Data Quality Reports for Session: 151183 User: paulc Completed: 06/27/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D110418.6SGP/MWRHF/C1 - Reprocessed: Rapid drift in 90 GHz noise diodesgpmwrhfC1.b1


DQRID : D110418.6
Start DateStart TimeEnd DateEnd Time
03/18/2011000009/20/20112359
Subject:
SGP/MWRHF/C1 - Reprocessed: Rapid drift in 90 GHz noise diode
DataStreams:sgpmwrhfC1.b1
Description:
The noise diode of the 90 GHz channel was showing rapid drifts that affected the accuracy 
of the calibration. Although the algorithm is designed to follow slow drifts in the noise 
diode, in this case (because the variation was fast) the computed brightness temperature 
was probably about 4-5 K colder than it should be.

Ingest was modified and data were reprocessed and archived in September 2011
Measurements:sgpmwrhfC1.b1:
  • Sky brightness temperature (90 GHz)(tbsky90)


Back To Table of Contents



END OF DATA