Data Quality Reports for Session: 178107 User: balmesk1 Completed: 06/27/2016


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D160620.1SGP/RL/C1 - Data acquisition system malfunctionsgprllogC1.a0


DQRID : D160620.1
Start DateStart TimeEnd DateEnd Time
05/21/2016040006/14/20161530
Subject:
SGP/RL/C1 - Data acquisition system malfunction
DataStreams:sgprllogC1.a0
Description:
Due to a malfunction in the data acquisition system, the narrow-field-of-view (N2, Dep, 
Copol, H2O, T1, and T2) photon counting signals experienced frequent intermittent drop outs 
during this period. The problem was solved when the 15-V power supply in one of the 
data-acquisition racks was replaced.
Measurements:sgprllogC1.a0:
  • counts


Back To Table of Contents



END OF DATA