Data Quality Reports for Session: 178742 User: yfq Completed: 08/01/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D140602.1MAO/CCN/M1 - Communication lossmaoaosccn100M1.a1Missing
D140606.2MAO/CPC/CCN/CLAP/PSAP/NEPHELOMETER/M1 - power outagemaoaosccnavgM1.c1, maoaosccn100M1.a1Missing
D140606.3MAO/CPC/CLAP/CCN/PSAP/NEPHELOMETER/M1 - power outagemaoaosccnavgM1.c1, maoaosccn100M1.a1Missing
D140815.1MAO/AOS/S1/M1 - Instrument maintenancemaoaosccn100M1.a1, maoaosccnavgM1.c1Incorrect
D150126.2MAO/CCN/M1 - Missing datamaoaosccn100M1.a1Missing
D150126.3MAO/AOS/CCN/CLAP/CPC/NEPHELOMETER/PSAP/M1 - Power outage. No pump flowmaoaosccn100M1.a1Incorrect
D150126.4MAO/AOS/CCN/CLAP/CPC/NEPHELOMETER/PSAP/M1 - Computer software issuemaoaosccn100M1.a1Missing
D150220.8MAO/AOS/CCN/CLAP/NEPHELOMETER/PSAP/M1 - Dilution compressor flow cycling on and offmaoaosccn100M1.a1Suspect
D150306.1MAO/CCN/M1 - Instrument undercounting of lowest super saturation valuesmaoaosccn100M1.a1Incorrect
D150415.2MAO/AOS/M1 - Weak solder connection in dilution flow meter comport cablemaoaosccn100M1.a1Incorrect
D150415.3MAO/CCN/M1 - Instrument system maintenancemaoaosccn100M1.a1Incorrect
D150430.1MAO/CCN/CPC/NEPHELOMETER/PSAP/M1 - Pump failuremaoaosccn100M1.a1Incorrect
D150921.2MAO/AOS/M1/S1 - Forest Firemaoaosccn100M1.a1Does not affect quality
D151113.1MAO/CCN/M1 - Instrument pump failuremaoaosccn100M1.a1Missing
D160301.1MAO/AOS/CCN/CLAP/CPC/NEPHELOMETER/PSAP/M1 - Reprocess: Dilution flow calibration updatemaoaosccn1colM1.b1Incorrect


DQRID : D140602.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/01/2014120002/04/20142100Missing
Subject:
MAO/CCN/M1 - Communication loss
DataStreams:maoaosccn100M1.a1
Description:
The instrument lost communication with the computer and had to be rebooted.
Measurements:maoaosccn100M1.a1:
  • CCN_ss_set
  • Size of upper limit of each CCN bin(CCN_upper_size_limit)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • CCN_Q_sheath
  • Altitude above mean sea level(alt)
  • CCN_overflow
  • CCN_first_stage_monitor_voltage
  • CCN_proportional_valve_voltage
  • CCN_inlet_set
  • CCN_temperature_gradient
  • CCN_OPC_set
  • CCN_nafion_set
  • base_time
  • AOS number concentration of CCN(N_CCN)
  • North latitude(lat)
  • CCN_T_OPC
  • CCN_t_set_TEC3
  • CCN_t_set_TEC2
  • CCN_t_set_TEC1
  • CCN_Q_sample
  • CCN_alarm_code
  • time
  • CCN_temp_unstable_flag
  • East longitude(lon)
  • CCN_T_sample
  • CCN_baseline_monitor
  • CCN_t_read_TEC3
  • CCN_t_read_TEC2
  • CCN_t_read_TEC1
  • CCN_T_inlet
  • time_offset
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_T_nafion
  • CCN_laser_current


Back To Table of Contents

DQRID : D140606.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/20/2014000002/20/20142029Missing
Subject:
MAO/CPC/CCN/CLAP/PSAP/NEPHELOMETER/M1 - power outage
DataStreams:maoaosccnavgM1.c1, maoaosccn100M1.a1
Description:
Data are missing due to the power outage.
Measurements:maoaosccn100M1.a1:
  • CCN_ss_set
  • Size of upper limit of each CCN bin(CCN_upper_size_limit)
  • CCN_Q_sheath
  • Altitude above mean sea level(alt)
  • CCN_proportional_valve_voltage
  • CCN_first_stage_monitor_voltage
  • CCN_OPC_set
  • CCN_nafion_set
  • base_time
  • AOS number concentration of CCN(N_CCN)
  • CCN_T_OPC
  • CCN_alarm_code
  • time
  • CCN_baseline_monitor
  • CCN_T_sample
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • CCN_overflow
  • CCN_inlet_set
  • CCN_temperature_gradient
  • North latitude(lat)
  • CCN_t_set_TEC3
  • CCN_t_set_TEC2
  • CCN_t_set_TEC1
  • CCN_Q_sample
  • CCN_temp_unstable_flag
  • East longitude(lon)
  • CCN_t_read_TEC3
  • CCN_t_read_TEC2
  • CCN_t_read_TEC1
  • CCN_T_inlet
  • time_offset
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_T_nafion
  • CCN_laser_current

maoaosccnavgM1.c1:
  • supersaturation_setpoint
  • time
  • time_offset
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • East longitude(lon)
  • N_CPC_std
  • Condensation particle concentration number(N_CPC)
  • AOS number concentration of CCN(N_CCN)
  • column_temperature_gradient_setting
  • N_CCN_std
  • base_time
  • column_temperature_gradient_measured_std
  • North latitude(lat)
  • droplet_size
  • column_temperature_gradient_measured
  • supersaturation_calculated
  • Altitude above mean sea level(alt)


Back To Table of Contents

DQRID : D140606.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/21/2014150702/22/20141435Missing
Subject:
MAO/CPC/CLAP/CCN/PSAP/NEPHELOMETER/M1 - power outage
DataStreams:maoaosccnavgM1.c1, maoaosccn100M1.a1
Description:
Data are missing due to the power outage.
Measurements:maoaosccn100M1.a1:
  • CCN_ss_set
  • Size of upper limit of each CCN bin(CCN_upper_size_limit)
  • CCN_Q_sheath
  • Altitude above mean sea level(alt)
  • CCN_proportional_valve_voltage
  • CCN_first_stage_monitor_voltage
  • CCN_OPC_set
  • CCN_nafion_set
  • base_time
  • AOS number concentration of CCN(N_CCN)
  • CCN_T_OPC
  • CCN_alarm_code
  • time
  • CCN_baseline_monitor
  • CCN_T_sample
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • CCN_overflow
  • CCN_inlet_set
  • CCN_temperature_gradient
  • North latitude(lat)
  • CCN_t_set_TEC3
  • CCN_t_set_TEC2
  • CCN_t_set_TEC1
  • CCN_Q_sample
  • CCN_temp_unstable_flag
  • East longitude(lon)
  • CCN_t_read_TEC3
  • CCN_t_read_TEC2
  • CCN_t_read_TEC1
  • CCN_T_inlet
  • time_offset
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_T_nafion
  • CCN_laser_current

maoaosccnavgM1.c1:
  • supersaturation_setpoint
  • time
  • time_offset
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • East longitude(lon)
  • N_CPC_std
  • Condensation particle concentration number(N_CPC)
  • AOS number concentration of CCN(N_CCN)
  • column_temperature_gradient_setting
  • N_CCN_std
  • base_time
  • column_temperature_gradient_measured_std
  • North latitude(lat)
  • droplet_size
  • column_temperature_gradient_measured
  • supersaturation_calculated
  • Altitude above mean sea level(alt)


Back To Table of Contents

DQRID : D140815.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/09/2014160008/15/20141500Incorrect
Subject:
MAO/AOS/S1/M1 - Instrument maintenance
DataStreams:maoaosccn100M1.a1, maoaosccnavgM1.c1
Description:
Instrument maintenance and cleaning inlets affecting data.
Measurements:maoaosccn100M1.a1:
  • CCN_ss_set
  • Size of upper limit of each CCN bin(CCN_upper_size_limit)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • CCN_Q_sheath
  • CCN_overflow
  • CCN_proportional_valve_voltage
  • CCN_first_stage_monitor_voltage
  • CCN_inlet_set
  • CCN_temperature_gradient
  • CCN_OPC_set
  • CCN_nafion_set
  • AOS number concentration of CCN(N_CCN)
  • CCN_T_OPC
  • CCN_t_set_TEC3
  • CCN_t_set_TEC2
  • CCN_t_set_TEC1
  • CCN_Q_sample
  • CCN_alarm_code
  • CCN_temp_unstable_flag
  • CCN_T_sample
  • CCN_baseline_monitor
  • CCN_t_read_TEC3
  • CCN_t_read_TEC2
  • CCN_t_read_TEC1
  • CCN_T_inlet
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_T_nafion
  • CCN_laser_current

maoaosccnavgM1.c1:
  • N_CPC_std
  • Condensation particle concentration number(N_CPC)
  • AOS number concentration of CCN(N_CCN)
  • supersaturation_setpoint
  • N_CCN_std
  • column_temperature_gradient_setting
  • column_temperature_gradient_measured_std
  • droplet_size
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • column_temperature_gradient_measured
  • supersaturation_calculated


Back To Table of Contents

DQRID : D150126.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/11/2015035001/11/20151226Missing
Subject:
MAO/CCN/M1 - Missing data
DataStreams:maoaosccn100M1.a1
Description:
Computer error with embedded CCN computer. Instrument was rebooted.
Measurements:maoaosccn100M1.a1:
  • CCN_ss_set
  • Size of upper limit of each CCN bin(CCN_upper_size_limit)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • CCN_Q_sheath
  • Altitude above mean sea level(alt)
  • CCN_overflow
  • CCN_first_stage_monitor_voltage
  • CCN_proportional_valve_voltage
  • CCN_inlet_set
  • CCN_temperature_gradient
  • CCN_OPC_set
  • CCN_nafion_set
  • base_time
  • AOS number concentration of CCN(N_CCN)
  • North latitude(lat)
  • CCN_T_OPC
  • CCN_t_set_TEC3
  • CCN_t_set_TEC2
  • CCN_t_set_TEC1
  • CCN_Q_sample
  • CCN_alarm_code
  • time
  • CCN_temp_unstable_flag
  • East longitude(lon)
  • CCN_T_sample
  • CCN_baseline_monitor
  • CCN_t_read_TEC3
  • CCN_t_read_TEC2
  • CCN_t_read_TEC1
  • CCN_T_inlet
  • time_offset
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_T_nafion
  • CCN_laser_current


Back To Table of Contents

DQRID : D150126.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/15/2015071201/15/20151507Incorrect
Subject:
MAO/AOS/CCN/CLAP/CPC/NEPHELOMETER/PSAP/M1 - Power outage. No pump flow
DataStreams:maoaosccn100M1.a1
Description:
Raw power which drives the pumps and heaters went out. There was no flow through the 
instruments.
Measurements:maoaosccn100M1.a1:
  • CCN_ss_set
  • Size of upper limit of each CCN bin(CCN_upper_size_limit)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • CCN_Q_sheath
  • CCN_overflow
  • CCN_first_stage_monitor_voltage
  • CCN_proportional_valve_voltage
  • CCN_inlet_set
  • CCN_temperature_gradient
  • CCN_OPC_set
  • CCN_nafion_set
  • AOS number concentration of CCN(N_CCN)
  • CCN_T_OPC
  • CCN_t_set_TEC3
  • CCN_t_set_TEC2
  • CCN_t_set_TEC1
  • CCN_Q_sample
  • CCN_alarm_code
  • CCN_temp_unstable_flag
  • CCN_T_sample
  • CCN_baseline_monitor
  • CCN_t_read_TEC3
  • CCN_t_read_TEC2
  • CCN_t_read_TEC1
  • CCN_T_inlet
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_T_nafion
  • CCN_laser_current


Back To Table of Contents

DQRID : D150126.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/17/2015120001/18/20151843Missing
Subject:
MAO/AOS/CCN/CLAP/CPC/NEPHELOMETER/PSAP/M1 - Computer software issue
DataStreams:maoaosccn100M1.a1
Description:
The computer software stopped collecting data and was rebooted.
Measurements:maoaosccn100M1.a1:
  • CCN_ss_set
  • Size of upper limit of each CCN bin(CCN_upper_size_limit)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • CCN_Q_sheath
  • Altitude above mean sea level(alt)
  • CCN_overflow
  • CCN_first_stage_monitor_voltage
  • CCN_proportional_valve_voltage
  • CCN_inlet_set
  • CCN_temperature_gradient
  • CCN_OPC_set
  • CCN_nafion_set
  • base_time
  • AOS number concentration of CCN(N_CCN)
  • North latitude(lat)
  • CCN_T_OPC
  • CCN_t_set_TEC3
  • CCN_t_set_TEC2
  • CCN_t_set_TEC1
  • CCN_Q_sample
  • CCN_alarm_code
  • time
  • CCN_temp_unstable_flag
  • East longitude(lon)
  • CCN_T_sample
  • CCN_baseline_monitor
  • CCN_t_read_TEC3
  • CCN_t_read_TEC2
  • CCN_t_read_TEC1
  • CCN_T_inlet
  • time_offset
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_T_nafion
  • CCN_laser_current


Back To Table of Contents

DQRID : D150220.8
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/02/2015162002/03/20151550Suspect
Subject:
MAO/AOS/CCN/CLAP/NEPHELOMETER/PSAP/M1 - Dilution compressor flow cycling on and off
DataStreams:maoaosccn100M1.a1
Description:
The dilution flow was cycling on and off. This led to unstable concentrations. Dilution 
flow is about 30% of total flow so concentration uncertainty is on the order of 30%.
Measurements:maoaosccn100M1.a1:
  • CCN_ss_set
  • Size of upper limit of each CCN bin(CCN_upper_size_limit)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • CCN_Q_sheath
  • CCN_overflow
  • CCN_first_stage_monitor_voltage
  • CCN_proportional_valve_voltage
  • CCN_inlet_set
  • CCN_temperature_gradient
  • CCN_OPC_set
  • CCN_nafion_set
  • AOS number concentration of CCN(N_CCN)
  • CCN_T_OPC
  • CCN_t_set_TEC3
  • CCN_t_set_TEC2
  • CCN_t_set_TEC1
  • CCN_Q_sample
  • CCN_alarm_code
  • CCN_temp_unstable_flag
  • CCN_T_sample
  • CCN_baseline_monitor
  • CCN_t_read_TEC3
  • CCN_t_read_TEC2
  • CCN_t_read_TEC1
  • CCN_T_inlet
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_T_nafion
  • CCN_laser_current


Back To Table of Contents

DQRID : D150306.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/25/2014000003/06/20151619Incorrect
Subject:
MAO/CCN/M1 - Instrument undercounting of lowest super saturation values
DataStreams:maoaosccn100M1.a1
Description:
The slow temperature decay of the OPC induced droplet evaporation and caused an under 
counting of the droplets at the lowest %ss, 0.15.
Measurements:maoaosccn100M1.a1:
  • CCN_ss_set


Back To Table of Contents

DQRID : D150415.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/19/2015041903/02/20152121Incorrect
04/06/2015071104/11/20151833Incorrect
Subject:
MAO/AOS/M1 - Weak solder connection in dilution flow meter comport cable
DataStreams:maoaosccn100M1.a1
Description:
The dilution flow meter reading was erratic. Plots of the data suggest that the flow 
varied by about 5%. Problem was corrected when wire was re-soldered.
Measurements:maoaosccn100M1.a1:
  • AOS number concentration of CCN(N_CCN)
  • AOS CCN droplet count by bin size(N_CCN_dN)


Back To Table of Contents

DQRID : D150415.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/24/2015145303/28/20151717Incorrect
Subject:
MAO/CCN/M1 - Instrument system maintenance
DataStreams:maoaosccn100M1.a1
Description:
AOS system down for maintenance
Measurements:maoaosccn100M1.a1:
  • CCN_ss_set
  • Size of upper limit of each CCN bin(CCN_upper_size_limit)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • CCN_Q_sheath
  • CCN_overflow
  • CCN_first_stage_monitor_voltage
  • CCN_proportional_valve_voltage
  • CCN_inlet_set
  • CCN_temperature_gradient
  • CCN_OPC_set
  • CCN_nafion_set
  • AOS number concentration of CCN(N_CCN)
  • CCN_T_OPC
  • CCN_t_set_TEC3
  • CCN_t_set_TEC2
  • CCN_t_set_TEC1
  • CCN_Q_sample
  • CCN_alarm_code
  • CCN_temp_unstable_flag
  • CCN_T_sample
  • CCN_baseline_monitor
  • CCN_t_read_TEC3
  • CCN_t_read_TEC2
  • CCN_t_read_TEC1
  • CCN_T_inlet
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_T_nafion
  • CCN_laser_current


Back To Table of Contents

DQRID : D150430.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/03/2015105505/03/20152136Incorrect
04/29/2015141204/30/20151740Incorrect
Subject:
MAO/CCN/CPC/NEPHELOMETER/PSAP/M1 - Pump failure
DataStreams:maoaosccn100M1.a1
Description:
Pump failed, no flow through instruments.
Measurements:maoaosccn100M1.a1:
  • CCN_ss_set
  • Size of upper limit of each CCN bin(CCN_upper_size_limit)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • CCN_Q_sheath
  • CCN_overflow
  • CCN_first_stage_monitor_voltage
  • CCN_proportional_valve_voltage
  • CCN_inlet_set
  • CCN_temperature_gradient
  • CCN_OPC_set
  • CCN_nafion_set
  • AOS number concentration of CCN(N_CCN)
  • CCN_T_OPC
  • CCN_t_set_TEC3
  • CCN_t_set_TEC2
  • CCN_t_set_TEC1
  • CCN_Q_sample
  • CCN_alarm_code
  • CCN_temp_unstable_flag
  • CCN_T_sample
  • CCN_baseline_monitor
  • CCN_t_read_TEC3
  • CCN_t_read_TEC2
  • CCN_t_read_TEC1
  • CCN_T_inlet
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_T_nafion
  • CCN_laser_current


Back To Table of Contents

DQRID : D150921.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/04/2015000010/08/20152359Does not affect quality
09/03/2015180009/03/20152359Does not affect quality
09/13/2015165309/13/20151810Does not affect quality
Subject:
MAO/AOS/M1/S1 - Forest Fire
DataStreams:maoaosccn100M1.a1
Description:
Smoke from forest burning can be seen on the TSI images rising and then subsequently 
moving over the site during this period.  Many of the AOS instruments observed a sharp spike 
in data values.

October 4-8 multiple fires in the area resulted in hazy conditions during this entire 
period.

Most notable in the following instruments and variables

ACSM - Total Organics
AETH - Black Carbon Concentrations at all wavelengths
CCN100 - Spike in number concentrations occurred a few hours after the smoke plume
CLAP3W - Absorption Coefficient
CPC - Similar to the CCN100, spike in particle concentration occurred a few hours later
CPCF - Particle Concentration
CPCU - Particle Concentration
NEPHELOMETERS - Scattering/Backscattering Coeff
PSAP3W - Absorption Coefficient
SMPS - Concentration and Size Distribution
UHSAS - Size Distribution
Measurements:maoaosccn100M1.a1:
  • AOS number concentration of CCN(N_CCN)
  • AOS CCN droplet count by bin size(N_CCN_dN)


Back To Table of Contents

DQRID : D151113.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/10/2015175611/11/20152107Missing
Subject:
MAO/CCN/M1 - Instrument pump failure
DataStreams:maoaosccn100M1.a1
Description:
There is no data during this time
Measurements:maoaosccn100M1.a1:
  • CCN_ss_set
  • Size of upper limit of each CCN bin(CCN_upper_size_limit)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • CCN_Q_sheath
  • Altitude above mean sea level(alt)
  • CCN_overflow
  • CCN_first_stage_monitor_voltage
  • CCN_proportional_valve_voltage
  • CCN_inlet_set
  • CCN_temperature_gradient
  • CCN_OPC_set
  • CCN_nafion_set
  • base_time
  • AOS number concentration of CCN(N_CCN)
  • North latitude(lat)
  • CCN_T_OPC
  • CCN_t_set_TEC3
  • CCN_t_set_TEC2
  • CCN_t_set_TEC1
  • CCN_Q_sample
  • CCN_alarm_code
  • time
  • CCN_temp_unstable_flag
  • East longitude(lon)
  • CCN_T_sample
  • CCN_baseline_monitor
  • CCN_t_read_TEC3
  • CCN_t_read_TEC2
  • CCN_t_read_TEC1
  • CCN_T_inlet
  • time_offset
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_T_nafion
  • CCN_laser_current


Back To Table of Contents

DQRID : D160301.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/01/2015000011/08/20150000Incorrect
02/19/2015042003/01/20152122Incorrect
05/15/2015120007/01/20150000Incorrect
11/08/2015000012/01/20150000Incorrect
Subject:
MAO/AOS/CCN/CLAP/CPC/NEPHELOMETER/PSAP/M1 - Reprocess: Dilution flow calibration update
DataStreams:maoaosccn1colM1.b1
Description:
Dilution flow calibrations are incorrect. Dilution flow corrections need to be applied to 
b1 data.

Flow: 11.1 lpm 2015-02-19 04:19:12 to 2015-03-02 21:21:36 UTC
Flow: 9.75 lpm 2015-05-15 00:00:00 to 2015-07-01 23:59:59 UTC
Flow: 3.7178 slpm 2015-07-01 00:00:00 to 2015-11-08 23:59:59 UTC
Flow: 9.324 slpm 2015-11-08 00:00:00 UTC to 2015-12-01 23:59:59 UTC
Measurements:maoaosccn1colM1.b1:
  • time_offset
  • T_set_TEC2
  • eta
  • T_inlet
  • T_set_TEC3
  • first_stage_monitor_voltage
  • Altitude above mean sea level(alt)
  • T_set_TEC1
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • reported_temperature_gradient
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • laser_current
  • eta_target
  • overflow
  • size_bin_bounds
  • North latitude(lat)
  • dT_target_estimated
  • base_time
  • proportional_valve_voltage
  • T_sample
  • dilution_correction_factor
  • temp_unstable
  • seconds_after_transition
  • supersaturation_calculated
  • P_sample
  • T_read_TEC3
  • size_bin
  • T_read_TEC2
  • Q_sample
  • dT_OPC
  • eta_lookup_table
  • East longitude(lon)
  • time
  • temperature_std
  • T_read_TEC1
  • supersaturation_set_point
  • T_set_gradient
  • T_nafion
  • Q_sheath
  • supersaturation_calculated_target
  • AOS number concentration of CCN(N_CCN)
  • T_read_gradient
  • T_OPC
  • qc_N_CCN


Back To Table of Contents



END OF DATA