Data Quality Reports for Session: 151086 User: james1blee Completed: 06/26/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040225.10SGP/TWR/C1 - Temperature Sensor Not Connected to Dataloggersgp1twr10xC1.b1, sgp30twr10xC1.b1
D040304.30SGP/TWR/C1 - Southeast Side RH Too Largesgp1twr10xC1.b1, sgp30twr10xC1.b1
D040830.4SGP/TWR/C1 - Tower SE-side lowered for maintenancesgp1twr10xC1.b1, sgp30twr10xC1.b1
D071217.2SGP/TWR/C1 - Southeast Side (10X) Temperature and RH Spikessgp1twr10xC1.b1, sgp30twr10xC1.b1


DQRID : D040225.10
Start DateStart TimeEnd DateEnd Time
02/05/2004163102/09/20041926
Subject:
SGP/TWR/C1 - Temperature Sensor Not Connected to Datalogger
DataStreams:sgp1twr10xC1.b1, sgp30twr10xC1.b1
Description:
One of the southeast side 25 m Temperature sensor leads was pulled out from the CR10X 
datalogger terminal strip.
Measurements:sgp1twr10xC1.b1:
  • Temperature, air, 25-m height, 30-min intervals(temp_25m)
  • Pressure, vapor, 25-m height, 30-min intervals(vap_pres_25m)

sgp30twr10xC1.b1:
  • Temperature, air, 25-m height, 30-min intervals(temp_25m)
  • Pressure, vapor, 25-m height, 30-min intervals(vap_pres_25m)


Back To Table of Contents

DQRID : D040304.30
Start DateStart TimeEnd DateEnd Time
08/15/2003134502/10/20042310
Subject:
SGP/TWR/C1 - Southeast Side RH Too Large
DataStreams:sgp1twr10xC1.b1, sgp30twr10xC1.b1
Description:
The T/RH probes installed on the 25 m and 60 m levels of the southeast side of the tower 
on 8/15/03 gave RH readings that were too large for RH above 65%. The difference between 
the west and southeast side RH was within the combined accuracies of the sensors up to 
about 80% RH.  Above 80% RH the southeast side RH was much larger, with the difference 
increasing to 10% larger (or more) at 100% RH.  The vapor pressure from the southeast side 
(calculated with RH) was also too large.

The probes were replaced on 2/10/04 and the two sides of the tower now agree
fairly well, with RH not exceeding 100% during saturation.
Measurements:sgp1twr10xC1.b1:
  • Pressure, vapor, 60-m height, 30-min intervals(vap_pres_60m)
  • Humidity, relative, 25-m height, 30-min intervals(rh_25m)
  • Humidity, relative, 60-m height, 30-min intervals(rh_60m)
  • Pressure, vapor, 25-m height, 30-min intervals(vap_pres_25m)

sgp30twr10xC1.b1:
  • Humidity, relative, 25-m height, 30-min intervals(rh_25m)
  • Pressure, vapor, 60-m height, 30-min intervals(vap_pres_60m)
  • Humidity, relative, 60-m height, 30-min intervals(rh_60m)
  • Pressure, vapor, 25-m height, 30-min intervals(vap_pres_25m)


Back To Table of Contents

DQRID : D040830.4
Start DateStart TimeEnd DateEnd Time
11/19/2003210411/19/20032118
12/08/2003194112/09/20031422
02/10/2004185602/10/20042308
02/13/2004175102/13/20042138
02/20/2004145802/23/20041840
08/05/2004141208/05/20041451
09/27/2004143809/27/20042056
02/14/2006200202/14/20062105
05/31/2006175605/31/20061852
10/11/2006144610/11/20062039
07/26/2007134007/26/20071900
08/20/2007145008/20/20071555
Subject:
SGP/TWR/C1 - Tower SE-side lowered for maintenance
DataStreams:sgp1twr10xC1.b1, sgp30twr10xC1.b1
Description:
The southeast side tower carriage was lowered for maintenance and/or data are off-scale 
during maintenance.
Measurements:sgp1twr10xC1.b1:
  • Pressure, vapor, 60-m height, 30-min intervals(vap_pres_60m)
  • Humidity, relative, 25-m height, 30-min intervals(rh_25m)
  • Pressure, vapor, 25-m height, 30-min intervals(vap_pres_25m)
  • Humidity, relative, 60-m height, 30-min intervals(rh_60m)
  • Temperature, air, 25-m height, 30-min intervals(temp_25m)
  • Temperature, air, 60-m height, 30-min intervals(temp_60m)

sgp30twr10xC1.b1:
  • sd_vap_pres_25m
  • Pressure, vapor, 60-m height, 30-min intervals(vap_pres_60m)
  • Temperature, air, 60-m height, 30-min intervals(temp_60m)
  • sd_rh_25m
  • Pressure, vapor, 25-m height, 30-min intervals(vap_pres_25m)
  • sd_temp_25m
  • sd_rh_60m
  • sd_temp_60m
  • sd_vap_pres_60m
  • Humidity, relative, 25-m height, 30-min intervals(rh_25m)
  • Temperature, air, 25-m height, 30-min intervals(temp_25m)
  • Humidity, relative, 60-m height, 30-min intervals(rh_60m)


Back To Table of Contents

DQRID : D071217.2
Start DateStart TimeEnd DateEnd Time
12/04/2007170001/24/20082030
Subject:
SGP/TWR/C1 - Southeast Side (10X) Temperature and RH Spikes
DataStreams:sgp1twr10xC1.b1, sgp30twr10xC1.b1
Description:
Southeast side 25 m and 60 m (10X datastreams) carriage connectors may not have been well 
connected; temperatures and relative humidities spiked sometimes.  This problem occurred 
after the carriages were lowered on 12/04.
Measurements:sgp1twr10xC1.b1:
  • Pressure, vapor, 60-m height, 30-min intervals(vap_pres_60m)
  • Humidity, relative, 25-m height, 30-min intervals(rh_25m)
  • Pressure, vapor, 25-m height, 30-min intervals(vap_pres_25m)
  • Humidity, relative, 60-m height, 30-min intervals(rh_60m)
  • Temperature, air, 25-m height, 30-min intervals(temp_25m)
  • Temperature, air, 60-m height, 30-min intervals(temp_60m)

sgp30twr10xC1.b1:
  • sd_vap_pres_25m
  • Pressure, vapor, 60-m height, 30-min intervals(vap_pres_60m)
  • Temperature, air, 60-m height, 30-min intervals(temp_60m)
  • sd_rh_25m
  • Pressure, vapor, 25-m height, 30-min intervals(vap_pres_25m)
  • sd_temp_25m
  • sd_rh_60m
  • sd_temp_60m
  • sd_vap_pres_60m
  • Humidity, relative, 25-m height, 30-min intervals(rh_25m)
  • Temperature, air, 25-m height, 30-min intervals(temp_25m)
  • Humidity, relative, 60-m height, 30-min intervals(rh_60m)


Back To Table of Contents



END OF DATA