Data Quality Reports for Session: 178945 User: oktemr1 Completed: 08/11/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D140611.16SGP/MPLCMASK/C1 - Reprocessing Notice (RID 1136)sgp30smplcmask1zwangC1.c1Does not affect quality
D140611.17NSA/MPLCMASK/C1 - Reprocessing Notice (RID 1137)nsa30smplcmask1zwangC1.c1Does not affect quality
D140611.20MAO/MPLCMASK/M1 - Reprocessing Notice (RID 1139)mao30smplcmask1zwangM1.c1Does not affect quality
D140616.1TWP/MPLCMASK/C1/C2/C3 - Reprocessing Notice (RID 1148)twp30smplcmask1zwangC1.c1, twp30smplcmask1zwangC3.c1Does not affect quality


DQRID : D140611.16
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/09/2006000006/06/20142359Does not affect quality
Subject:
SGP/MPLCMASK/C1 - Reprocessing Notice (RID 1136)
DataStreams:sgp30smplcmask1zwangC1.c1
Description:
See below for details of reprocessing that was completed on 20140807.  Original DQR 
category: 4 (Red, Incorrect).

-----------------------------------------------------------

Linear depolarization ratio calculation incorrect in the VAP processing code.
Measurements:sgp30smplcmask1zwangC1.c1:
  • linear_depol_ratio


Back To Table of Contents

DQRID : D140611.17
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/29/2006000006/06/20142359Does not affect quality
Subject:
NSA/MPLCMASK/C1 - Reprocessing Notice (RID 1137)
DataStreams:nsa30smplcmask1zwangC1.c1
Description:
See below for details of reprocessing that was completed on 20140807.  Original DQR 
category:
4 (Red, Incorrect).

-----------------------------------------------------------

Linear depolarization ratio calculation is incorrect in the VAP processing code.
Measurements:nsa30smplcmask1zwangC1.c1:
  • linear_depol_ratio


Back To Table of Contents

DQRID : D140611.20
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/19/2013000006/06/20142359Does not affect quality
Subject:
MAO/MPLCMASK/M1 - Reprocessing Notice (RID 1139)
DataStreams:mao30smplcmask1zwangM1.c1
Description:
See below for details of reprocessing that was completed on 20140807.  Original DQR 
category:
4 (Red, Incorrect).

-----------------------------------------------------------

Linear depolarization ratio calculation is incorrect in the VAP processing code.
Measurements:mao30smplcmask1zwangM1.c1:
  • linear_depol_ratio


Back To Table of Contents

DQRID : D140616.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/31/2006000006/06/20142359Does not affect quality
Subject:
TWP/MPLCMASK/C1/C2/C3 - Reprocessing Notice (RID 1148)
DataStreams:twp30smplcmask1zwangC1.c1, twp30smplcmask1zwangC3.c1
Description:
See below for details of reprocessing that was completed on 20140822.  Original DQR 
category: 4 (Red, Incorrect ).

-----------------------------------------------------------

The Linear depolarization ratio calculation in the VAP is incorrect.
Measurements:twp30smplcmask1zwangC1.c1:
  • linear_depol_ratio

twp30smplcmask1zwangC3.c1:
  • linear_depol_ratio


Back To Table of Contents



END OF DATA