D140530.2MAO/SMPS/S1 - clogged impactor at the SMPS aerosol sample inletmaoaossmpsS1.a1Suspect
D140815.1MAO/AOS/S1/M1 - Instrument maintenancemaoaossmpsS1.a1, maoaossmpstrhS1.a1Incorrect
D141203.3MAO/ALL/S1 - Inlet down/off for cleaning and maintenance.maoaossmpstrhS1.a1, maoaossmpsS1.a1Incorrect
D151203.4MAO/SMPS/S1 - Reprocess: SMPS datamaoaossmpsS1.a1Does not affect quality


DQRID : D140530.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/07/2014143005/13/20142043Suspect
Subject:
MAO/SMPS/S1 - clogged impactor at the SMPS aerosol sample inlet
DataStreams:maoaossmpsS1.a1
Description:
The clogging of the impactor leads to a decrease in the aerosol sample flow rate of the 
SMPS, which is nominally 1 lpm.  This decrease in the SMPS sample flow rate results in an 
underestimation of the ambient size distribution and integrated number concentration.
Measurements:maoaossmpsS1.a1:
  • aerosol_flow
  • CPC_inlet_flow
  • Total concentration(total_concentration)
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • CPC_sample_flow


Back To Table of Contents

DQRID : D140815.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/09/2014160008/15/20141500Incorrect
Subject:
MAO/AOS/S1/M1 - Instrument maintenance
DataStreams:maoaossmpsS1.a1, maoaossmpstrhS1.a1
Description:
Instrument maintenance and cleaning inlets affecting data.
Measurements:maoaossmpstrhS1.a1:
  • Sample temperature(sample_temperature)
  • Sample relative humidity(sample_rh)

maoaossmpsS1.a1:
  • gas_viscosity
  • median
  • low_voltage
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • mean
  • sheath_flow
  • aerosol_flow
  • d50
  • CPC_inlet_flow
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • geometric_mean_std
  • sample_pressure
  • status_flag
  • mode
  • CPC_sample_flow


Back To Table of Contents

DQRID : D141203.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/10/2014132508/10/20142115Incorrect
Subject:
MAO/ALL/S1 - Inlet down/off for cleaning and maintenance.
DataStreams:maoaossmpstrhS1.a1, maoaossmpsS1.a1
Description:
Inlet stack of MAOS A lowered. All instruments on internal cabin air. No atmospheric 
validity on HTDMA, PSAP, f(RH) dry neph, wet neph, aethalometer, PASS3, SP2, ACSM, CCN-200, 
Impactor, CPC, UCPC, SMPS.
Measurements:maoaossmpstrhS1.a1:
  • Sample temperature(sample_temperature)
  • Sample relative humidity(sample_rh)

maoaossmpsS1.a1:
  • gas_viscosity
  • median
  • low_voltage
  • geometric_mean
  • lower_size
  • Sample temperature(sample_temperature)
  • high_voltage
  • mean
  • sheath_flow
  • aerosol_flow
  • d50
  • CPC_inlet_flow
  • Total concentration(total_concentration)
  • mean_free_path
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • diameter_midpoint
  • geometric_mean_std
  • sample_pressure
  • status_flag
  • mode
  • CPC_sample_flow


Back To Table of Contents

DQRID : D151203.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/29/2014000012/28/20150000Does not affect quality
Subject:
MAO/SMPS/S1 - Reprocess: SMPS data
DataStreams:maoaossmpsS1.a1
Description:
No affect on data quality
Measurements:maoaossmpsS1.a1:
  • status_flag


Back To Table of Contents