Data Quality Reports for Session: 154314 User: koontz Completed: 11/06/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D130709.1SGP/SIRS/C1 - NIP sensor failuresgpsirs20sC1.a0, sgpsirsC1.b1


DQRID : D130709.1
Start DateStart TimeEnd DateEnd Time
06/28/2013012507/05/20131540
Subject:
SGP/SIRS/C1 - NIP sensor failure
DataStreams:sgpsirs20sC1.a0, sgpsirsC1.b1
Description:
The short_direct_normal data are missing due to an instrument failure (NIP pyhrheliometer) 
likely due to a nearby lightning strike near the time of initial failure.

The short_direct_normal data are not recoverable during this time period.  However, direct 
normal data can be reconstructed from the down_short_diffuse_hemisp and 
down_short_hemisp using the following relationship for each data point:
short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle)
Measurements:sgpsirs20sC1.a0:
  • inst_direct_normal

sgpsirsC1.b1:
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
  • short_direct_normal_std
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)


Back To Table of Contents



END OF DATA