Data Quality Reports for Session: 152707 User: hs32 Completed: 08/31/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D060113.1TWP/VCEIL/C3 - low receiver sensitivitytwpvceil25kC3.b1
D060717.5TWP/VCEIL/C2 - Time offsettwpvceil25kC2.b1
D061220.34TWP/VCEIL/C3 - standard instrument installedtwpvceil25kC3.b1
D070201.1TWP/VCEIL/C3 - Reduced sensitivity to high cloudstwpvceil25kC3.b1
D070201.2TWP/VCEIL/C2 - optical cross-talk compensationtwpvceil25kC2.b1
D100215.4TWP/VCEIL/C1 - Reprocess: Incorrect serial_number and sum_backscatter fieldstwpvceil25kC1.b1
D100215.5TWP/VCEIL/C2 - Reprocess: Incorrect serial_number and sum_backscatter fieldstwpvceil25kC2.b1
D100215.6TWP/VCEIL/C3 - Reprocess: Incorrect serial_number and sum_backscatter fieldstwpvceil25kC3.b1


DQRID : D060113.1
Start DateStart TimeEnd DateEnd Time
12/12/2005055505/18/20062300
Subject:
TWP/VCEIL/C3 - low receiver sensitivity
DataStreams:twpvceil25kC3.b1
Description:
The ceilometer reported frequent receiver alarms. There are periods when the receiver 
sensitivity dropped to about 50%. The problem was corrected by replacing the instrument with 
a spare.
Measurements:twpvceil25kC3.b1:
  • receiver_sensitivity
  • Normalized corrected relative backscatter(backscatter)
  • sum_backscatter
  • Second lowest cloud base height(second_cbh)
  • Third cloud base height(third_cbh)
  • Lowest cloud base height detected.(first_cbh)
  • background_light
  • alt_highest_signal
  • vertical_visibility


Back To Table of Contents

DQRID : D060717.5
Start DateStart TimeEnd DateEnd Time
05/13/2006222407/17/20061912
Subject:
TWP/VCEIL/C2 - Time offset
DataStreams:twpvceil25kC2.b1
Description:
On 7/14/2006, I found that the ceilometer time was 8m 36s slow, the PC time display (lower 
right taskbar) was frozen at "22:24", and AboutTime was not running. Corrected it on 
7/17 by running AboutTime so that 19:03:29 became 19:12:05. This also corrected the frozen 
display on the Windows98 taskbar. 

I am not sure of the start time of this problem but there were several other Windows 
related issues that had to be corrected on 5/14 with indications of inappropriate access via 
Remote Administrator on 5/13. I do not know if the time offset was a sudden or gradual 
change.
Measurements:twpvceil25kC2.b1:
  • status_string
  • detection_status
  • sum_backscatter
  • laser_pulse_energy
  • tilt_angle
  • alt_highest_signal
  • laser_temperature
  • vertical_visibility
  • range
  • Third cloud base height(third_cbh)
  • measurement_parameters
  • Second lowest cloud base height(second_cbh)
  • status_flag
  • Lowest cloud base height detected.(first_cbh)
  • background_light
  • Normalized corrected relative backscatter(backscatter)
  • window_contamination
  • base_time
  • receiver_sensitivity
  • time_offset


Back To Table of Contents

DQRID : D061220.34
Start DateStart TimeEnd DateEnd Time
05/18/2006230012/20/20060400
Subject:
TWP/VCEIL/C3 - standard instrument installed
DataStreams:twpvceil25kC3.b1
Description:
A spare ceilometer without the tropics window option was installed. The manufacturer 
recommends the tropics window be used at +/- 25 deg lattitude to prevent direct solar 
radiation into the laser optics when the instrument pointed at zenith. No significant signal 
degradation is apparent during the autumnal equinox but, as a precaution, the unit was 
replaced with the original with a tropics window.
Measurements:twpvceil25kC3.b1:
  • Normalized corrected relative backscatter(backscatter)
  • laser_pulse_energy
  • laser_temperature
  • sum_backscatter
  • background_light
  • vertical_visibility


Back To Table of Contents

DQRID : D070201.1
Start DateStart TimeEnd DateEnd Time
12/20/2006040101/12/20070120
Subject:
TWP/VCEIL/C3 - Reduced sensitivity to high clouds
DataStreams:twpvceil25kC3.b1
Description:
The instrument produced a significant number of optical cross-talk warnings and displayed 
a reduced capacity to detect high clouds.  The internal cables were inspected, the 
compensation fiber and optics monitor were replaced, the crosstalk compensation was adjusted, 
and the system tested with an optical termination hood, all with no improvement.  While 
waiting for a new receiver board, receiver ribbon cable, and compensation fiber cable from 
Vaisala, a spare ceilometer with a standard, "non-tropics", window was installed.
Measurements:twpvceil25kC3.b1:
  • receiver_sensitivity
  • Second lowest cloud base height(second_cbh)
  • Third cloud base height(third_cbh)
  • detection_status
  • alt_highest_signal


Back To Table of Contents

DQRID : D070201.2
Start DateStart TimeEnd DateEnd Time
12/13/2006000002/05/20070028
Subject:
TWP/VCEIL/C2 - optical cross-talk compensation
DataStreams:twpvceil25kC2.b1
Description:
The ceilometer gave frequent optical cross-talk compensation warnings, was insensitive to 
high clouds, and the tropics window was deteriorated. The instrument was replaced with a 
spare.
Measurements:twpvceil25kC2.b1:
  • Second lowest cloud base height(second_cbh)
  • alt_highest_signal
  • Normalized corrected relative backscatter(backscatter)
  • window_contamination
  • vertical_visibility
  • Third cloud base height(third_cbh)


Back To Table of Contents

DQRID : D100215.4
Start DateStart TimeEnd DateEnd Time
10/11/1996000010/31/20092359
Subject:
TWP/VCEIL/C1 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:twpvceil25kC1.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
   
This DQR also applies to a1 level data.
Measurements:twpvceil25kC1.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100215.5
Start DateStart TimeEnd DateEnd Time
10/28/1998000010/31/20092359
Subject:
TWP/VCEIL/C2 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:twpvceil25kC2.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
   
This DQR also applies to a1 level data.
Measurements:twpvceil25kC2.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100215.6
Start DateStart TimeEnd DateEnd Time
02/27/2002000010/31/20092359
Subject:
TWP/VCEIL/C3 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:twpvceil25kC3.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
   
This DQR also applies to a1 level data.