Data Quality Reports for Session: 147153 User: dumka Completed: 01/18/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D121029.2PGH/GNDRAD/M1 - Reprocessed: Incorrect calibrationspghgndrad20sM1.a0, pghgndrad60sM1.b1


DQRID : D121029.2
Start DateStart TimeEnd DateEnd Time
06/09/2011103303/31/20122359
Subject:
PGH/GNDRAD/M1 - Reprocessed: Incorrect calibrations
DataStreams:pghgndrad20sM1.a0, pghgndrad60sM1.b1
Description:
During the entire India deployment the Gndrad system was configured with incorrect 
calibration coefficients for the shortwave instruments. The program hadn't been updated since 
the previous deployment in the Azores.

The instrument mentor, site personal and OSS support have attempted to best determine the 
serial numbers of instruments installed at PGH during the deployment period.  The 
following instrument data will need to be reprocessed with the following correction multipliers:
PIR-UIR(up_long_hemisp): OK, no correction required
PSP-US (up_short_hemsip): 117.97(correct)/129.42(incorrect) = 0.91153
IRT: calibration should be validated by instrument mentor (Vic Morris): mult=25, 
offset=223.2

Correct serial numbers:
PSP-US (up_short_hemsip): 33941F3

Reprocessing completed 20130115
Measurements:pghgndrad20sM1.a0:
  • inst_up_short_hemisp
  • inst_sfc_ir_signal

pghgndrad60sM1.b1:
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • sfc_ir_temp_std
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Radiation, shortwave, at 10-m height, upwelling hemispheric irrad., 1-min avg(up_short_hemisp)
  • Upwelling Shortwave Hemispheric Irradiance, Unventilated Pyranometer, Maxima(up_short_hemisp_max)
  • Upwelling Shortwave Hemispheric Irradiance, Unventilated Pyranometer, Minima(up_short_hemisp_min)
  • up_short_hemisp_std
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)


Back To Table of Contents



END OF DATA