Data Quality Reports for Session: 151125 User: adam1schwa Completed: 06/24/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D041117.2SGP/MWR/C1 - Reprocess: wrong retrievalssgpmwrlosC1.a1, sgpmwrlosC1.b1, sgpmwrtipC1.a1, sgpqmemwrcolC1.c1
D050722.1SGP/MWR/C1 - REPROCESS - Revised Retrieval Coefficientssgp1mwravgC1.c1, sgp5mwravgC1.c1, sgpmwrlosC1.a1, sgpmwrlosC1.b1, sgpmwrtipC1.a1,
sgpqmemwrcolC1.c1
D051011.6SGP/MWR/C1 - New software version (4.15) installedsgpmwrlosC1.b1, sgpmwrtipC1.a1


DQRID : D041117.2
Start DateStart TimeEnd DateEnd Time
09/21/2004164311/11/20042100
Subject:
SGP/MWR/C1 - Reprocess: wrong retrievals
DataStreams:sgpmwrlosC1.a1, sgpmwrlosC1.b1, sgpmwrtipC1.a1, sgpqmemwrcolC1.c1
Description:
When the computer and core configuration were upgraded, retrieval coefficients for BF1 
were accidently included in the configuration file.
The correct coefficients for CF1 were applied when the configuration file was updated.
Measurements:sgpmwrtipC1.a1:
  • liqtip
  • vaptip

sgpqmemwrcolC1.c1:
  • mean_liq_mwr
  • mean_vap_mwr

sgpmwrlosC1.a1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)

sgpmwrlosC1.b1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)


Back To Table of Contents

DQRID : D050722.1
Start DateStart TimeEnd DateEnd Time
04/16/2002200006/28/20052300
Subject:
SGP/MWR/C1 - REPROCESS - Revised Retrieval Coefficients
DataStreams:sgp1mwravgC1.c1, sgp5mwravgC1.c1, sgpmwrlosC1.a1, sgpmwrlosC1.b1, sgpmwrtipC1.a1,
sgpqmemwrcolC1.c1
Description:
IN THE BEGINNING (June 1992), the retrieval coefficients used to derive the precipitable 
water vapor (PWV) and liquid water path (LWP) from the MWR brightness temperatures were 
based on the Liebe and Layton (1987) water vapor and oxygen absorption model and the Grant 
(1957) liquid water absorption model.  

Following the SHEBA experience, revised retrievals based on the more recent Rosenkranz 
(1998) water vapor and oxygen absorption models and the Liebe (1991) liquid waer absorption 
model were developed.  The Rosenkranz water vapor absorption model resulted a 2 percent 
increase in PWV relative to the earlier Liebe and Layton model.  The Liebe liquid water 
absorption model decreased the LWP by 10% relative to the Grant model.  However, the 
increased oxygen absorption caused a 0.02-0.03 mm (20-30 g/m2) reduction in LWP, which was 
particularly significant for low LWP conditions (i.e. thin clouds encountered at SHEBA).

Recently, it has been shown (Liljegren, Boukabara, Cady-Pereira, and Clough, TGARS v. 43, 
pp 1102-1108, 2005) that the half-width of the 22 GHz water vapor line from the HITRAN 
compilation, which is 5 percent smaller than the Liebe and Dillon (1969) half-width used in 
Rosenkranz (1998), provided a better fit to the microwave brightness temperature 
measurements at 5 frequencies in the range 22-30 GHz, and yielded more accurate retrievals.  
Accordingly, revised MWR retrieval coefficients have been developed using MONORTM, which 
utilizes the HITRAN compilation for its spectroscopic parameters.  These new retrievals 
provide 3 percent less PWV and 2.6 percent greater LWP than the previous retrievals based on 
Rosenkranz (1998).

Although the MWR data will be reprocessed to apply the new monortm-based retrievals, for 
most purposes it will be sufficient to correct the data using the following factors:

PWV_MONORTM = 0.9695 * PWV_ROSENKRANZ
LWP_MONORTM = 1.026  * LWP_ROSENKRANZ

The Rosenkranz-based retrieval coefficients became active as follows (BCR 456):
SGP/C1 (Lamont)     4/16/2002, 2000
SGP/B1 (Hillsboro)  4/12/2002, 1600
SGP/B4 (Vici)       4/15/2002, 2300
SGP/B5 (Morris)     4/15/2002, 2300
SGP/B6 (Purcell)    4/16/2002, 2200
SGP/E14(Lamont)     4/16/2002, 0000
NSA/C1 (Barrow)     4/25/2002, 1900 
NSA/C2 (Atqasuk)    4/18/2002, 1700
TWP/C1 (Manus)      5/04/2002, 0200
TWP/C2 (Nauru)      4/27/2002, 0600
TWP/C3 (Darwin)     inception

The MONORTM-based retrieval coefficients became active as follows (BCR 984):

SGP/C1 (Lamont)     6/28/2005, 2300
SGP/B1 (Hillsboro)  6/24/2005, 2100
SGP/B4 (Vici)       6/24/2005, 2100
SGP/B5 (Morris)     6/24/2005, 2100
SGP/B6 (Purcell)    6/24/2005, 1942
SGP/E14(Lamont)     6/28/2005, 2300
NSA/C1 (Barrow)     6/29/2005, 0000 
NSA/C2 (Atqasuk)    6/29/2005, 0000
TWP/C1 (Manus)      6/30/2005, 2100
TWP/C2 (Nauru)      6/30/2005, 2100
TWP/C3 (Darwin)     6/30/2005, 2100
PYE/M1 (Pt. Reyes)  4/08/2005, 1900**

** At Pt. Reyes, the original retrieval coefficients implemented in March 2005 were based 
on a version of the Rosenkranz model that had been modified to use the HITRAN half-width 
at 22 GHz and to be consistent with the water vapor continuum in MONORTM.  These 
retrievals yield nearly identical results to the MONORTM retrievals.  Therefore the Pt. Reyes 
data prior to 4/08/2005 may not require reprocessing.
Measurements:sgp5mwravgC1.c1:
  • Total liquid water along LOS path(liq)
  • Total water vapor along LOS path(vap)

sgpmwrtipC1.a1:
  • liqtip
  • vaptip

sgpqmemwrcolC1.c1:
  • mean_liq_mwr
  • mean_vap_mwr

sgp1mwravgC1.c1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)

sgpmwrlosC1.a1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)

sgpmwrlosC1.b1:
  • Total water vapor along LOS path(vap)
  • Total liquid water along LOS path(liq)


Back To Table of Contents

DQRID : D051011.6
Start DateStart TimeEnd DateEnd Time
07/31/2002202708/04/20051959
Subject:
SGP/MWR/C1 - New software version (4.15) installed
DataStreams:sgpmwrlosC1.b1, sgpmwrtipC1.a1
Description:
A problem began with the installation of MWR.EXE version 4.12 in July 2002. The software 
had been upgraded from a "DOS" to a "Windows"-compiled program to address an earlier 
problem.  The software upgrade corrected the earlier problem but introduced a new one that 
caused line-of-sight observing cycles to be skipped, a 15% reduction in the number of tip 
curves, and saturation of CPU usage. Software versions 4.13 and 4.14 also produced these 
problems.

The new MWR software, version 4.15, was installed on 08/04/2005. As a consequence of this 
upgrade, the tip curve frequency increased. The tip cycle time decreased from ~60s to 
~50s.
Measurements:sgpmwrtipC1.a1:
  • tc23
  • liqtip
  • tnd23I
  • tbsky23tip
  • tipsky23
  • tipsky31
  • tkxc
  • tknd
  • tkbb
  • tnd31I
  • r23
  • bbn31
  • tnd_nom23
  • bb23
  • tkair
  • tnd31
  • tc31
  • vaptip
  • tbsky31tip
  • bbn23
  • r31
  • tnd_nom31
  • bb31
  • tnd23

sgpmwrlosC1.b1:
  • bbn23
  • tkxc
  • tnd_nom31
  • tc31
  • Radiation, longwave, brightness temperature, 23.8 GHz(tbsky23)
  • Total liquid water along LOS path(liq)
  • tkbb
  • sky31
  • tc23
  • bb31
  • bbn31
  • Total water vapor along LOS path(vap)
  • tkair
  • bb23
  • tnd_nom23
  • sky23
  • 31.4 GHz sky brightness temperature(tbsky31)
  • tnd23
  • IR Brightness Temperature(sky_ir_temp)
  • tnd31
  • tknd


Back To Table of Contents



END OF DATA