Data Quality Reports for Session: 178782 User: zhangc1 Completed: 08/03/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160318.12SGP/CCN/C1 - % super saturation values change after instrument calibrationsgpaosccn1colC1.b1Does not affect quality
D160406.2SGP/AOS/C1 - Wildfire smoke plumes advecting over facilitysgpaosccn1colC1.b1Does not affect quality
D160408.3SGP/CCN/C1 - Problem with flow control regulationsgpaosccn1colC1.b1Incorrect


DQRID : D160318.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/17/2016160003/18/20161800Does not affect quality
Subject:
SGP/CCN/C1 - % super saturation values change after instrument calibration
DataStreams:sgpaosccn1colC1.b1
Description:
% super saturation scan values during this time period were 0.12, 0.23, 0.4, 0.85, 0.51, 
0.29 and 0.18. Before and after this time period the % super saturation scan values are 
0.15, 0.25, 0.4, 0.8, 0.5, 0.3 and 0.2.
Measurements:sgpaosccn1colC1.b1:
  • supersaturation_set_point


Back To Table of Contents

DQRID : D160406.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/18/2016040003/18/20161100Does not affect quality
03/25/2016010003/25/20160700Does not affect quality
Subject:
SGP/AOS/C1 - Wildfire smoke plumes advecting over facility
DataStreams:sgpaosccn1colC1.b1
Description:
Smoke from nearby wildfires advected through SGP C1 during the listed affected time 
ranges, resulting in high signals observable in many of the listed AOS datastreams.
Measurements:sgpaosccn1colC1.b1:
  • AOS number concentration of CCN(N_CCN)
  • AOS CCN droplet count by bin size(N_CCN_dN)


Back To Table of Contents

DQRID : D160408.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/04/2016003005/10/20161600Incorrect
Subject:
SGP/CCN/C1 - Problem with flow control regulation
DataStreams:sgpaosccn1colC1.b1
Description:
A problem with flow control regulation. The instrument was returned to manufacture DMT for 
repair. The column was found cracked and replaced. Data scans from 5/6/2016 15:36 to 
5/10/2016 16:00 use the default scan mode that steps from 0.2 to 1.0 %ss. After this time, 
the pyramid stepping scheme is implemented.
Measurements:sgpaosccn1colC1.b1:
  • T_inlet
  • reported_temperature_gradient
  • dT_OPC
  • overflow
  • Q_sample
  • supersaturation_set_point
  • eta_target
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • T_read_gradient
  • T_OPC
  • T_nafion
  • size_bin_bounds
  • temperature_std
  • proportional_valve_voltage
  • dT_target_estimated
  • T_read_TEC1
  • T_read_TEC2
  • T_read_TEC3
  • seconds_after_transition
  • eta
  • T_set_gradient
  • supersaturation_calculated
  • size_bin
  • T_set_TEC3
  • T_set_TEC2
  • first_stage_monitor_voltage
  • T_set_TEC1
  • AOS number concentration of CCN(N_CCN)
  • T_sample
  • eta_lookup_table
  • supersaturation_calculated_target
  • Q_sheath
  • laser_current
  • dilution_correction_factor
  • P_sample
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • temp_unstable


Back To Table of Contents



END OF DATA