Data Quality Reports for Session: 180072 User: dfeldman Completed: 09/29/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D140823.1SGP/SASHE/C1 - Instrument off-line for repairsgpsashenirC1.b1Missing
D140823.2SGP/SASHE/C1 - Severe shadowband alignment issuesgpsashenirC1.b1Incorrect
D150121.1MAO/PGH/PVC/SGP/SASHE/C1 - Reprocess: Stray light contamination at extreme wavelengthssgpsashenirC1.b1Suspect


DQRID : D140823.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/28/2014215507/01/20141822Missing
Subject:
SGP/SASHE/C1 - Instrument off-line for repair
DataStreams:sgpsashenirC1.b1
Description:
Instrument was returned for repair of azimuth stage.
Measurements:sgpsashenirC1.b1:
  • time
  • chiller_dewpoint
  • Atmospheric pressure(atmos_pressure)
  • base_time
  • collector_x_tilt_std
  • chiller_temperature
  • time_offset
  • airmass
  • solar_zenith
  • cosine_correction_computed
  • Direct horizontal solar irradiance from InGaAs linear array spectrometer(direct_horizontal_spectrometer_nir)
  • East longitude(lon)
  • number_of_scans_spectrometer_nir
  • Direct normal solar irradiance, InGaAs linear array spectrometer(direct_normal_spectrometer_nir)
  • collector_y_tilt_std
  • collector_temperature
  • mio_rh
  • North latitude(lat)
  • mio_temperature_trh
  • collector_dewpoint
  • ad_temperature_spectrometer_nir
  • test_incident_angle
  • Total hemispheric irradiance component, InGaAs linear array spectrometer(total_hemisp_spectrometer_nir)
  • cosine_solar_zenith_angle
  • solar_azimuth
  • bench_temperature_spectrometer_nir
  • cosine_correction_lowsun
  • band_azimuth
  • wavelength_nir
  • diffuse_correction
  • chiller_rh
  • Diffuse hemispheric irradiance component, InGaAs linear array spectrometer(diffuse_hemisp_spectrometer_nir)
  • collector_y_tilt
  • cosine_correction_hisun
  • integration_time_spectrometer_nir
  • cosine_test_incident_angle
  • collector_x_tilt
  • mio_temperature_mems
  • collector_rh
  • responsivity_spectrometer_nir
  • Altitude above mean sea level(alt)


Back To Table of Contents

DQRID : D140823.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/03/2014120008/12/20141559Incorrect
Subject:
SGP/SASHE/C1 - Severe shadowband alignment issue
DataStreams:sgpsashenirC1.b1
Description:
A severe shadowband misalignment prevented identification of direct and diffuse irradiance 
components for the entire solar day.  Total hemispheric irradiance is nominally 
acceptable.
Measurements:sgpsashenirC1.b1:
  • Direct horizontal solar irradiance from InGaAs linear array spectrometer(direct_horizontal_spectrometer_nir)
  • Diffuse hemispheric irradiance component, InGaAs linear array spectrometer(diffuse_hemisp_spectrometer_nir)
  • Direct normal solar irradiance, InGaAs linear array spectrometer(direct_normal_spectrometer_nir)


Back To Table of Contents

DQRID : D150121.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/22/2011000008/19/20142359Suspect
Subject:
MAO/PGH/PVC/SGP/SASHE/C1 - Reprocess: Stray light contamination at extreme wavelengths
DataStreams:sgpsashenirC1.b1
Description:
The irradiance measurements at near the limits of the spectrometer wavelength ranges are 
subject to contamination from stray light within the spectrometer itself. This affects not 
only the basic measurement, but also can affect the calibration such that reported 
values may be either higher or lower than actual. Typically, this yields a high bias for 
measurements at the low wavelength limit and a low bias at the high wavelength limit. This 
causes opposite effects for AOD with a low bias at low wavelength and a high bias for high 
wavelengths.
Measurements:sgpsashenirC1.b1:
  • Direct horizontal solar irradiance from InGaAs linear array spectrometer(direct_horizontal_spectrometer_nir)
  • Total hemispheric irradiance component, InGaAs linear array spectrometer(total_hemisp_spectrometer_nir)
  • Diffuse hemispheric irradiance component, InGaAs linear array spectrometer(diffuse_hemisp_spectrometer_nir)
  • Direct normal solar irradiance, InGaAs linear array spectrometer(direct_normal_spectrometer_nir)


Back To Table of Contents