Data Quality Reports for Session: 148603 User: manish3391 Completed: 03/09/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D120807.1PGH/AOS/M1 - Trailer temp too low, CCN affectedpghaosccn100M1.a1


DQRID : D120807.1
Start DateStart TimeEnd DateEnd Time
12/11/2011180002/10/20120600
Subject:
PGH/AOS/M1 - Trailer temp too low, CCN affected
DataStreams:pghaosccn100M1.a1
Description:
The trailer temperature wasn't regulating very well. Night time temperatures fell to 10 C. 
The CCN doesn't operate well at low temperatures. As a consequence the instrument 
temperatures were below the acceptable range for the model calculation of %ss. The instrument 
still reported a %ss values, but this value is likely off by as much as 0.1%, especially 
at night when the trailer temperatures were low.
Measurements:pghaosccn100M1.a1:
  • CCN_ss_set


Back To Table of Contents



END OF DATA