Data Quality Reports for Session: 179103 User: kwintenv1 Completed: 08/18/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D110418.6SGP/MWRHF/C1 - Reprocessed: Rapid drift in 90 GHz noise diodesgpmwrhfC1.b1Does not affect quality


DQRID : D110418.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/18/2011000009/20/20112359Does not affect quality
Subject:
SGP/MWRHF/C1 - Reprocessed: Rapid drift in 90 GHz noise diode
DataStreams:sgpmwrhfC1.b1
Description:
The noise diode of the 90 GHz channel was showing rapid drifts that affected the accuracy 
of the calibration. Although the algorithm is designed to follow slow drifts in the noise 
diode, in this case (because the variation was fast) the computed brightness temperature 
was probably about 4-5 K colder than it should be.

Ingest was modified and data were reprocessed and archived in September 2011
Measurements:sgpmwrhfC1.b1:
  • 90.0 GHz sky brightness temperature(tbsky90)


Back To Table of Contents