Data Quality Reports for Session: 153689 User: bsyin1 Completed: 10/09/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D010831.3SGP/MFR10M/C1 - Head temperature problemssgpmfr10mC1.a0, sgpmfr10mC1.b1
D020620.42SGP/MFR25M/C1 - Loose connectionssgpmfr25mC1.a0, sgpmfr25mC1.b1
D030916.7SGP/MFR10M/C1 - Artifact in datasgpmfr10mC1.a0, sgpmfr10mC1.b1
D050218.3SGP/MFR10M/C1 - Data discontinuitiessgpmfr10mC1.a0, sgpmfr10mC1.b1
D050218.4SGP/MFR10M/C1 - Dip in data at solar noonsgpmfr10mC1.a0, sgpmfr10mC1.b1
D060411.7SGP/MFR10M/C1 - Missing datasgpmfr10mC1.b1, sgpmfr10mC1.a0
D090226.8SGP/MFR25M/C1 - Missing datasgpmfr25mC1.a0, sgpmfr25mC1.b1
D090520.3SGP/MFR10M/C1 - Heater failuresgpmfr10mC1.b1, sgpmfr10mC1.a0
D091112.3SGP/MFR10M/C1 - Bad connectionsgpmfr10mC1.a0, sgpmfr10mC1.b1
D100722.1SGP/MFR10M/C1 - Poor connectionsgpmfr10mC1.a0, sgpmfr10mC1.b1
D120229.2SGP/MFR10M/C1 - Noise in ch415, sensor degradationsgpmfr10mC1.a0, sgpmfr10mC1.b1
D120229.3SGP/MFR10M/C1 - Channel 615 deadsgpmfr10mC1.a0, sgpmfr10mC1.b1
D120229.4SGP/MFR10M/C1 - Broadband channel failuresgpmfr10mC1.a0, sgpmfr10mC1.b1
D941227.4SGP/MFR10M/C1 - Tower lowered for maintenancesgpmfr10mC1.a0, sgpmfr10mC1.b1
D950327.1SGP/MFR10M/C1 - New datalogger PROMsgpmfr10mC1.a0, sgpmfr10mC1.b1
D960914.17SGP/MFR10m/C1 - Small intermittent time stamp errorsgpmfr10mC1.a0, sgpmfr10mC1.b1
D960914.18SGP/MFR25M/C1 - Intermittent small time stamp errorsgpmfr25mC1.a0, sgpmfr25mC1.b1
D961008.1SGP/MFR10M/C1 - Corrupt or truncated datasgpmfr10mC1.a0, sgpmfr10mC1.b1
D961010.1SGP/MFR25M/C1 - poor data qualitysgpmfr25mC1.a0, sgpmfr25mC1.b1
D990906.1SGP/MFR10M/C1 - Head heater not workingsgpmfr10mC1.a0, sgpmfr10mC1.b1


DQRID : D010831.3
Start DateStart TimeEnd DateEnd Time
06/11/2001000009/30/20012359
Subject:
SGP/MFR10M/C1 - Head temperature problems
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
Head temperature control was not functioning during the specified period.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • callang_flags
  • cosine_solar_zenith_angle
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • callang_flags
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D020620.42
Start DateStart TimeEnd DateEnd Time
03/24/2002090804/11/20022202
Subject:
SGP/MFR25M/C1 - Loose connections
DataStreams:sgpmfr25mC1.a0, sgpmfr25mC1.b1
Description:
No solar data after 03/24/2002 and no head temp data after 03/25/2002. Connections were 
reseated on 04/02/2002, but the data still looked bad.  More maintenance occurred on 
04/11/2002, when a reseating and tightening of the connections was performed.
Measurements:sgpmfr25mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • lon
  • logger_volt
  • cosine_solar_zenith_angle
  • time_offset
  • base_time
  • lat
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • alt
  • callang_flags
  • head_temp

sgpmfr25mC1.a0:
  • alt
  • base_time
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • lat
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • callang_flags
  • logger_volt
  • time_offset
  • lon


Back To Table of Contents

DQRID : D030916.7
Start DateStart TimeEnd DateEnd Time
04/14/2003000009/16/20032359
Subject:
SGP/MFR10M/C1 - Artifact in data
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
A cable was hanging under the instrument and was in the field of view of the instrument. 
This resulted in some erroneous increases in the data, frequently around 22GMT. These 
increases are likely from specular reflections off the cable.
Measurements:sgpmfr10mC1.b1:
  • logger_volt
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • base_time
  • cosine_solar_zenith_angle
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • time_offset
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • alt
  • callang_flags
  • lon
  • head_temp
  • lat

sgpmfr10mC1.a0:
  • callang_flags
  • time_offset
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • logger_volt
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • lon
  • base_time
  • head_temp
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • lat
  • alt


Back To Table of Contents

DQRID : D050218.3
Start DateStart TimeEnd DateEnd Time
06/27/2003143006/27/20031530
01/20/2005190001/20/20052000
Subject:
SGP/MFR10M/C1 - Data discontinuities
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
There is a discontinuity in the data (all channels) just after 1500 UTC on June 27, 2003 
and again, just before 1930 UTC on Jan 20, 2005.  There is nothing unusual in the temp or 
volt data.

At these times the tower was lowered for cleaning so the initial assumption was the 
instrument was dirty, and a cleaning resulted in the discontinuity.  This theory doesn't hold 
up well though as the tower is lowered for instrument cleaning regularly and these are the 
only two times in the past few years that a discontinuity coinciding with a lowering of 
the tower has resulted.  

Also, down pointing instruments are also not prone to collecting dust and dirt.  This 
instrument is located on the E13 SIRS tower and the down pointing PSP does not show a 
discontinuity at either time, which further discounts the dirty instrument theory.  Site 
operations were asked to check the instrument mount to see if it was loose possibly causing the 
instrument to change its position, and they reported that the mount seemed secure and 
stable.  So as of now, the cause of this issue is a mystery.

Compare with other down pointing instruments to help ascertain if data on one side of the 
discontinuity or the other are "better."  Perform an error analysis to determine if this 
would introduce a significant error in the role the data stream plays in a user's research.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D050218.4
Start DateStart TimeEnd DateEnd Time
09/01/2004000004/30/20050000
Subject:
SGP/MFR10M/C1 - Dip in data at solar noon
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
On clear days there is a dip in the data (all channels) at solar noon that is more 
pronounced during the winter months.  While we currently don't know the cause of this dip, we 
believe it is related to reflections off the tower pole.  The working theory says at solar 
noon the north-south crossbar shades the tower pole cutting off reflections into the 
detector.  If true, this means data on either side of the dip are being enhanced by 
reflections and the data within the dip are actually "correct."  An experiment is being planned 
for next fall that will wrap the upper portion of the tower in a non-reflecting cloth.  
This should determine if the dip is related to reflections from the tower pole.

Since at this point we don't know if the dip consists of bad or good data, users should be 
aware there is a problem and treat all the data as suspect.  It should noted that the 
magnitude of the problem is relatively small, likely no greater than 2.5%, so the dataset 
as whole is probably still useful as is for most users.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D060411.7
Start DateStart TimeEnd DateEnd Time
08/25/1994203009/11/19941631
12/20/1994223012/29/19941248
05/23/1995163005/26/19951327
07/17/1995083001/07/19960533
01/19/1996153003/01/19961731
07/16/1994023007/19/19941831
04/02/1997173004/04/19970019
05/23/1997140805/27/19970104
05/30/1997140806/02/19970314
10/31/1998111311/03/19981935
07/23/2004210207/26/20041458
06/12/2011050006/14/20111959
12/11/2000131312/13/20002115
08/18/2001000208/20/20011602
01/30/2002180202/11/20021502
03/09/2002200203/12/20020702
11/14/2002200211/18/20021602
12/25/1998201312/28/19982014
03/14/1999002203/16/19990135
04/14/1999201305/04/19991914
06/12/1999061606/14/19991414
07/15/1999181307/19/19991813
11/24/2000151311/27/20001814
Subject:
SGP/MFR10M/C1 - Missing data
DataStreams:sgpmfr10mC1.b1, sgpmfr10mC1.a0
Description:
Data are missing and unrecoverable.
Measurements:sgpmfr10mC1.b1:
  • logger_volt
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • base_time
  • cosine_solar_zenith_angle
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • time_offset
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • alt
  • callang_flags
  • lon
  • head_temp
  • lat

sgpmfr10mC1.a0:
  • callang_flags
  • time_offset
  • logger_volt
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • lon
  • base_time
  • head_temp
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • lat
  • alt


Back To Table of Contents

DQRID : D090226.8
Start DateStart TimeEnd DateEnd Time
02/13/2004170202/20/20041525
07/02/2004230207/06/20041502
12/22/2007180812/24/20070211
02/04/2009160902/12/20091709
11/18/1995013011/20/19950232
11/22/1995233111/24/19951715
11/25/1995043111/27/19950931
12/06/1995133112/10/19951653
02/28/1996203003/01/19961531
03/02/1996023203/04/19960450
04/05/1996033204/08/19962010
01/04/1998090901/07/19982115
01/11/1998201101/14/19981712
02/04/1998191202/06/19982212
05/01/1998181105/04/19981611
10/31/1998111411/03/19981313
12/25/1998211512/28/19982015
01/02/1999111401/04/19991506
09/30/1996193210/08/19962033
11/16/1996153211/18/19961533
02/28/1997113103/03/19971933
08/12/1997151008/14/19971932
10/12/1997031010/15/19971416
11/01/1997011111/03/19972212
03/13/1999212303/16/19990325
10/18/2000141411/09/20002316
08/18/2001000208/20/20011602
01/30/2002180202/19/20021702
03/09/2002200203/12/20020702
09/03/1994142009/06/19941223
10/08/1994232010/10/19940225
12/20/1994223112/29/19941634
05/23/1995163105/26/19951634
Subject:
SGP/MFR25M/C1 - Missing data
DataStreams:sgpmfr25mC1.a0, sgpmfr25mC1.b1
Description:
Data are missing and unrecoverable.
Measurements:sgpmfr25mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • lon
  • time
  • logger_volt
  • cosine_solar_zenith_angle
  • time_offset
  • base_time
  • lat
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • alt
  • callang_flags
  • head_temp

sgpmfr25mC1.a0:
  • alt
  • base_time
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • lat
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • callang_flags
  • time
  • logger_volt
  • time_offset
  • lon


Back To Table of Contents

DQRID : D090520.3
Start DateStart TimeEnd DateEnd Time
05/17/2009052805/20/20092300
05/24/2009030006/10/20091600
Subject:
SGP/MFR10M/C1 - Heater failure
DataStreams:sgpmfr10mC1.b1, sgpmfr10mC1.a0
Description:
The heater stopped working on 20090517.  Maintenance on 20090520 seemed to resolve the 
problem, but the heater failed again on 5/24.  Additional maintenance on 20090610 seems to 
have resolved the issue.

The instrument is designed to operate with a head temp of around 40 deg C.  As the head 
was at ambient temp during the period covered by this DQR, the mentor suggests looking at 
the 25m MFR if one is interested in these data.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D091112.3
Start DateStart TimeEnd DateEnd Time
11/05/2009180911/25/20092000
Subject:
SGP/MFR10M/C1 - Bad connection
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
The instrument was pulled for quarterly calibration on 20091105.  When the instrument was 
reinstalled, the head cable was not making a good connection to the head.  Cleaning the 
connector and reinstalling resolved the issue.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D100722.1
Start DateStart TimeEnd DateEnd Time
05/11/2010201405/13/20101505
Subject:
SGP/MFR10M/C1 - Poor connection
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
The instrument was pulled for calibration on 5/11.  Upon re-installation, a good 
connection was not made.  Problem resolved on 5/13.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D120229.2
Start DateStart TimeEnd DateEnd Time
09/09/2010000002/06/20122359
Subject:
SGP/MFR10M/C1 - Noise in ch415, sensor degradation
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
This instrument experienced general degradation over a period of more than a year.  Please 
see D120229.3 (DQPR-2693) and D120229.4 (DQPR-2763) for more information.

The 415nm channel was noisy.  The noise wasn't extreme, so initially at least, the mentor 
would not expect much impact in using the 415 data.  However, toward the end of the 
period covered by this DQR, we start seeing a general degradation of this instrument.  On 
20110720, after a calibration, channel 615 is dead.  On 20111101 the open channel stops 
functioning.  

This instrument is calibrated every three months.  The past couple/few calibrations showed 
a steady decline in the sensitivity of 415.  The declining sensitivity should be 
generally accounted for by each new calibration, but as time moved one, one should become less 
and less confident in those (415) data.

In general, the mentor would suggest using 20110720 as a cut-off date for reliable data 
(all channels).  Caution is given for all channels as the instrument is failing, though the 
channels not specifically mentioned may be OK.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)

sgpmfr10mC1.a0:
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)


Back To Table of Contents

DQRID : D120229.3
Start DateStart TimeEnd DateEnd Time
07/19/2011130402/06/20122359
Subject:
SGP/MFR10M/C1 - Channel 615 dead
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
This instrument experienced general degradation over a period of more than a year.  Please 
see D120229.2 (DQPR-2564) and D120229.4 (DQPR-2763) for more information.

On 20110719 channel 615 stopped functioning.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)

sgpmfr10mC1.a0:
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)


Back To Table of Contents

DQRID : D120229.4
Start DateStart TimeEnd DateEnd Time
11/01/2011200402/06/20122359
Subject:
SGP/MFR10M/C1 - Broadband channel failure
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
This instrument experienced general degradation over a period of more than a year.  Please 
see D120229.2 (DQPR-2564) and D120229.3 (DQPR-2693) for more information.

The open, or broadband, channel died on 20111101.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)

sgpmfr10mC1.a0:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)


Back To Table of Contents

DQRID : D941227.4
Start DateStart TimeEnd DateEnd Time
08/18/1994153008/18/19941630
10/19/1994142010/19/19941500
Subject:
SGP/MFR10M/C1 - Tower lowered for maintenance
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
The 10-m tower at the central facility central cluster was lowered
by site operations personnel for maintenance and repair activities.  
During this time interval, the multifilter radiometer on the tower 
had a field of view that were altered, possibly to include some sky 
radiation.  Data from this sensor should not be included in any 
analyses, for the specified time interval and location.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D950327.1
Start DateStart TimeEnd DateEnd Time
03/08/1995213103/08/19952231
Subject:
SGP/MFR10M/C1 - New datalogger PROM
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
The sgpmfr10m datalogger had its PROM upgraded during a visit by the instrument mentors. 
The new PROM is Version 13 with the capability of operating the MFR heater without a 
shadowband motor attached to the MFR head. No operating changes resulted from this upgrade.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • callang_flags
  • cosine_solar_zenith_angle
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • callang_flags
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D960914.17
Start DateStart TimeEnd DateEnd Time
03/26/1994000003/27/19962359
Subject:
SGP/MFR10m/C1 - Small intermittent time stamp error
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
The internal code that runs the MFR data logger had a bug that occasionally caused the 
time stamp associated with the data to be slightly in error.  This error is thought to be 
very small (say less than 30 seconds) and very infrequent.  (Unfortunately, we cannot give 
firm estimates of the magnitude and frequency of the error).

Also, the upwelling MFR measurements were contaminated by electrical noise originating 
from the heater contained in the sensor head.  During the time the sensor head was polled by 
the data logger, the heater would strobe "on and off", and the noise caused by this 
strobing could greatly contaminate the data.  Some of the data is so contaminated that it is 
virtually useless.   

After 96/03/27 these problems were corrected.  Those who use the data before this time 
should realize that the time stamp for each record of data might be slightly in error, and 
the data could be very noisy due to the heater switching on and off.
Measurements:sgpmfr10mC1.b1:
  • logger_volt
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • base_time
  • cosine_solar_zenith_angle
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • time_offset
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • callang_flags
  • head_temp

sgpmfr10mC1.a0:
  • callang_flags
  • time_offset
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • logger_volt
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • base_time
  • head_temp
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)


Back To Table of Contents

DQRID : D960914.18
Start DateStart TimeEnd DateEnd Time
03/23/1994000003/27/19962359
Subject:
SGP/MFR25M/C1 - Intermittent small time stamp error
DataStreams:sgpmfr25mC1.a0, sgpmfr25mC1.b1
Description:
The internal code that runs the MFR data logger had a bug that occasionally caused the 
time stamp associated with the data to be slightly in error.  This error is thought to be 
very small (say less than 30 seconds) and very infrequent.  (Unfortunately, we cannot give 
firm estimates of the magnitude and frequency of the error).

Also, the upwelling MFR measurements were contaminated by electrical noise originating 
from the heater contained in the sensor head.  During the time the sensor head was polled by 
the data logger, the heater would strobe "on and off", and the noise caused by this 
strobing could greatly contaminate the data.  Some of the data is so contaminated that it is 
virtually useless.   

After 96/03/27 these problems were corrected.  Those who use the data before this time 
should realize that the time stamp for each record of data might be slightly in error, and 
the data could be very noisy due to the heater switching on and off.
Measurements:sgpmfr25mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • logger_volt
  • cosine_solar_zenith_angle
  • time_offset
  • base_time
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • callang_flags
  • head_temp

sgpmfr25mC1.a0:
  • base_time
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • callang_flags
  • logger_volt
  • time_offset


Back To Table of Contents

DQRID : D961008.1
Start DateStart TimeEnd DateEnd Time
08/18/1996000010/03/19962359
Subject:
SGP/MFR10M/C1 - Corrupt or truncated data
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
On about 96/08/18, the data logger board for the downlooking 10m MFR was modified to test 
an upgrade.  Unfortunately this upgrade did not work as planned, and some data was either 
corrupted or truncated.  You can't miss the corrupted data (from which you can infer 
albedos much greater than one!).  The truncated data is truncated in time, by which we mean 
that only a half day's worth of data was collected.  The daily time span of these 
truncated data is from midnight to about 1 PM (local time). At about 96/10/03, these problems 
were corrected.

Note:  The "begin date" given for this DQR may be off a day or two, so beware.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • cosine_solar_zenith_angle
  • callang_flags
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • callang_flags
  • head_temp
  • logger_volt
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D961010.1
Start DateStart TimeEnd DateEnd Time
08/22/1996000010/08/19961900
Subject:
SGP/MFR25M/C1 - poor data quality
DataStreams:sgpmfr25mC1.a0, sgpmfr25mC1.b1
Description:
In the middle of August, the logger board that collects data from the 25m MFR was 
upgraded.  This upgrade did not work out as planned, and the data quality from the MFR are not 
good.  The poor data is easily identifiable - you'll know it when you see it.  After 
19961008, a new logger board was installed that solved the problem. 

Note: the exact date when these problems started is not known with precision; it could be 
several days prior to or after 960822.
Measurements:sgpmfr25mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • cosine_solar_zenith_angle
  • callang_flags

sgpmfr25mC1.a0:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • callang_flags
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)


Back To Table of Contents

DQRID : D990906.1
Start DateStart TimeEnd DateEnd Time
04/15/1999000009/02/19992221
Subject:
SGP/MFR10M/C1 - Head heater not working
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
On 19990415, the tower upon which the 10-m MFR was mounted was dismantled and replaced 
with a so-called "tippy tower".  After the 10-m MFR was reinstalled on the new tower, the 
mfr head heater did not work, or worked intermittently (at best).  A new mfr head was 
installed on 99.09.02 and that solved the problem.  

Data was collected from the old head in the morning up to the time it was replaced.  Once 
the new head was installed, data collection resumed.  There's a gap of about an hour 
between the time the old head was removed and the new head was installed.  If we try to gauge 
how well the old head was calibrated by comparing the two segments of data (old head, 
new head) it appears as if the old head remained in a good state of calibration even though 
the head heater was not working.
Measurements:sgpmfr10mC1.b1:
  • head_temp

sgpmfr10mC1.a0:
  • head_temp


Back To Table of Contents



END OF DATA