Data Quality Reports for Session: 153293 User: aeround Completed: 09/30/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D100616.2NSA/VCEIL/C1 - Small gaps in datansavceil25kC1.b1
D100719.1NSA/VCEIL/C1 - Gaps in datansavceil25kC1.b1
D130502.3NSA/VCEIL/C1 - Reprocess: Time errornsavceil25kC1.a1, nsavceil25kC1.b1


DQRID : D100616.2
Start DateStart TimeEnd DateEnd Time
12/03/2009210706/15/20102143
Subject:
NSA/VCEIL/C1 - Small gaps in data
DataStreams:nsavceil25kC1.b1
Description:
When the ceilometer was replaced, small gaps in the netCDF files occurred due to the 
presence of '?' characters in the backscatter and header records of the raw data that could 
not be ingested. It is unknown whether this was caused by a poor CPU board or 
incompatibilities with the computer configuration (hyper-threading). The problem was corrected when 
the instrument and computer were both replaced. 

During this period, approximately 3% of data are missing each hour due to this problem.
Measurements:nsavceil25kC1.b1:
  • Third cloud base height(third_cbh)
  • alt_highest_signal
  • laser_temperature
  • receiver_sensitivity
  • Second lowest cloud base height(second_cbh)
  • Lowest cloud base height detected.(first_cbh)
  • vertical_visibility
  • tilt_angle
  • time_offset
  • measurement_parameters
  • status_flag
  • sum_backscatter
  • range
  • base_time
  • Normalized corrected relative backscatter(backscatter)
  • status_string
  • background_light
  • window_contamination
  • detection_status
  • laser_pulse_energy


Back To Table of Contents

DQRID : D100719.1
Start DateStart TimeEnd DateEnd Time
06/15/2010215906/17/20102201
Subject:
NSA/VCEIL/C1 - Gaps in data
DataStreams:nsavceil25kC1.b1
Description:
When the CT25K ceilometer was replaced with a new CL31, checksum errors began occurring at 
regular 6-min intervals and several things were done to diagnose the problem, including 
changing the serial cable and trying different computer types. It was finally corrected 
by replacing the computer with one in which hyperthreading was disabled by default in the 
BIOS.
Measurements:nsavceil25kC1.b1:
  • Third cloud base height(third_cbh)
  • alt_highest_signal
  • laser_temperature
  • receiver_sensitivity
  • Second lowest cloud base height(second_cbh)
  • Lowest cloud base height detected.(first_cbh)
  • vertical_visibility
  • tilt_angle
  • time_offset
  • measurement_parameters
  • status_flag
  • sum_backscatter
  • range
  • base_time
  • Normalized corrected relative backscatter(backscatter)
  • status_string
  • background_light
  • window_contamination
  • detection_status
  • laser_pulse_energy


Back To Table of Contents

DQRID : D130502.3
Start DateStart TimeEnd DateEnd Time
03/31/2013010004/15/20131540
Subject:
NSA/VCEIL/C1 - Reprocess: Time error
DataStreams:nsavceil25kC1.a1, nsavceil25kC1.b1
Description:
The computer time jumped forward an hour because the time zone was set to "(GMT) Greenwich 
Mean Time" with "Automatically adjust clock for Daylight Savings Time" selected. The 
time zone was changed to "(GMT) Coordinated Universal Time" and the clock reset to correct 
UTC time.

Correct time by subtracting 1 hour.
Measurements:nsavceil25kC1.b1:
  • time_offset

nsavceil25kC1.a1:
  • time_offset


Back To Table of Contents



END OF DATA