Data Quality Reports for Session: 178317 User: stemmlerj1 Completed: 07/11/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D160411.1ENA/UHSAS/C1 - Intermittent spikes in the lower size binsenaaosuhsasC1.a1
D160411.2ENA/UHSAS/C1 - Instrument not sampling ambient airenaaosuhsasC1.a1
D160427.7ENA/C1 - Reprocess: Slightly incorrect location fieldsenaaoscpcfC1.a1, enametC1.b1
D160610.10ENA/UHSAS/C1 - High counts in lower size binsenaaosuhsasC1.a1
D160610.9ENA/UHSAS/C1 - missing dataenaaosuhsasC1.a1
D160613.1ENA/CAPS-PMEX/NEPHELOMETER/PSAP/C1 - Impactor not switchingenaaosnephdryC1.a1, enaaosnephdryC1.b1
D160701.3ENA/AOS/C1 -enaaoscpcfC1.a1


DQRID : D160411.1
Start DateStart TimeEnd DateEnd Time
03/28/2016220004/01/20162330
Subject:
ENA/UHSAS/C1 - Intermittent spikes in the lower size bins
DataStreams:enaaosuhsasC1.a1
Description:
Intermittent spikes in the 3 lowest size bins. Other size bins should not be affected.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D160411.2
Start DateStart TimeEnd DateEnd Time
03/31/2016160004/01/20161025
03/30/2016090003/30/20161100
03/30/2016153003/30/20161900
Subject:
ENA/UHSAS/C1 - Instrument not sampling ambient air
DataStreams:enaaosuhsasC1.a1
Description:
Instrument was periodically offline or not sampling ambient air for troubleshooting 
purposes. Data missing or no valid data.
Measurements:enaaosuhsasC1.a1:
  • scattering_voltage
  • base_time
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)
  • sheath_flow
  • MET air temperature(temperature)
  • sampling_volume
  • purge_flow
  • unit_temperature
  • auxiliary
  • reference
  • current
  • sample_pressure
  • flow
  • upper_size_limit
  • lower_size_limit
  • differential_voltage
  • bin_num


Back To Table of Contents

DQRID : D160427.7
Start DateStart TimeEnd DateEnd Time
09/29/2013094104/30/20162359
Subject:
ENA/C1 - Reprocess: Slightly incorrect location fields
DataStreams:enaaoscpcfC1.a1, enametC1.b1
Description:
Some of the instruments are being processed with the previous "GRW-Azores" lat/lon/alt 
instead of the new "ENA-Azores" lat/lon/alt. The differences are minor, but will be 
corrected. The respective GRW and ENA lat, lon, and alt values are as follows:
 
GRW  lat = 39.0911 ;
ENA lat = 39.0916 ;
GRW  lon = -28.0297 ;
ENA  lon = -28.0257 ;
GRW  alt = 26 ;
ENA  alt = 30.48 ;
Measurements:enaaoscpcfC1.a1:
  • East longitude(lon)
  • Altitude above mean sea level(alt)
  • North latitude(lat)

enametC1.b1:
  • Altitude above mean sea level(alt)
  • North latitude(lat)
  • East longitude(lon)


Back To Table of Contents

DQRID : D160610.10
Start DateStart TimeEnd DateEnd Time
06/07/2016134006/08/20160200
Subject:
ENA/UHSAS/C1 - High counts in lower size bins
DataStreams:enaaosuhsasC1.a1
Description:
High counts in the 3 lowest size bins. Other size bins should not be affected.
Measurements:enaaosuhsasC1.a1:
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)


Back To Table of Contents

DQRID : D160610.9
Start DateStart TimeEnd DateEnd Time
06/06/2016221006/07/20161340
Subject:
ENA/UHSAS/C1 - missing data
DataStreams:enaaosuhsasC1.a1
Description:
Instrument's internal hard drive was full and no new data was being saved except a few 
samples occasionally.
Measurements:enaaosuhsasC1.a1:
  • scattering_voltage
  • time
  • base_time
  • Particle size distribution, limited by lower_size_limit and upper_size_limit(size_distribution)
  • sheath_flow
  • Altitude above mean sea level(alt)
  • MET air temperature(temperature)
  • North latitude(lat)
  • sampling_volume
  • time_offset
  • purge_flow
  • unit_temperature
  • auxiliary
  • current
  • reference
  • sample_pressure
  • lower_size_limit
  • flow
  • upper_size_limit
  • differential_voltage
  • East longitude(lon)
  • bin_num


Back To Table of Contents

DQRID : D160613.1
Start DateStart TimeEnd DateEnd Time
06/10/2016150006/12/20161100
Subject:
ENA/CAPS-PMEX/NEPHELOMETER/PSAP/C1 - Impactor not switching
DataStreams:enaaosnephdryC1.a1, enaaosnephdryC1.b1
Description:
Impactor power accidentally unplugged. Impactor not switching and stuck in 10um position. 
The quality of data variables is not affected.
Measurements:enaaosnephdryC1.a1:
  • impactor_setting

enaaosnephdryC1.b1:
  • impactor_setting
  • Impactor particle size cut(impactor_state)


Back To Table of Contents

DQRID : D160701.3
Start DateStart TimeEnd DateEnd Time
06/20/2016182006/21/20160835
Subject:
ENA/AOS/C1 -
DataStreams:enaaoscpcfC1.a1
Description:
During a measurement of the inlet flow rate, the cpcf inlet became plugged and the 
instrument was flooded with butanol.  After removing the butanol and allowing the system to dry, 
the cpc returned to normal operation.
Measurements:enaaoscpcfC1.a1:
  • Concentration(concentration)


Back To Table of Contents



END OF DATA