DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D130509.1 | NSA/IRT/C1 - Measurement below minimum threshold | nsairtC1.b1, nsairt200msC1.a1 | Suspect |
D140204.1 | OLI/IRT/M1 - IRT not insulated | oliirtM1.b1, oliirt200msM1.a1 | Suspect |
D140304.5 | NSA/IRT/C1 - Reprocessing Notice (RID 1097) | nsairtC1.b1, nsairt200msC1.a1 | Does not affect quality |
D141013.5 | OLI/IRT/M1 - Instrument Communication | oliirtM1.b1, oliirt200msM1.a1 | Missing |
D141016.5 | OLI/IRT/M1 - power failure | oliirtM1.b1, oliirt200msM1.a1 | Missing |
D141204.3 | NSA/IRT/C1 - Incorrect qc_flag values | nsairtC1.b1 | Does not affect quality |
D150123.6 | OLI/IRT/M1 - Missing data due to loss of power | oliirtM1.b1, oliirt200msM1.a1 | Missing |
D150327.6 | OLI/IRT/M1 - Environmental conditions resulted in missing data | oliirtM1.b1, oliirt200msM1.a1 | Missing |
D150417.7 | OLI/IRT/M1 - Environmental conditions resulted in missing data | oliirtM1.b1, oliirt200msM1.a1 | Missing |
D150424.4 | OLI/IRT/M1 - Environmental | oliirtM1.b1, oliirt200msM1.a1 | Missing |
D150731.1 | NSA/IRT/C1 - power outage | nsairtC1.b1, nsairt200msC1.a1 | Missing |
D160718.1 | NSA/IRT/MWR/C1 - Data Missing - Data System Upgrades | nsairtC1.b1, nsairt200msC1.a1 | Missing |
D160801.3 | OLI/IRT/M1 - Missing data | oliirtM1.b1, oliirt200msM1.a1 | Missing |
D161005.4 | OLI/IRT/M1 - data acquisition | oliirtM1.b1, oliirt200msM1.a1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/11/2008 | 1936 | 02/26/2014 | 2231 | Suspect |
Subject: | NSA/IRT/C1 - Measurement below minimum threshold |
DataStreams: | nsairtC1.b1, nsairt200msC1.a1 |
Description: | When the actual sky temperature is less than the minimum measureable of 213.2K (-60C), the IRT (serial) value is mostly -9999 (missing) with an occasionally valid but usually not accurate reading, resulting in the "noisy" data. This situation will occur only during periods of very clear skies with low precipitable water vapor (cold and dry). The problem was be solved by upgrading the IRT to a model with a lower measuring threshold of 173.2K (-100C). |
Measurements: | nsairt200msC1.a1:
nsairtC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/21/2013 | 0820 | 01/18/2014 | 0805 | Suspect |
Subject: | OLI/IRT/M1 - IRT not insulated |
DataStreams: | oliirtM1.b1, oliirt200msM1.a1 |
Description: | When the IRT was exposed to ambient air temperatures below ~-30C, the internal reference temperature measurement reached a minimum, constant value of -22.71 C. Without an AERI to compare with, the affect on sky temperature measurements may be unknown. The IRT was insulated on 1/14. |
Measurements: | oliirt200msM1.a1:
oliirtM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/26/2014 | 2232 | 03/03/2014 | 1920 | Does not affect quality |
Subject: | NSA/IRT/C1 - Reprocessing Notice (RID 1097) |
DataStreams: | nsairtC1.b1, nsairt200msC1.a1 |
Description: | See below for details of reprocessing that was completed on 20140822. Original DQR category: 4 (Red, Incorrect ). ----------------------------------------------------------- The upgraded IRT was installed but was not automatically configured correctly. Measurements are in degrees C not K. |
Measurements: | nsairt200msC1.a1:
nsairtC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/22/2014 | 1750 | 10/02/2014 | 1829 | Missing |
Subject: | OLI/IRT/M1 - Instrument Communication |
DataStreams: | oliirtM1.b1, oliirt200msM1.a1 |
Description: | The serial cable connector became disconnected, apparently causing the datalogger to fail. The datalogger was replaced on 7/23 and the IRT analog cable was reconnected but the serial cable was not connected to avoid the same event happening again. The DB9 connectors inside the IRT enclosure were cleaned of corrosion and the serial cable was eventually reconnected. |
Measurements: | oliirt200msM1.a1:
oliirtM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/04/2014 | 1800 | 10/14/2014 | 1343 | Missing |
Subject: | OLI/IRT/M1 - power failure |
DataStreams: | oliirtM1.b1, oliirt200msM1.a1 |
Description: | The site data system experienced a power failure. |
Measurements: | oliirt200msM1.a1:
oliirtM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/26/2014 | 2240 | 01/06/2015 | 2300 | Does not affect quality |
Subject: | NSA/IRT/C1 - Incorrect qc_flag values |
DataStreams: | nsairtC1.b1 |
Description: | When the IRTs were upgraded, the values of sky_ir_temp:min, sky_ir_temp:max, sky_ir_temp:resolution, and ref_ir_temp:min were not changed. The quality control fields for these limits are incorrect. |
Measurements: | nsairtC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/31/2014 | 0300 | 01/08/2015 | 2359 | Missing |
Subject: | OLI/IRT/M1 - Missing data due to loss of power |
DataStreams: | oliirtM1.b1, oliirt200msM1.a1 |
Description: | Data not available due to site losing power. |
Measurements: | oliirt200msM1.a1:
oliirtM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/01/2015 | 0400 | 03/18/2015 | 1700 | Missing |
Subject: | OLI/IRT/M1 - Environmental conditions resulted in missing data |
DataStreams: | oliirtM1.b1, oliirt200msM1.a1 |
Description: | Extreme weather prevented fueling of the emergency generator that was providing site power. The site data system was gracefully shut down. |
Measurements: | oliirt200msM1.a1:
oliirtM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/30/2015 | 2000 | 04/01/2015 | 1905 | Missing |
Subject: | OLI/IRT/M1 - Environmental conditions resulted in missing data |
DataStreams: | oliirtM1.b1, oliirt200msM1.a1 |
Description: | A transient power anomaly occurred and severe weather prevented site operators from getting to the site. |
Measurements: | oliirt200msM1.a1:
oliirtM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/19/2015 | 1250 | 04/21/2015 | 1807 | Missing |
Subject: | OLI/IRT/M1 - Environmental |
DataStreams: | oliirtM1.b1, oliirt200msM1.a1 |
Description: | The site was without power. |
Measurements: | oliirt200msM1.a1:
oliirtM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/18/2015 | 0200 | 07/22/2015 | 1906 | Missing |
Subject: | NSA/IRT/C1 - power outage |
DataStreams: | nsairtC1.b1, nsairt200msC1.a1 |
Description: | The IRT was not receiving power. |
Measurements: | nsairt200msC1.a1:
nsairtC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/12/2016 | 1600 | 07/13/2016 | 1952 | Missing |
Subject: | NSA/IRT/MWR/C1 - Data Missing - Data System Upgrades |
DataStreams: | nsairtC1.b1, nsairt200msC1.a1 |
Description: | Site data systems were being upgraded during this period. |
Measurements: | nsairt200msC1.a1:
nsairtC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/25/2016 | 1700 | 07/26/2016 | 1959 | Missing |
Subject: | OLI/IRT/M1 - Missing data |
DataStreams: | oliirtM1.b1, oliirt200msM1.a1 |
Description: | Missing data related to OLI system upgrades. |
Measurements: | oliirt200msM1.a1:
oliirtM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/16/2016 | 1750 | 10/05/2016 | 1359 | Missing |
Subject: | OLI/IRT/M1 - data acquisition |
DataStreams: | oliirtM1.b1, oliirt200msM1.a1 |
Description: | The irt_comm program stopped communicating with the IRT. The program was restarted. |
Measurements: | oliirt200msM1.a1:
oliirtM1.b1:
|