Data Quality Reports for Session: 178955 User: manasi1mah Completed: 08/11/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D101020.7SGP/AOS/C1 - Data processing problemsgpnoaaaosavgC1.b1Does not affect quality
D120608.1SGP/AOS/C1 - CPC leaksgpnoaaaosavgC1.b1Incorrect


DQRID : D101020.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/01/2009000010/01/20100000Does not affect quality
Subject:
SGP/AOS/C1 - Data processing problem
DataStreams:sgpnoaaaosavgC1.b1
Description:
The AOS data was reprocessed for the AMF, SGP and NSA sites. There was an error in the 
processing of the corrections for the PSAP absorption signal such that a low pass filter was 
implemented for the scattering data used in these corrections.The difference between the 
corrected and previous data is < 0.05 Mm-1 (below the instrument detection limit) and 
should not have a significant effect on the previous data. Data will be reprocessed at a 
later time when more significant changes to the data processing are implemented.
Measurements:sgpnoaaaosavgC1.b1:
  • Aerosol light absorption coefficient, green channel, 10 um particle diameter(Ba_G_Dry_10um_PSAP1W_1)
  • Absorption coefficient, green wavelength, low RH, 1 um size cut(Ba_G_Dry_1um_PSAP1W_1)


Back To Table of Contents

DQRID : D120608.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/01/2010154004/18/20112010Incorrect
Subject:
SGP/AOS/C1 - CPC leak
DataStreams:sgpnoaaaosavgC1.b1
Description:
There was a slow internal leak in the CPC instrument and the data slowly decreased over 
time. The CN concentration is a lower limit and is likely 5 to 20% lower than the actual 
value with higher errors toward the end of the DQR period. The problem was fixed by 
replacing the CPC saturator block.
Measurements:sgpnoaaaosavgC1.b1:
  • N_CPC_1_std
  • N_CPC_1_N
  • N_OPC_1_N


Back To Table of Contents



END OF DATA