DQRID : D050120.2
Start DateStart TimeEnd DateEnd Time
01/01/2004000005/01/20052359
Subject:
SGP/AOS/C1 - Failed RH sensor
DataStreams:sgpaosC1.a0
Description:
The RH sensor inside the humidified nephelometer began to slowly decline in value in the 
upper RH range beginning in early 2004 and becoming more noticable poor on 20041117. The 
sensor currently reaches a high RH value of about 60% when the actual RH is closer to 80%.

The sensor was replaced in the Spring of 2005, but the problem remained.  Effective 
1/1/2004, this sensor is not used to calculate scientific data and is therefore essentially 
abandoned in the field.  An end-date of 20050501 is being used as a placeholder.
Measurements:sgpaosC1.a0:
  • TSINephRHSamp_HRH


Back To Table of Contents

DQRID : D050401.1
Start DateStart TimeEnd DateEnd Time
03/04/2005000003/22/20051955
Subject:
SGP/AOS/C1 - wet nephelometer out of service
DataStreams:sgpaosC1.a0
Description:
The humidified nephelometer was removed and shipped back to CMDL to test and work on the 
relative humidity sensor inside the instrument.
Measurements:sgpaosC1.a0:
  • TSINephTSamp_HRH
  • HGupstream_RH
  • Backscattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_10um_HRH)
  • HGupstream_T
  • HGdownstream_T
  • Total scattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_1um_HRH)
  • TSINephPres_HRH
  • Total scattering coefficient at 1 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_1um_HRH)
  • Backscattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_1um_HRH)
  • Backscattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_1um_HRH)
  • HGdownstream_RH
  • Backscattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_10um_HRH)
  • Total scattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_10um_HRH)
  • TSINephRHSamp_HRH
  • Backscattering coefficient at 10 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_10um_HRH)
  • Total scattering coefficient at 10 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_10um_HRH)
  • Total scattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_10um_HRH)
  • Backscattering coefficient at 1 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_1um_HRH)
  • TSINephTin_HRH
  • Total scattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_1um_HRH)


Back To Table of Contents

DQRID : D051024.4
Start DateStart TimeEnd DateEnd Time
10/14/2005153610/28/20051940
Subject:
SGP/AOS/C1 - Nephelometer calibration drift
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
Bad span check on the humidified nepehelometer.  Instrument was recalibrated and returned 
to service on 10/28/2005 at 1940 UTC.
Measurements:sgpaosC1.a0:
  • TSINephTSamp_HRH
  • HGupstream_RH
  • Backscattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_10um_HRH)
  • Total scattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_1um_HRH)
  • TSINephPres_HRH
  • Total scattering coefficient at 1 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_1um_HRH)
  • Backscattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_1um_HRH)
  • Backscattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_1um_HRH)
  • HGdownstream_RH
  • Backscattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_10um_HRH)
  • Total scattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_10um_HRH)
  • TSINephRHSamp_HRH
  • Backscattering coefficient at 10 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_10um_HRH)
  • Total scattering coefficient at 10 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_10um_HRH)
  • Backscattering coefficient at 1 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_1um_HRH)
  • Total scattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_10um_HRH)
  • TSINephTin_HRH
  • Total scattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_1um_HRH)

sgpaosauxC1.a0:
  • BluBScatSpn_HRH
  • BluTScatZero_HRH
  • BluTScatSpn_HRH
  • BluRScatZero_HRH
  • BluBScatZero_HRH


Back To Table of Contents

DQRID : D980528.1
Start DateStart TimeEnd DateEnd Time
06/09/1999082407/26/19992205
09/08/2000042610/13/20001630
07/01/1997093607/03/19971556
08/01/1997132908/08/19971348
05/05/1998092606/16/19981505
05/17/2002060205/20/20051847
05/23/2002201105/31/20021258
09/09/2002000010/07/20021905
Subject:
SGP/AOS/C1 - Optical Particle Counter (PCASP-X) down
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The AOS Optical Particle Spectrometer (PCASP-X sizing probe) was down during these 
periods.  This lowers the laser reference voltage and results in inaccurate sizing.
Measurements:sgpaosC1.a0:
  • Particle concentration (0.50 um < Dp < 0.60 um) from PMS PCASP optical particle
    counter, channel 13(pa50_p60conc)
  • Particle concentration (1.80 um < Dp < 2.00 um) from PMS PCASP optical particle
    counter, channel 22(pa180_p200conc)
  • Particle concentration (6.50 um < Dp < 8.00 um) from PMS PCASP optical particle
    counter, channel 30(pa650_p800conc)
  • Particle concentration (2.00 um < Dp < 2.30 um) from PMS PCASP optical particle
    counter, channel 23(pa200_p230conc)
  • Particle concentration (0.30 um < Dp < 0.35 um) from PMS PCASP optical particle
    counter, channel 9(pa30_p35conc)
  • Particle concentration (0.16 um < Dp < 0.18 um) from PMS PCASP optical particle
    counter, channel 4(pa16_p18conc)
  • Particle concentration (2.30 um < Dp < 2.60 um) from PMS PCASP optical particle
    counter, channel 24(pa230_p260conc)
  • Particle concentration (3.00 um < Dp < 3.50 um) from PMS PCASP optical particle
    counter, channel 26(pa300_p350conc)
  • Particle concentration (0.70 um < Dp < 0.80 um) from PMS PCASP optical particle
    counter, channel 15(pa70_p80conc)
  • Particle concentration (1.00 um < Dp < 1.30 um) from PMS PCASP optical particle
    counter, channel 18(pa100_p130conc)
  • Particle concentration (0.20 um < Dp < 0.23 um) from PMS PCASP optical particle
    counter, channel 6(pa20_p23conc)
  • Particle concentration (Dp > 10 um) from PMS PCASP optical particle counter,
    channel 0(Pa1000Conc)
  • Particle concentration (0.45 um < Dp < 0.50 um) from PMS PCASP optical particle
    counter, channel 12(pa45_p50conc)
  • Particle concentration (5.00 um < Dp < 6.50 um) from PMS PCASP optical particle
    counter, channel 29(pa500_p650conc)
  • Particle concentration (0.10 um < Dp < 0.12 um) from PMS PCASP optical particle
    counter, channel 1(pa10_p12conc)
  • Particle concentration (2.60 um < Dp < 3.00 um) from PMS PCASP optical particle
    counter, channel 25(pa260_p300conc)
  • Particle concentration (1.40 um < Dp < 1.60 um) from PMS PCASP optical particle
    counter, channel 20(pa140_p160conc)
  • Particle concentration (0.14 um < Dp < 0.16 um) from PMS PCASP optical particle
    counter, channel 3(pa14_p16conc)
  • Particle concentration (0.40 um < Dp < 0.45 um) from PMS PCASP optical particle
    counter, channel 11(pa40_p45conc)
  • Particle concentration (0.90 um < Dp < 1.00 um) from PMS PCASP optical particle
    counter, channel 17(pa90_p100conc)
  • Particle concentration (1.60 um < Dp < 1.80 um) from PMS PCASP optical particle
    counter, channel 21(pa160_p180conc)
  • Particle concentration (3.50 um < Dp < 4.00 um) from PMS PCASP optical particle
    counter, channel 27(pa350_p400conc)
  • Particle concentration (0.12 um < Dp < 0.14 um) from PMS PCASP optical particle
    counter, channel 2(pa12_p14conc)
  • Particle concentration (0.23 um < Dp < 0.26 um) from PMS PCASP optical particle
    counter, channel 7(pa23_p26conc)
  • Particle concentration (8.00 um < Dp < 10.00 um) from PMS PCASP optical particle
    counter, channel 31(pa800_p1000conc)
  • Particle concentration (1.30 um < Dp < 1.40 um) from PMS PCASP optical particle
    counter, channel 19(pa130_p140conc)
  • Particle concentration (4.00 um < Dp < 5.00 um) from PMS PCASP optical particle
    counter, channel 28(pa400_p500conc)
  • Particle concentration (0.18 um < Dp < 0.20 um) from PMS PCASP optical particle
    counter, channel 5(pa18_p20conc)
  • Particle concentration (0.60 um < Dp < 0.70 um) from PMS PCASP optical particle
    counter, channel 14(pa60_p70conc)
  • Particle concentration (0.26 um < Dp < 0.30 um) from PMS PCASP optical particle
    counter, channel 8(pa26_p30conc)
  • Particle concentration (0.80 um < Dp < 0.90 um) from PMS PCASP optical particle
    counter, channel 16(pa80_p90conc)
  • Particle concentration (0.35 um < Dp < 0.40 um) from PMS PCASP optical particle
    counter, channel 10(pa35_p40conc)

sgpaosauxC1.a0:
  • OPCFlow
  • OPCSheath
  • OPCLaserVolt