Data Quality Reports for Session: 153762 User: manasi1mah Completed: 10/13/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D091106.1SGP/DISDROMETER/C1 - False reports of precipsgpdisdrometerC1.b1
D120215.4SGP/MERGESONDE/C1 - Reprocess: Precip data incorrectsgpmergesonde1maceC1.c1


DQRID : D091106.1
Start DateStart TimeEnd DateEnd Time
06/01/2008000007/01/20090000
Subject:
SGP/DISDROMETER/C1 - False reports of precip
DataStreams:sgpdisdrometerC1.b1
Description:
A number of days during this period show indications of rain on dry days.  This occurred 
during windy periods when vibrations of the splash guard resulted in many small drops 
being recorded.  The splash guard has been removed and the Instrument Handbook has been 
updated with a larger section describing this issue (Section 5.3 on known problems).
Measurements:sgpdisdrometerC1.b1:
  • Precipitation(precip_dis)
  • d_max
  • nd
  • Rainfall rate(rain_rate)
  • liq_water
  • num_drop
  • n_0


Back To Table of Contents

DQRID : D120215.4
Start DateStart TimeEnd DateEnd Time
01/01/2006000005/30/20122359
Subject:
SGP/MERGESONDE/C1 - Reprocess: Precip data incorrect
DataStreams:sgpmergesonde1maceC1.c1
Description:
All precip values are incorrectly reported during the period:  

From 20060101-20101231 all precip values are incorrectly reported as missing, or -9999.  

From 20110101-present all precip values are reported as 0 even though the qc_precip 
variable flags the data as missing, or -9999. 

The precipitation data from the co-located surface met station were not actually missing, 
but the MERGESONDE VAP did not properly ingest these data.  While the 0 precip values 
reported after 20110101 in MERGESONDE may be correct on days with no precip, they are not 
actually tied to the measurements from the co-located rain gauge and are due to a VAP 
software error.

To correct this issue, the MERGESONDE software will need to be corrected so the data can 
be reprocessed.  Because precip is a secondary variable in MERGESONDE, this issue is 
considered low priority and the VAP will not be reprocessed until a sufficient number of 
issues in the VAP have been identified.
Measurements:sgpmergesonde1maceC1.c1:
  • Precipitation, total(precip)


Back To Table of Contents



END OF DATA