DQRID : D140121.3
Start DateStart TimeEnd DateEnd Time
01/15/2014000002/11/20140000
12/28/2013000001/09/20140059
Subject:
TWP/RL/C3 - Missing T1 and/or T2 (rotational Raman) channels
DataStreams:twprlprofmerge1turnC3.c0
Description:
Data for the T1 and/or T2 (rotation Raman) channels are missing due to failures of 
multiple high voltage power supplies for PMTs.  As a result, it is not be possible to derive 
temperature profiles from the Raman lidar during the effected period.

Temperature profile data from the Raman lidar are incorrect. Do not use for analysis.
Measurements:twprlprofmerge1turnC3.c0:
  • t1_counts_high
  • electronic_background_t2_analog_high
  • t2_counts_high_glue_coef_offset
  • t1_counts_high_glue_coef_offset
  • t2_counts_high_glue_coef_slope
  • electronic_background_t1_analog_high
  • t2_counts_high
  • t1_counts_high_glue_coef_slope


Back To Table of Contents

DQRID : D140223.1
Start DateStart TimeEnd DateEnd Time
01/11/2014180001/15/20142300
Subject:
TWP/RL/C3 - Invalid depolarization channel data
DataStreams:twprlprofmerge1turnC3.c0
Description:
The depolarization channel was not working properly during the period in question due to a 
failure of the high voltage power supply for the PMT.
Measurements:twprlprofmerge1turnC3.c0:
  • electronic_background_depolarization_analog_high
  • depolarization_counts_high_glue_coef_offset
  • depolarization_counts_high
  • depolarization_counts_high_glue_coef_slope


Back To Table of Contents