Data Quality Reports for Session: 153835 User: lkberg Completed: 10/15/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D121016.3PVC/IRT/M1 - Missing datapvcirt200msM1.a1, pvcirtM1.b1
D121018.11PVC/IRT/M1 - Reprocess: Incorrect Configurationpvcirt200msM1.a1, pvcirtM1.b1
D121218.5PVC/IRT/M1 - Instrument failurepvcirt200msM1.a1, pvcirtM1.b1


DQRID : D121016.3
Start DateStart TimeEnd DateEnd Time
07/10/2012230009/05/20121306
Subject:
PVC/IRT/M1 - Missing data
DataStreams:pvcirt200msM1.a1, pvcirtM1.b1
Description:
Data are missing and unrecoverable.
Measurements:pvcirt200msM1.a1:
  • lon
  • IR Brightness Temperature(sky_ir_temp)
  • alt
  • lat
  • time
  • time_offset
  • ref_ir_temp
  • base_time

pvcirtM1.b1:
  • IR Brightness Temperature(sky_ir_temp)
  • ref_ir_temp
  • lon
  • sky_ir_temp_std
  • sky_ir_temp_max
  • sky_ir_temp_min
  • base_time
  • time
  • samples_in_average
  • lat
  • time_offset
  • alt


Back To Table of Contents

DQRID : D121018.11
Start DateStart TimeEnd DateEnd Time
06/29/2012224010/18/20121259
Subject:
PVC/IRT/M1 - Reprocess: Incorrect Configuration
DataStreams:pvcirt200msM1.a1, pvcirtM1.b1
Description:
The SKYRAD IRT was installed temporarily (waiting for the new enclosure) in a vertical 
orientation, without the use of the gold mirror. When irt_comm was started, the emissivity 
parameter in the startup file was not changed to 1.000 compensate for the absence of the 
mirror, resulting in a positive bias of about 5 K.

These data may be corrected by adjusting the radiance values for the difference in 
emmissivity.
Measurements:pvcirt200msM1.a1:
  • IR Brightness Temperature(sky_ir_temp)

pvcirtM1.b1:
  • IR Brightness Temperature(sky_ir_temp)
  • sky_ir_temp_max
  • sky_ir_temp_min


Back To Table of Contents

DQRID : D121218.5
Start DateStart TimeEnd DateEnd Time
10/05/2012142012/04/20121434
Subject:
PVC/IRT/M1 - Instrument failure
DataStreams:pvcirt200msM1.a1, pvcirtM1.b1
Description:
The IRT failed because of water intrusion. It was replaced with a unit installed inside a 
ventilated enclosure.
Measurements:pvcirt200msM1.a1:
  • IR Brightness Temperature(sky_ir_temp)
  • ref_ir_temp

pvcirtM1.b1:
  • IR Brightness Temperature(sky_ir_temp)
  • ref_ir_temp
  • sky_ir_temp_std
  • sky_ir_temp_max
  • sky_ir_temp_min


Back To Table of Contents



END OF DATA