Data Quality Reports for Session: 180609 User: dzhang4 Completed: 10/19/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160608.2AWR/HSRL/M1 - Instrument malfunctionawrhsrlM1.01Incorrect


DQRID : D160608.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/04/2016092306/06/20161100Incorrect
Subject:
AWR/HSRL/M1 - Instrument malfunction
DataStreams:awrhsrlM1.01
Description:
All data collected during this period is bad.  Problems with the output shutter/DAQ/Seed 
voltage were reported by the AMF2 operator. He restarted the computer on June 4th. The 
system lock point was lost at that time. The lock point was manually reset from ~-9V to 2.8V 
on Monday June 6 around 11 UTC.
Measurements:awrhsrlM1.01:
  • Raw data stream - documentation not supported


Back To Table of Contents



END OF DATA