km, from water vapor density prof.(dewpointTemperature)
  • Water vapor, tot. precipitable, from Tbright, 23.835 & 30.0 GHz, stat. retrvl.(totalPrecipitableWater2)
  • Height, cloud base, height where IR temp. matches retrieved temp. profile(cloudBaseHeight)
  • Pressure, equilibrium level(equilibriumLevelPres)
  • Mixing ratio, water vap., vert. prof. sfc-10km, frm retr. water vap. dens. prof(waterVaporMixingRatio)
  • North latitude(lat)
  • Altitude above mean sea level(alt)
  • Pressure, associated geometric height, at altitude, NGM model output(height)
  • Humidity, relative, prof. from sfc.-10km, from water vap. dens. and temp. prof.(relativeHumidity)
  • Pressure, lifting condensation level(liftingCondensationLevelPres)
  • elevation
  • blackbodyTemperature
  • time_offset
  • wetWindowFlag
  • Microwave brightness temperature(brightnessTemperature)
  • Temperature, air, at instrument height (1 or 2 m)(surfaceTemperature)
  • Liquid water path, from Tbright at 23.835 and 30.0 GHz only, using stat. retrvl.(liquidWaterPath2)
  • dataQualityFlags
  • Density, water vapor, at instrument height (1 or 2 m)(surfaceWaterVaporDensity)
  • surface pressure(pressure)
  • base_time
  • Pressure, level of free convection(levelFreeConvectionPres)
  • liquidWaterContentRmsError
  • Liquid water content, cloud, vert. prof. sfc.-10km, from Tbright stat. retrvl.(liquidWaterContent)
  • Height, equilibrium level(equilibriumLevel)
  • totalPrecipitableWaterRmsError
  • MET air temperature(temperature)
  • frequency
  • East longitude(lon)
  • azimuth
  • totalPrecipitableWater2RmsError
  • Height, lifting condensation level(liftingCondensationLevel)
  • Water vapor, total precipitable, derived from Tbright using stat. retrvl.(totalPrecipitableWater)
  • Temperature, virtual, vert. prof. sfc-10km, fr. wtr. vap. dens. and temp. prof.(virtualTemperature)
  • time
  • Height, level of free convection(levelFreeConvection)
  • CAPE, Convective available potential energy(cape)
  • Pressure, barometric. at instrument height (1 or 2 m)(surfacePressure)
  • waterVaporDensityRmsError
  • temperatureRmsError
  • infraredTemperature
  • liquidWaterPath2RmsError
  • Density, water vapor, prof. from sfc.-10 km, der. from Tbright stat. retrvl.(waterVaporDensity)
  • liquidWaterPathRmsError


Back To Table of Contents

DQRID : D071016.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/17/2007150010/09/20071840Does not affect quality
Subject:
FKB/MWRP/M1 - Reprocessed: Updated 30-GHz calibration coefficient
DataStreams:fkbmwrpM1.b1
Description:
On 9/17 the instrument was calibrated with liquid nitrogen. For the 30-GHz channel the 
results of calibration were not optimal and were causing a negative bias in the retrieved 
LWP.

On 10/9 the 30-GHz calibration parameters were updated with values derived from tip 
curves.

Data between 9/17 and 10/9 were reprocessed to correct the 30-GHz calibration.  
Reprocessed data were archived 20080122.
Measurements:fkbmwrpM1.b1:
  • Liquid water path, from brightness temperatures using statistical retrieval(liquidWaterPath)
  • Microwave brightness temperature(brightnessTemperature)
  • Liquid water path, from Tbright at 23.835 and 30.0 GHz only, using stat. retrvl.(liquidWaterPath2)


Back To Table of Contents

DQRID : D071207.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/06/2007000012/15/20070000Does not affect quality
Subject:
FKB/MWRP/M1 - Heavy rain and fog at the site
DataStreams:fkbmwrpM1.b1
Description:
The months of November and December have been very rainy and wet at the AMF site. During 
rain events the radiometer does not work properly because of water accumulating on the 
window. In addition condensation and fog will affect the data causing very high brightness 
temperatures. Although the instrument is equipped with rain detection system to flag bad 
data, it is important that the user pays extra attention to the quality of the data during 
periods of continuous rain, drizzle or fog by applying the quality checks and screening 
for high brightness temperatures.
Measurements:fkbmwrpM1.b1:
  • Liquid water path, from brightness temperatures using statistical retrieval(liquidWaterPath)
  • Temperature, dew point, vert. prof. sfc-10km, from water vapor density prof.(dewpointTemperature)
  • Liquid water content, cloud, vert. prof. sfc.-10km, from Tbright stat. retrvl.(liquidWaterContent)
  • Water vapor, tot. precipitable, from Tbright, 23.835 & 30.0 GHz, stat. retrvl.(totalPrecipitableWater2)
  • MET air temperature(temperature)
  • Microwave brightness temperature(brightnessTemperature)
  • Mixing ratio, water vap., vert. prof. sfc-10km, frm retr. water vap. dens. prof(waterVaporMixingRatio)
  • Water vapor, total precipitable, derived from Tbright using stat. retrvl.(totalPrecipitableWater)
  • Humidity, relative, prof. from sfc.-10km, from water vap. dens. and temp. prof.(relativeHumidity)
  • Liquid water path, from Tbright at 23.835 and 30.0 GHz only, using stat. retrvl.(liquidWaterPath2)
  • Temperature, virtual, vert. prof. sfc-10km, fr. wtr. vap. dens. and temp. prof.(virtualTemperature)
  • Density, water vapor, prof. from sfc.-10 km, der. from Tbright stat. retrvl.(waterVaporDensity)


Back To Table of Contents

DQRID : D080116.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/01/2007000012/31/20072300Suspect
Subject:
FKB/MWRP/M1 - Possible slight drift of V-band calibration
DataStreams:fkbmwrpM1.b1
Description:
During the month of November and December the bias between model computations and 
measurements for the 51.25 and 52.28 GHz channels has slightly increased to about 3 K. This might 
be due to a slight drift in the calibration of these channels.

The calibration error associated with these data is increased, to 2-3 K, for these two 
months. See this plot for a comparison of measurements (solid lines) and model (red 
dots) for the 51.25 (black), 52.28 (blue) and 53.85-GHz (green) channels.
Measurements:fkbmwrpM1.b1:
  • Microwave brightness temperature(brightnessTemperature)


Back To Table of Contents

DQRID : D080116.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/19/2007000012/31/20070000Incorrect
Subject:
FKB/MWRP/M1 - Periodic decrease in the 30-GHz brightness temperature
DataStreams:fkbmwrpM1.b1
Description:
Starting around 12/19 until the end of December, the 30-GHz brightness temperature started 
to show a periodic oscillation between 9 and 11 AM GMT. The cause of this is probably 
thermal instability of the radiometer since it does correspond with some larger 
oscillations of the mixer temperature. During this time external temperatures displayed large 
variations going from -10 C at night to about +5 C during the day. The corresponding liquid 
water path retrievals are affected showing periodic negative values.

Liquid water path retrievals between 12/19 and 12/31 are the most affected variable since 
they are largely affected by the 30-GHz channel. A time series plot of the data will 
clearly show the periodic fluctuation of the LWP variable.
Measurements:fkbmwrpM1.b1:
  • Liquid water path, from brightness temperatures using statistical retrieval(liquidWaterPath)
  • Liquid water content, cloud, vert. prof. sfc.-10km, from Tbright stat. retrvl.(liquidWaterContent)
  • Microwave brightness temperature(brightnessTemperature)
  • Liquid water path, from Tbright at 23.835 and 30.0 GHz only, using stat. retrvl.(liquidWaterPath2)


Back To Table of Contents



END OF DATA