Data Quality Reports for Session: 178832 User: sonia Completed: 08/05/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D140618.2ENA/CPC/NEPH/PSAP/C1 - Impactor not installedenaaosuhsasC1.a1Does not affect quality
D140731.2ENA/UHSAS/C1 - Reprocess: Upper and lower size limit incorrectenaaosuhsasC1.a1Incorrect
D140926.3ENA/UHSAS/C1 - Computer FailureenaaosuhsasC1.a1Missing
D150130.3ENA/DL/C1 - Instrument down due to power supply failureenadlfptC1.b1Missing
D150506.2ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site VisitenaaoshtdmaC1.a1, enaaosuhsasC1.a1Incorrect
D150526.6ENA/DL/C1 - System down due to power supply issueenadlfptC1.b1Missing
D150812.1ENA/HTDMA/C1 - Incorrect air flowenaaoshtdmaC1.a1Incorrect
D150909.1ENA/UHSAS/C1 - Reprocess: Non-optimal bin sizesenaaosuhsasC1.a1Does not affect quality
D151120.6ENA/HTDMA/C1 - Incorrect RH and air flowenaaoshtdmaC1.a1Incorrect
D151214.1ENA/UHSAS/C1 - High counts on three lowest channelsenaaosuhsasC1.a1Suspect
D151228.1ENA/UHSAS/C1 - spikes in lower size binsenaaosuhsasC1.a1Suspect
D160107.1ENA/UHSAS/C1 - HEPA filter attached to inletenaaosuhsasC1.a1Incorrect
D160122.2ENA/DL/C1 - Instrument removed for hurricaneenadlfptC1.b1Missing
D160126.5ENA/UHSAS/C1 - Noisy counts / no countsenaaosuhsasC1.a1Incorrect
D160209.1ENA/UHSAS/C1 - High counts in lower binsenaaosuhsasC1.a1Suspect
D160211.2ENA/UHSAS/C1 - High counts in lower binsenaaosuhsasC1.a1Incorrect
D160223.1ENA/UHSAS/C1 - Intermittent spikes in the lowest channelsenaaosuhsasC1.a1Suspect
D160314.1ENA/UHSAS/C1 - High counts in the 3 lowest size binsenaaosuhsasC1.a1Suspect
D160319.1ENA/UHSAS/C1 - High counts in the 3 lowest size binsenaaosuhsasC1.a1Suspect
D160321.1ENA/UHSAS/C1 - High counts in 3 lowest size binsenaaosuhsasC1.a1Suspect
D160411.1ENA/UHSAS/C1 - Intermittent spikes in the lower size binsenaaosuhsasC1.a1Suspect
D160411.2ENA/UHSAS/C1 - Instrument not sampling ambient airenaaosuhsasC1.a1Incorrect
D160427.7ENA/C1 - Reprocess: Slightly incorrect location fieldsenaaoshtdmaC1.a1Suspect
D160520.10ENA/DL/DLPROF/C1 - Reprocess: Incorrect serial numberenadlfptC1.b1Does not affect quality
D160610.10ENA/UHSAS/C1 - High counts in lower size binsenaaosuhsasC1.a1Suspect
D160610.9ENA/UHSAS/C1 - missing dataenaaosuhsasC1.a1Missing


DQRID : D140618.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/01/2013000002/19/20141351Does not affect quality
Subject:
ENA/CPC/NEPH/PSAP/C1 - Impactor not installed
DataStreams:enaaosuhsasC1.a1
Description:
No impactor installed.  All data is the same as when impactor at 10 um setting.
Measurements:enaaosuhsasC1.a1:
  • scattering_voltage
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)
  • sheath_flow
  • MET air temperature(temperature)
  • sampling_volume
  • purge_flow
  • unit_temperature
  • auxiliary
  • reference
  • current
  • sample_pressure
  • flow
  • lower_size_limit
  • upper_size_limit
  • differential_voltage


Back To Table of Contents

DQRID : D140731.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/17/2014145507/16/20141459Incorrect
Subject:
ENA/UHSAS/C1 - Reprocess: Upper and lower size limit incorrect
DataStreams:enaaosuhsasC1.a1
Description:
The lower size limit and upper size limit (99 bins) is incorrect because the values are 
not read from the header records.
Measurements:enaaosuhsasC1.a1:
  • upper_size_limit
  • lower_size_limit


Back To Table of Contents

DQRID : D140926.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/17/2014150009/23/20142007Missing
Subject:
ENA/UHSAS/C1 - Computer Failure
DataStreams:enaaosuhsasC1.a1
Description:
The CPC unit for the UHSAS (and other systems) failed.  The UHSAS was down at this time 
and came back up after the CPC unit was fixed.
Measurements:enaaosuhsasC1.a1:
  • scattering_voltage
  • time
  • base_time
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)
  • sheath_flow
  • Altitude above mean sea level(alt)
  • MET air temperature(temperature)
  • North latitude(lat)
  • sampling_volume
  • time_offset
  • purge_flow
  • unit_temperature
  • auxiliary
  • current
  • reference
  • sample_pressure
  • lower_size_limit
  • flow
  • upper_size_limit
  • differential_voltage
  • East longitude(lon)
  • bin_num


Back To Table of Contents

DQRID : D150130.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/30/2014040001/26/20151352Missing
Subject:
ENA/DL/C1 - Instrument down due to power supply failure
DataStreams:enadlfptC1.b1
Description:
No data was collected during the period in question because the instrument was down due to 
a power supply failure.
Measurements:enadlfptC1.b1:
  • North latitude(lat)
  • intensity
  • base_time
  • Altitude above mean sea level(alt)
  • azimuth
  • time
  • Attenuated backscatter(attenuated_backscatter)
  • Radial velocity(radial_velocity)
  • range
  • East longitude(lon)
  • elevation
  • time_offset


Back To Table of Contents

DQRID : D150506.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/22/2015170004/27/20151700Incorrect
Subject:
ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site Visit
DataStreams:enaaoshtdmaC1.a1, enaaosuhsasC1.a1
Description:
There was visit of 7 AOS instrument mentors to the ENA AOS during this period of 
instrument repair, calibration and modification. Instruments were occasionally off, or with no 
sample flow, or receiving calibration samples, etc. There was no productive sampling of 
ambient air during this period other than for testing. No data should be used for this period 
as ambient samples.

Note: Two datastream are missing. The CAPS Extinction Monitor and the Greenhouse gas 
instrument.
Measurements:enaaosuhsasC1.a1:
  • scattering_voltage
  • MET air temperature(temperature)
  • sampling_volume
  • auxiliary
  • reference
  • upper_size_limit
  • flow
  • differential_voltage
  • bin_num
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)
  • sheath_flow
  • purge_flow
  • unit_temperature
  • current
  • sample_pressure
  • lower_size_limit

enaaoshtdmaC1.a1:
  • scan_min_diameter_setting
  • scan_time_setting
  • dry_rh_setting
  • Concentration of aerosol(aerosol_concentration)
  • dry_diameter_setting
  • bin_center
  • scan_type
  • bin
  • scan_max_diameter_setting
  • bin_width
  • atomizer
  • Total concentration(total_concentration)
  • num_bins
  • humid_rh
  • sample_flow


Back To Table of Contents

DQRID : D150526.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/20/2015220005/22/20151345Missing
Subject:
ENA/DL/C1 - System down due to power supply issue
DataStreams:enadlfptC1.b1
Description:
The system was down during the period in question due to an issue with the power supply.
Measurements:enadlfptC1.b1:
  • North latitude(lat)
  • intensity
  • base_time
  • Altitude above mean sea level(alt)
  • azimuth
  • time
  • Attenuated backscatter(attenuated_backscatter)
  • Radial velocity(radial_velocity)
  • range
  • East longitude(lon)
  • elevation
  • time_offset


Back To Table of Contents

DQRID : D150812.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/28/2015000008/12/20151450Incorrect
Subject:
ENA/HTDMA/C1 - Incorrect air flow
DataStreams:enaaoshtdmaC1.a1
Description:
Low sample flow, no ambient aerosol size selection and HSEMS was measuring cabin air. 
Additionally, HSEMS air flows, temperatures and RH were incorrect.
Measurements:enaaoshtdmaC1.a1:
  • scan_min_diameter_setting
  • scan_time_setting
  • scan_max_diameter_setting
  • dry_rh_setting
  • Concentration of aerosol(aerosol_concentration)
  • dry_diameter_setting
  • bin_width
  • bin_center
  • Total concentration(total_concentration)
  • atomizer
  • scan_type
  • humid_rh
  • num_bins
  • sample_flow
  • bin


Back To Table of Contents

DQRID : D150909.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/20/2014160009/30/20152359Does not affect quality
Subject:
ENA/UHSAS/C1 - Reprocess: Non-optimal bin sizes
DataStreams:enaaosuhsasC1.a1
Description:
The UHSAS was programmed with a bin sizes appropriate to the aerial UHSAS. This is not 
really a problem. The bin sizes are not consistent with the other console UHSAS. The problem 
is in blindly comparing the data between the various UHSAS. This is not a problem of 
mislabeling or incorrect size bins. The data are correct for the stated size bins.
Measurements:enaaosuhsasC1.a1:
  • upper_size_limit
  • lower_size_limit


Back To Table of Contents

DQRID : D151120.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/27/2015000011/19/20150000Incorrect
Subject:
ENA/HTDMA/C1 - Incorrect RH and air flow
DataStreams:enaaoshtdmaC1.a1
Description:
The HT-DMA internal RH sensor was sending out data in a format that was not correctly 
interpreted by the instrument software. This led the software to assume an incorrect RH 
reading and set the humid and dry air flows at incorrect values, affecting RH and other 
internal flows.
Measurements:enaaoshtdmaC1.a1:
  • scan_min_diameter_setting
  • scan_time_setting
  • scan_max_diameter_setting
  • dry_rh_setting
  • Concentration of aerosol(aerosol_concentration)
  • dry_diameter_setting
  • bin_width
  • bin_center
  • Total concentration(total_concentration)
  • atomizer
  • scan_type
  • humid_rh
  • num_bins
  • sample_flow
  • bin


Back To Table of Contents

DQRID : D151214.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/12/2015201712/13/20152100Suspect
12/08/2015173912/08/20152346Suspect
12/13/2015204112/14/20150903Suspect
12/11/2015100112/12/20150058Suspect
12/10/2015111712/11/20150636Suspect
Subject:
ENA/UHSAS/C1 - High counts on three lowest channels
DataStreams:enaaosuhsasC1.a1
Description:
High counts in the UHSAS three lowest channels. Probably a contamination (a fiber on the 
jet nozzle) inside the optical chamber. Rest of the channels should not be affected.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D151228.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/25/2015120012/28/20151540Suspect
Subject:
ENA/UHSAS/C1 - spikes in lower size bins
DataStreams:enaaosuhsasC1.a1
Description:
Intermittent spikes in the lowest 3 size bins. Probably a contamination that was cleared 
when a HEPA filter was installed for zero testing. Other bins should not be affected.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D160107.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/30/2015190001/04/20161600Incorrect
Subject:
ENA/UHSAS/C1 - HEPA filter attached to inlet
DataStreams:enaaosuhsasC1.a1
Description:
HEPA filter was attached to the UHSAS inlet for troubleshooting the issues related to DQR 
D151228.1. No ambient measurements were made during this period.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D160122.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/14/2016000001/18/20161628Missing
Subject:
ENA/DL/C1 - Instrument removed for hurricane
DataStreams:enadlfptC1.b1
Description:
The instrument was taken offline and stored indoors to avoid potential damage during a 
hurricane passage.
Measurements:enadlfptC1.b1:
  • North latitude(lat)
  • intensity
  • base_time
  • Altitude above mean sea level(alt)
  • azimuth
  • time
  • Attenuated backscatter(attenuated_backscatter)
  • Radial velocity(radial_velocity)
  • range
  • East longitude(lon)
  • elevation
  • time_offset


Back To Table of Contents

DQRID : D160126.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/25/2016130001/26/20161745Incorrect
Subject:
ENA/UHSAS/C1 - Noisy counts / no counts
DataStreams:enaaosuhsasC1.a1
Description:
UHSAS was showing spikes in the lowest channels. An HEPA filter was attached to the inlet 
for troubleshooting. No valid data during the period with spikes and when the filter was 
attached.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D160209.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/05/2016045302/05/20161445Suspect
Subject:
ENA/UHSAS/C1 - High counts in lower bins
DataStreams:enaaosuhsasC1.a1
Description:
High counts in the 3 lowest bins/channels. Possible contamination. Other bins should not 
be affected.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D160211.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/10/2016134502/11/20161245Incorrect
Subject:
ENA/UHSAS/C1 - High counts in lower bins
DataStreams:enaaosuhsasC1.a1
Description:
High counts in the 4 lowest bins/channels. Possible contamination. Instrument set up to 
sample through a filter for troubleshooting. No valid data from the period.
Measurements:enaaosuhsasC1.a1:
  • scattering_voltage
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)
  • sheath_flow
  • MET air temperature(temperature)
  • sampling_volume
  • purge_flow
  • unit_temperature
  • auxiliary
  • reference
  • current
  • sample_pressure
  • flow
  • upper_size_limit
  • lower_size_limit
  • differential_voltage
  • bin_num


Back To Table of Contents

DQRID : D160223.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/21/2016213402/21/20162238Suspect
02/22/2016012202/22/20160300Suspect
02/21/2016055502/21/20160636Suspect
02/19/2016232702/20/20160249Suspect
Subject:
ENA/UHSAS/C1 - Intermittent spikes in the lowest channels
DataStreams:enaaosuhsasC1.a1
Description:
Intermittent spikes in the 3 lowest size bins. Rest of the bins should be unaffected.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D160314.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/11/2016124603/11/20161542Suspect
03/13/2016142503/13/20161708Suspect
03/13/2016120603/13/20161240Suspect
03/12/2016031003/12/20160737Suspect
Subject:
ENA/UHSAS/C1 - High counts in the 3 lowest size bins
DataStreams:enaaosuhsasC1.a1
Description:
High counts in the lowest 3 size bins. The rest of the bins should be unaffected.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D160319.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/18/2016052503/18/20160735Suspect
Subject:
ENA/UHSAS/C1 - High counts in the 3 lowest size bins
DataStreams:enaaosuhsasC1.a1
Description:
High counts in the 3 lowest size bins. Other size bins should not be affected.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D160321.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/20/2016045003/20/20160822Suspect
Subject:
ENA/UHSAS/C1 - High counts in 3 lowest size bins
DataStreams:enaaosuhsasC1.a1
Description:
High counts in the 3 lowest size bins. Other size bins should not be affected.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D160411.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/28/2016220004/01/20162330Suspect
Subject:
ENA/UHSAS/C1 - Intermittent spikes in the lower size bins
DataStreams:enaaosuhsasC1.a1
Description:
Intermittent spikes in the 3 lowest size bins. Other size bins should not be affected.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D160411.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/31/2016160004/01/20161025Incorrect
03/30/2016090003/30/20161100Incorrect
03/30/2016153003/30/20161900Incorrect
Subject:
ENA/UHSAS/C1 - Instrument not sampling ambient air
DataStreams:enaaosuhsasC1.a1
Description:
Instrument was periodically offline or not sampling ambient air for troubleshooting 
purposes. Data missing or no valid data.
Measurements:enaaosuhsasC1.a1:
  • scattering_voltage
  • base_time
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)
  • sheath_flow
  • MET air temperature(temperature)
  • sampling_volume
  • purge_flow
  • unit_temperature
  • auxiliary
  • reference
  • current
  • sample_pressure
  • flow
  • upper_size_limit
  • lower_size_limit
  • differential_voltage
  • bin_num


Back To Table of Contents

DQRID : D160427.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/29/2013094104/30/20162359Suspect
Subject:
ENA/C1 - Reprocess: Slightly incorrect location fields
DataStreams:enaaoshtdmaC1.a1
Description:
Some of the instruments are being processed with the previous "GRW-Azores" lat/lon/alt 
instead of the new "ENA-Azores" lat/lon/alt. The differences are minor, but will be 
corrected. The respective GRW and ENA lat, lon, and alt values are as follows:
 
GRW  lat = 39.0911 ;
ENA lat = 39.0916 ;
GRW  lon = -28.0297 ;
ENA  lon = -28.0257 ;
GRW  alt = 26 ;
ENA  alt = 30.48 ;
Measurements:enaaoshtdmaC1.a1:
  • East longitude(lon)
  • North latitude(lat)
  • Altitude above mean sea level(alt)


Back To Table of Contents

DQRID : D160520.10
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/21/2014000005/19/20162200Does not affect quality
Subject:
ENA/DL/DLPROF/C1 - Reprocess: Incorrect serial number
DataStreams:enadlfptC1.b1
Description:
The instrument serial number that appears in the netcdf files global attributes is 
incorrect. The correct serial number is 0514-84.
Measurements:enadlfptC1.b1:
  • base_time


Back To Table of Contents

DQRID : D160610.10
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/07/2016134006/08/20160200Suspect
Subject:
ENA/UHSAS/C1 - High counts in lower size bins
DataStreams:enaaosuhsasC1.a1
Description:
High counts in the 3 lowest size bins. Other size bins should not be affected.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D160610.9
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/06/2016221006/07/20161340Missing
Subject:
ENA/UHSAS/C1 - missing data
DataStreams:enaaosuhsasC1.a1
Description:
Instrument's internal hard drive was full and no new data was being saved except a few 
samples occasionally.
Measurements:enaaosuhsasC1.a1:
  • scattering_voltage
  • time
  • base_time
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)
  • sheath_flow
  • Altitude above mean sea level(alt)
  • MET air temperature(temperature)
  • North latitude(lat)
  • sampling_volume
  • time_offset
  • purge_flow
  • unit_temperature
  • auxiliary
  • current
  • reference
  • sample_pressure
  • lower_size_limit
  • flow
  • upper_size_limit
  • differential_voltage
  • East longitude(lon)
  • bin_num


Back To Table of Contents



END OF DATA