Data Quality Reports for Session: 178919 User: ql_data Completed: 08/09/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160708.11SGP/MWR3C/E39 - Condensation on lenssgpmwr3cE39.b1Incorrect
D160708.4SGP/MWR3C/E37 - Uncertain IRT datasgpmwr3cE37.b1Suspect
D160708.5SGP/MWR3C/E37 - Instrument malfunctionsgpmwr3cE37.b1Incorrect


DQRID : D160708.11
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/07/2016100006/07/20161400Incorrect
06/19/2016070006/19/20161400Incorrect
06/04/2016030006/04/20161400Incorrect
06/16/2016100006/16/20161400Incorrect
05/31/2016050005/31/20161400Incorrect
04/14/2016100004/14/20161400Incorrect
05/10/2016060005/10/20161400Incorrect
06/08/2016070006/08/20161400Incorrect
06/06/2016100006/06/20161400Incorrect
04/21/2016100004/21/20161400Incorrect
06/15/2016070006/15/20161400Incorrect
04/15/2016100004/15/20161400Incorrect
06/21/2016090006/21/20161300Incorrect
06/13/2016070006/13/20161400Incorrect
06/03/2016030006/03/20161400Incorrect
05/29/2016050005/29/20161400Incorrect
07/02/2016110007/02/20161300Incorrect
07/04/2016100007/04/20161300Incorrect
06/14/2016070006/14/20161400Incorrect
06/12/2016050006/12/20161400Incorrect
05/27/2016100005/27/20161400Incorrect
05/28/2016080005/28/20161400Incorrect
06/20/2016070006/20/20161400Incorrect
06/11/2016060006/11/20161400Incorrect
05/26/2016100005/26/20161400Incorrect
Subject:
SGP/MWR3C/E39 - Condensation on lens
DataStreams:sgpmwr3cE39.b1
Description:
The dewblower was malfunctioning and there was condensation on the lens when the RH was 
higher that ~80%.
Measurements:sgpmwr3cE39.b1:
  • 89.0 GHz sky brightness temperature(tbsky89)
  • 30.0 GHz sky brightness temperature(tbsky30)
  • 23.8 GHz sky brightness temperature(tbsky23)
  • Column precip water vapor(pwv)
  • Liquid water path best-estimate value, hourly mean(lwp)


Back To Table of Contents

DQRID : D160708.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/28/2016193506/07/20161550Suspect
Subject:
SGP/MWR3C/E37 - Uncertain IRT data
DataStreams:sgpmwr3cE37.b1
Description:
The IRT data appear to disagree with the co-located MET IRT.
Measurements:sgpmwr3cE37.b1:
  • infrared_temperature


Back To Table of Contents

DQRID : D160708.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/30/2016160006/17/20161508Incorrect
Subject:
SGP/MWR3C/E37 - Instrument malfunction
DataStreams:sgpmwr3cE37.b1
Description:
The instrument malfunctioned and was pulled out and sent to the vendor
Measurements:sgpmwr3cE37.b1:
  • Column precip water vapor(pwv)
  • 23.8 GHz sky brightness temperature(tbsky23)
  • Liquid water path best-estimate value, hourly mean(lwp)
  • 30.0 GHz sky brightness temperature(tbsky30)


Back To Table of Contents



END OF DATA