Data Quality Reports for Session: 154756 User: immorhuang Completed: 11/24/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D130110.1NSA/SEBS/E10 - Tower Failure, Data Not AffectednsasebsE10.b1
D130110.2NSA/ECOR/E10 - Missing datansa30ecorE10.b1
D130411.1NSA/ECOR/E10 - CO2 and H2O Fluxes Missingnsa30ecorE10.b1
D131005.2NSA/ECOR/E10 - Missing Pressure and Temperature Datansa30ecorE10.b1


DQRID : D130110.1
Start DateStart TimeEnd DateEnd Time
12/22/2012220001/03/20131600
Subject:
NSA/SEBS/E10 - Tower Failure, Data Not Affected
DataStreams:nsasebsE10.b1
Description:
The tripod tower top holding the net radiometer broke and collapsed.

However, at this time of year radiation is so small that effects on the radiation and 
therefore net radiation measurements can not be detected.

Therefore this DQR has been designated as transparent.
Measurements:nsasebsE10.b1:
  • up_long
  • Radiation, shortwave, downwelling total hemispheric irradiance, 1-min avg(down_short_hemisp)
  • net radiation
  • down_long
  • Radiation, shortwave, at 10-m height, upwelling hemispheric irrad., 1-min avg(up_short_hemisp)
  • Radiation, shortwave, upwelling, at level, gridded(albedo)
  • Surface energy balance(surface_energy_balance)


Back To Table of Contents

DQRID : D130110.2
Start DateStart TimeEnd DateEnd Time
12/22/2012220001/03/20131600
Subject:
NSA/ECOR/E10 - Missing data
DataStreams:nsa30ecorE10.b1
Description:
Data are missing and unrecoverable.
Measurements:nsa30ecorE10.b1:
  • variance of variable c(var_c)
  • covariance qc(cvar_qc)
  • n_bad_son_ic
  • variance of variable q(var_q)
  • variance of variable t(var_t)
  • variance of variable v(var_v)
  • variance of variable u(var_u)
  • variance of variable w(var_w)
  • lon
  • Corrected sensible heat flux(h)
  • real_rho
  • k
  • real_cp
  • Latent heat flux(lv_e)
  • wind_spd
  • phi
  • mean_rot_v
  • Mean horizontal wind speed(mean_rot_u)
  • mean_rot_w
  • Mean H2O density(mean_q)
  • rotated covariance vt(cvar_rot_vt)
  • rotated covariance vw(cvar_rot_vw)
  • Mean sonic temperature (t), i.e. virtual temperature(mean_t)
  • mean_u
  • wind_dir
  • mean_v
  • mean_w
  • lv
  • rotated covariance vq(cvar_rot_vq)
  • kurt_c
  • skew_c
  • n_good_c
  • time
  • rotated covariance vc(cvar_rot_vc)
  • skew_q
  • covariance tq(cvar_tq)
  • skew_u
  • skew_t
  • skew_w
  • covariance tc(cvar_tc)
  • skew_v
  • lat
  • rotated covariance wt(cvar_rot_wt)
  • rotated covariance wq(cvar_rot_wq)
  • Mean CO2 density(mean_c)
  • standard deviation of wind elevation angle(std_elev)
  • time_offset
  • Pressure, condensation, at altitude, RUC model output(cp)
  • friction velocity(ustar)
  • rotated covariance wc(cvar_rot_wc)
  • mean_spk_q
  • covariance uc(cvar_uc)
  • CO2 flux(fc)
  • alt
  • covariance uq(cvar_uq)
  • mean_spk_c
  • covariance uw(cvar_uw)
  • n_spk_t
  • n_spk_u
  • n_spk_v
  • n_spk_w
  • atm_pres
  • covariance ut(cvar_ut)
  • covariance uv(cvar_uv)
  • mean_spk_v
  • mean_spk_w
  • covaraince vc(cvar_vc)
  • mean_spk_t
  • n_spk_q
  • mean_spk_u
  • n_good_u
  • n_good_t
  • n_good_w
  • n_good_v
  • n_good_q
  • kurt_q
  • kurt_t
  • n_bad_irga_light
  • kurt_u
  • covariance vq(cvar_vq)
  • rotated covariance uc(cvar_rot_uc)
  • kurt_v
  • kurt_w
  • rotated covariance uq(cvar_rot_uq)
  • covariance vw(cvar_vw)
  • covariance vt(cvar_vt)
  • covariance wc(cvar_wc)
  • rotated covariance uw(cvar_rot_uw)
  • rotated covariance ut(cvar_rot_ut)
  • rotated covariance uv(cvar_rot_uv)
  • Water vapor mixing ratio observed by the Raman lidar(mr)
  • elev
  • covaraince wq(cvar_wq)
  • covariance wt(cvar_wt)
  • real_mr
  • standard deviation of wind direction(std_wind_dir)
  • real_lv
  • base_time
  • temp_irga
  • rho
  • n_bad_w
  • n_bad_v
  • n_bad_u
  • n_bad_t
  • rotated variance v(var_rot_v)
  • rotated variance u(var_rot_u)
  • n_bad_q
  • n_bad_irga
  • rotated variance w(var_rot_w)
  • n_spk_c
  • n_bad_c
  • mean_cooler


Back To Table of Contents

DQRID : D130411.1
Start DateStart TimeEnd DateEnd Time
01/03/2013163004/09/20131935
Subject:
NSA/ECOR/E10 - CO2 and H2O Fluxes Missing
DataStreams:nsa30ecorE10.b1
Description:
The CO2/H2O sensor head cable was disconnected when the ECOR tripod failure occurred on 
12/22/2012.  The problem was rectified by plugging in the cable and cleaning the LI-7500 
sensor head windows.
Measurements:nsa30ecorE10.b1:
  • Mean H2O density(mean_q)
  • variance of variable c(var_c)
  • mean_spk_q
  • covariance uc(cvar_uc)
  • CO2 flux(fc)
  • covariance qc(cvar_qc)
  • covariance uq(cvar_uq)
  • mean_spk_c
  • rotated covariance vq(cvar_rot_vq)
  • covaraince wq(cvar_wq)
  • variance of variable q(var_q)
  • skew_c
  • kurt_c
  • n_good_c
  • covaraince vc(cvar_vc)
  • n_spk_q
  • rotated covariance vc(cvar_rot_vc)
  • n_bad_q
  • skew_q
  • n_good_q
  • covariance tq(cvar_tq)
  • kurt_q
  • n_spk_c
  • covariance vq(cvar_vq)
  • covariance tc(cvar_tc)
  • rotated covariance uc(cvar_rot_uc)
  • rotated covariance wq(cvar_rot_wq)
  • Latent heat flux(lv_e)
  • Mean CO2 density(mean_c)
  • n_bad_c
  • rotated covariance uq(cvar_rot_uq)
  • covariance wc(cvar_wc)
  • rotated covariance wc(cvar_rot_wc)


Back To Table of Contents

DQRID : D131005.2
Start DateStart TimeEnd DateEnd Time
09/26/2013010010/01/20131930
Subject:
NSA/ECOR/E10 - Missing Pressure and Temperature Data
DataStreams:nsa30ecorE10.b1
Description:
IRGA temperature and pressure was missing during this period. Power cycling the IRGA 
solved this problem.
Measurements:nsa30ecorE10.b1:
  • atm_pres
  • temp_irga


Back To Table of Contents



END OF DATA