DQR IDSubjectData Streams Affected
D050331.3PYE/MPL/M1 - wrong temperature calibrationpyemplM1.a1
D050803.4PYE/NFOV/M1 - Intermittent missing datapyenfov2chM1.00, pyenfov2chM1.b1
D050803.5PYE/NFOV/M1 - Intermittent data gapspyenfov2chM1.00, pyenfov2chM1.b1
D050902.2PYE/MWRP/M1 - Bias in Ground Relative Humidity MeasurementpyemwrpM1.00, pyemwrpM1.b1
D060317.1PYE/MPL/M1 - Erroneous temperaturespyemplM1.a1
D101020.6PYE/AOS/M1 - Data processing problempyeaosM1.a1, pyenoaaaosM1.b1, pyenoaaaosavgM1.b1


DQRID : D050331.3
Start DateStart TimeEnd DateEnd Time
02/22/2005000005/05/20052359
Subject:
PYE/MPL/M1 - wrong temperature calibration
DataStreams:pyemplM1.a1
Description:
Wrong calibration coefficients were applied to the temperature values.  These are 
engineering housekeeping data only.  The scientifically relevant measurements were unaffected.
Measurements:pyemplM1.a1:
  • laser_temp
  • detector_temp
  • filter_temp
  • instrument_temp


Back To Table of Contents

DQRID : D050803.4
Start DateStart TimeEnd DateEnd Time
07/18/2005190009/10/20050055
Subject:
PYE/NFOV/M1 - Intermittent missing data
DataStreams:pyenfov2chM1.00, pyenfov2chM1.b1
Description:
During the Pt Reyes deployment of the NFOV, the instrument experienced intermittent data 
gaps due to communications problems.  Data gaps occurred for the following date ranges:
20050718-20050723
20050727-20050730
20050803-20050804
20050813-20050816
20050820-20050822
20050826-20050827
20050830-20050902
20050908-20050910
Measurements:pyenfov2chM1.b1:
  • time_offset
  • Calibrated Zenith Radiance, 673nm(radiance_673nm)
  • head_temp
  • time
  • base_time
  • tube_temp
  • raw_counts_870nm
  • cosine_solar_zenith_angle
  • raw_counts_673nm
  • Calibrated Zenith Radiance, 870nm(radiance_870nm)

pyenfov2chM1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D050803.5
Start DateStart TimeEnd DateEnd Time
07/18/2005190007/30/20051200
Subject:
PYE/NFOV/M1 - Intermittent data gaps
DataStreams:pyenfov2chM1.00, pyenfov2chM1.b1
Description:
Intermittent data collection of the NFOV.  Data collection not successful regularly during 
day light hours but generally caught up during the nighttime hours.  Occasionally the 
system stopped collecting altogether and a reboot of the network components was required to 
restart collection. Data gaps are largest 18 through 23 of July and 27 through 30 July.
Measurements:pyenfov2chM1.b1:
  • Calibrated Zenith Radiance, 673nm(radiance_673nm)
  • time_offset
  • base_time
  • tube_temp
  • cosine_solar_zenith_angle
  • raw_counts_870nm
  • lat
  • Calibrated Zenith Radiance, 870nm(radiance_870nm)
  • alt
  • head_temp
  • time
  • lon
  • raw_counts_673nm

pyenfov2chM1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D050902.2
Start DateStart TimeEnd DateEnd Time
06/17/2005000009/03/20050000
Subject:
PYE/MWRP/M1 - Bias in Ground Relative Humidity Measurement
DataStreams:pyemwrpM1.00, pyemwrpM1.b1
Description:
On June 17 2005, the MWRP was re-installed at PYE after the Super Blower was replaced by 
the old style dew blower. The relative humidity ground sensor in the dew blower is not as 
accurate. This resulted in a negative relative humidity bias of about 5.9 %. In other 
words the MWRP sensor was underestimating the relative humidity. The bias was determined by 
comparing MWRP measurements with SMET and radiosonde measurements. The MP.cfg file for 
the MWRP was changed to reflect this bias on 09/02/2005 and will take effect on 09/03/2005 
0000 GMT
Measurements:pyemwrpM1.00:
  • AVHRR satellite images for Darwin from BOM (RAW)(raw)

pyemwrpM1.b1:
  • Density, water vapor, at instrument height (1 or 2 m)(surfaceWaterVaporDensity)
  • Humidity, relative, at instrument height (1 or 2 m)(surfaceRelativeHumidity)


Back To Table of Contents

DQRID : D060317.1
Start DateStart TimeEnd DateEnd Time
02/22/2005000003/15/20052359
Subject:
PYE/MPL/M1 - Erroneous temperatures
DataStreams:pyemplM1.a1
Description:
The temperature readings for the laser, detector, and instrument are erroneous.  Incorrect 
calibration factors; data should not be used.  This time period is prior to the official 
start of the PYE deployment.
Measurements:pyemplM1.a1:
  • laser_temp
  • detector_temp
  • instrument_temp


Back To Table of Contents

DQRID : D101020.6
Start DateStart TimeEnd DateEnd Time
03/09/2005000009/15/20050000
Subject:
PYE/AOS/M1 - Data processing problem
DataStreams:pyeaosM1.a1, pyenoaaaosM1.b1, pyenoaaaosavgM1.b1
Description:
The AOS data was reprocessed for the AMF, SGP and NSA sites. There was an error in the 
processing of the corrections for the PSAP absorption signal such that a low pass filter was 
implemented for the scattering data used in these corrections.The difference between the 
corrected and previous data is < 0.05 Mm-1 (below the instrument detection limit) and 
should not have a significant effect on the previous data. Data will be reprocessed at a 
later time when more significant changes to the data processing are implemented.
Measurements:pyeaosM1.a1:
  • Absorption coefficient, blue wavelength, 3 wavelength PSAP, low RH, 10 um size
    cut(Ba_B_Dry_10um_PSAP3W_1)
  • Absorption coefficient, blue wavelength, 3 wavelength PSAP, low RH, 1 um size
    cut(Ba_B_Dry_1um_PSAP3W_1)
  • Absorption coefficient, red wavelength, 3 wavelength PSAP, low RH, 10 um size
    cut(Ba_R_Dry_10um_PSAP3W_1)
  • Absorption coefficient, green wavelength, 3 wavelength PSAP, low RH, 10 um size
    cut(Ba_G_Dry_10um_PSAP3W_1)
  • Aerosol light absorption coefficient, green channel, 1 um particle diameter(Ba_G_Dry_1um_PSAP1W_1)
  • Aerosol light absorption coefficient, green channel, 10 um particle diameter(Ba_G_Dry_10um_PSAP1W_1)
  • Absorption coefficient, green wavelength, 3 wavelength PSAP, low RH, 1 um size
    cut(Ba_G_Dry_1um_PSAP3W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_1um_PSAP3W_1)

pyenoaaaosavgM1.b1:
  • Aerosol light absorption coefficient, green channel, 1 um particle diameter(Ba_G_Dry_1um_PSAP1W_1)
  • Aerosol light absorption coefficient, green channel, 10 um particle diameter(Ba_G_Dry_10um_PSAP1W_1)

pyenoaaaosM1.b1:
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_1um_PSAP3W_1)
  • Aerosol light absorption coefficient, green channel, 10 um particle diameter(Ba_G_Dry_10um_PSAP1W_1)
  • Absorption coefficient, blue wavelength, 3 wavelength PSAP, low RH, 10 um size
    cut(Ba_B_Dry_10um_PSAP3W_1)
  • Absorption coefficient, red wavelength, 3 wavelength PSAP, low RH, 10 um size
    cut(Ba_R_Dry_10um_PSAP3W_1)
  • Absorption coefficient, green wavelength, 3 wavelength PSAP, low RH, 1 um size
    cut(Ba_G_Dry_1um_PSAP3W_1)
  • Absorption coefficient, green wavelength, 3 wavelength PSAP, low RH, 10 um size
    cut(Ba_G_Dry_10um_PSAP3W_1)
  • Aerosol light absorption coefficient, green channel, 1 um particle diameter(Ba_G_Dry_1um_PSAP1W_1)
  • Absorption coefficient, blue wavelength, 3 wavelength PSAP, low RH, 1 um size
    cut(Ba_B_Dry_1um_PSAP3W_1)


Back To Table of Contents