Data Quality Reports for Session: 154394 User: aamatthews Completed: 11/08/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D060928.3SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdrysgpsondewnpnC1.b1
D120215.4SGP/MERGESONDE/C1 - Reprocess: Precip data incorrectsgpmergesonde1maceC1.c1


DQRID : D060928.3
Start DateStart TimeEnd DateEnd Time
09/28/2006172812/31/99992359
Subject:
SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry
DataStreams:sgpsondewnpnC1.b1
Description:
The output precision of the following variables was increased 20060928

alt  changed from 1m to 0.1m
dp   changed from 0.1degC to 0.01degC
pres changed from 0.1 hPa to 0.01 hPa
RH   changed from 1%RH to 0.01%RH
tdry changed from .1degC to 0.01degC

Note that the increase in precision does not necessarily imply an increase in accuracy.  
Users are cautioned to continue to use their best scientific judgement when using these or 
any other data.
Measurements:sgpsondewnpnC1.b1:
  • alt
  • Dry Bulb Temperature(tdry)
  • Dewpoint Temperature(dp)
  • Humidity, relative(rh)
  • Pressure, atmospheric, constant value of model level, RUC model output(pres)


Back To Table of Contents

DQRID : D120215.4
Start DateStart TimeEnd DateEnd Time
01/01/2006000005/30/20122359
Subject:
SGP/MERGESONDE/C1 - Reprocess: Precip data incorrect
DataStreams:sgpmergesonde1maceC1.c1
Description:
All precip values are incorrectly reported during the period:  

From 20060101-20101231 all precip values are incorrectly reported as missing, or -9999.  

From 20110101-present all precip values are reported as 0 even though the qc_precip 
variable flags the data as missing, or -9999. 

The precipitation data from the co-located surface met station were not actually missing, 
but the MERGESONDE VAP did not properly ingest these data.  While the 0 precip values 
reported after 20110101 in MERGESONDE may be correct on days with no precip, they are not 
actually tied to the measurements from the co-located rain gauge and are due to a VAP 
software error.

To correct this issue, the MERGESONDE software will need to be corrected so the data can 
be reprocessed.  Because precip is a secondary variable in MERGESONDE, this issue is 
considered low priority and the VAP will not be reprocessed until a sufficient number of 
issues in the VAP have been identified.
Measurements:sgpmergesonde1maceC1.c1:
  • Precipitation, total(precip)


Back To Table of Contents