Data Quality Reports for Session: 178747 User: songj1 Completed: 08/02/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D140707.2SGP/VDIS/C1 - Blown light bulb resulted in incorrectsgpvdisC1.b1Incorrect
D141031.4SGP/VDIS/C1 - Instrument problemsgpvdisC1.b1Missing
D160203.4SGP/MET/E13 - False tips in high windssgpmetE13.b1Does not affect quality


DQRID : D140707.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/03/2014060006/12/20140000Incorrect
Subject:
SGP/VDIS/C1 - Blown light bulb resulted in incorrect
DataStreams:sgpvdisC1.b1
Description:
The light bulb burnt out resulting in no observations or incorrect observations.
Measurements:sgpvdisC1.b1:
  • Number density(num_density)
  • Moment 2 from the observed distribution(moment2)
  • Number of drops per bin(num_drops)
  • Moment 1 from the observed distribution(moment1)
  • Moment 4 from the observed distribution(moment4)
  • Rain amount(rain_amount)
  • drop_diameter
  • Moment 3 from the observed distribution(moment3)
  • Moment 6 from the observed distribution(moment6)
  • Intercept parameter, assuming an ideal Marshall-Palmer type distribution(intercept_parameter)
  • Moment 5 from the observed distribution(moment5)
  • Liquid water content(liquid_water_content)
  • Median volume diameter, assuming an ideal Marshall-Palmer type distribution(median_volume_diameter)
  • Liquid water distribution mean, assuming an ideal Marshall-Palmer type
    distribution(liquid_water_distribution_mean)
  • Rain rate(rain_rate)
  • Sum of num_drops(total_drops)
  • diameter_max
  • Slope parameter, assuming an ideal Marshall-Palmer type distribution(slope_parameter)
  • diameter_min
  • radar_reflectivity


Back To Table of Contents

DQRID : D141031.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/12/2014000010/13/20142359Missing
09/06/2014000009/07/20142359Missing
09/09/2014000009/14/20142359Missing
10/05/2014000010/07/20142359Missing
08/30/2014000009/01/20142359Missing
Subject:
SGP/VDIS/C1 - Instrument problem
DataStreams:sgpvdisC1.b1
Description:
Alignment issues resulted in a failure to collect data.
Measurements:sgpvdisC1.b1:
  • Moment 2 from the observed distribution(moment2)
  • Moment 1 from the observed distribution(moment1)
  • Moment 4 from the observed distribution(moment4)
  • Rain amount(rain_amount)
  • Moment 3 from the observed distribution(moment3)
  • Moment 6 from the observed distribution(moment6)
  • Moment 5 from the observed distribution(moment5)
  • Altitude above mean sea level(alt)
  • Median volume diameter, assuming an ideal Marshall-Palmer type distribution(median_volume_diameter)
  • East longitude(lon)
  • Liquid water distribution mean, assuming an ideal Marshall-Palmer type
    distribution(liquid_water_distribution_mean)
  • Rain rate(rain_rate)
  • time_offset
  • base_time
  • diameter_max
  • radar_reflectivity
  • Number of drops per bin(num_drops)
  • Number density(num_density)
  • time
  • drop_diameter
  • Intercept parameter, assuming an ideal Marshall-Palmer type distribution(intercept_parameter)
  • Liquid water content(liquid_water_content)
  • North latitude(lat)
  • Sum of num_drops(total_drops)
  • Slope parameter, assuming an ideal Marshall-Palmer type distribution(slope_parameter)
  • diameter_min


Back To Table of Contents

DQRID : D160203.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/21/1993000001/12/20161958Does not affect quality
Subject:
SGP/MET/E13 - False tips in high winds
DataStreams:sgpmetE13.b1
Description:
The TBRG's were mounted at a height that made them prone to vibrations caused by high 
winds which may have caused false tips.  The end date of this DQR indicates when the rain 
gauge was lowered to mitigate these effects.  

Effects on the total precipitation amounts are assumed to be negligible.
Measurements:sgpmetE13.b1:
  • tbrg_precip_total
  • Corrected total precipitation measured by TBRG(tbrg_precip_total_corr)


Back To Table of Contents



END OF DATA