Data Quality Reports for Session: 179849 User: decarlop1 Completed: 09/21/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160203.1AWR/LDIS/MET/S1 - Precipitation Measurements SuspectawrmetS1.b1Suspect
D160224.3AWR/CPC/M1 - Failed pumpawraoscpcfM1.a1, awraoscpcfM1.b1Incorrect
D160308.3AWR/CPC/M1 - Instrument Startup, No Data CollectedawraoscpcfM1.a1Missing
D160502.1AWR/CPC/M1 - CPC FloodingawraoscpcfM1.b1, awraoscpcfM1.a1Suspect
D160722.1AWR/CPC/NEPHELOMETER/PSAP/M1 - Low impactor flowawraoscpcfM1.b1, awraoscpcfM1.a1Suspect


DQRID : D160203.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/04/2015000001/19/20162359Suspect
Subject:
AWR/LDIS/MET/S1 - Precipitation Measurements Suspect
DataStreams:awrmetS1.b1
Description:
Analysis of the PARS2 and the PWD precipitation data indicate that the results are 
suspect, possibly due to frost/blowing snow.  Both instruments reported precipitation for a 
majority of the days.

The TSI movies were reviewed and the days of noticeable precipitation are:

12/26/15
12/31/15 - Potentially just frost buildup
1/6/16
1/11/16
1/12/16
1/13/16
Measurements:awrmetS1.b1:
  • PWD 1 minute mean precipitation rate(pwd_precip_rate_mean_1min)
  • pwd_cumul_snow
  • pwd_pw_code_15min
  • pwd_cumul_rain
  • PWD 1 minute mean visibility(pwd_mean_vis_1min)
  • pwd_pw_code_inst
  • pwd_pw_code_1hr
  • pwd_err_code
  • PWD 10 minute mean visibility(pwd_mean_vis_10min)


Back To Table of Contents

DQRID : D160224.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/29/2016195904/07/20162319Incorrect
03/15/2016020603/18/20160024Incorrect
03/12/2016164203/15/20160020Incorrect
03/01/2016000403/01/20162329Incorrect
02/21/2016033802/22/20162249Incorrect
Subject:
AWR/CPC/M1 - Failed pump
DataStreams:awraoscpcfM1.a1, awraoscpcfM1.b1
Description:
The pumps that draw air into the instrument for particles to be counted failed.  The 
values reported by the instrument are not representative of the particle number concentration 
in the atmosphere
Measurements:awraoscpcfM1.a1:
  • AOS number concentration of CN (concentration)(concentration)

awraoscpcfM1.b1:
  • AOS number concentration of CN (concentration)(concentration)


Back To Table of Contents

DQRID : D160308.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/15/2015204911/16/20152059Missing
Subject:
AWR/CPC/M1 - Instrument Startup, No Data Collected
DataStreams:awraoscpcfM1.a1
Description:
At the instrument startup of the AMF2 deployment, the computers were turned on before the 
instruments were turned on.  The data collection starts automatically on instrument 
startup, but if the instruments are not on and connected, the data text files will contain 
missing data codes such as NaN or -9999.
Measurements:awraoscpcfM1.a1:
  • ambient_pressure
  • cabinet_temp
  • condensation_temp
  • AOS number concentration of CN (concentration)(concentration)
  • fraction_full
  • laser_current
  • orifice_pressure
  • liquid_level
  • optics_temp
  • nozzle_pressure
  • saturation_temp


Back To Table of Contents

DQRID : D160502.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/18/2016042304/20/20161218Suspect
Subject:
AWR/CPC/M1 - CPC Flooding
DataStreams:awraoscpcfM1.b1, awraoscpcfM1.a1
Description:
Following a routine flow rate check, the CPC showed signs of flooding with the orifice 
pressure dropping below 10 kPa.  The butanol fill bottle was removed from the system.  After 
two days of continuous running the system dried out and the orifice pressures were 
restored to nominal values.  The data from this period is suspect.
Measurements:awraoscpcfM1.a1:
  • AOS number concentration of CN (concentration)(concentration)

awraoscpcfM1.b1:
  • AOS number concentration of CN (concentration)(concentration)


Back To Table of Contents

DQRID : D160722.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/18/2016040007/21/20160500Suspect
Subject:
AWR/CPC/NEPHELOMETER/PSAP/M1 - Low impactor flow
DataStreams:awraoscpcfM1.b1, awraoscpcfM1.a1
Description:
CPC inlet not properly connected to the sample lines. Flow through the impactor below 
nominal. Instruments sampling cabin air at least part of the time.
Measurements:awraoscpcfM1.a1:
  • ambient_pressure
  • condensation_temp
  • liquid_level
  • dilution_valve_position
  • dilution_flow_setpoint
  • fraction_full
  • orifice_pressure
  • nozzle_pressure
  • dilution_flow_reading
  • cabinet_temp
  • AOS number concentration of CN (concentration)(concentration)
  • laser_current
  • optics_temp
  • saturation_temp

awraoscpcfM1.b1:
  • saturation_temp
  • fraction_full
  • dilution_slope
  • AOS number concentration of CN (concentration)(concentration)
  • dilution_flow_setpoint
  • optics_temp
  • cabinet_temp
  • nozzle_pressure
  • orifice_pressure
  • dilution_intercept
  • dilution_valve_position
  • dilution_flow_reading
  • laser_current
  • aerosol_flow_calibrated
  • ambient_pressure
  • dilution_correction_factor
  • condensation_temp
  • liquid_level


Back To Table of Contents