D111017.1GAN/GNDRAD/M1 - IRT calibration errorgangndrad60sM1.b1
D111101.1GAN/GNDRAD/M1 - Incorrect up_long_hemispgangndrad20sM1.a0, gangndrad60sM1.b1


DQRID : D111017.1
Start DateStart TimeEnd DateEnd Time
09/23/2011074210/09/20110413
Subject:
GAN/GNDRAD/M1 - IRT calibration error
DataStreams:gangndrad60sM1.b1
Description:
The IRT surface temperature reported was incorrect because the analog output configuration 
did not agree with the data logger scale factors. The IRT configuration was corrected on 
10/9.
Measurements:gangndrad60sM1.b1:
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)


Back To Table of Contents

DQRID : D111101.1
Start DateStart TimeEnd DateEnd Time
09/25/2011000010/05/20110444
Subject:
GAN/GNDRAD/M1 - Incorrect up_long_hemisp
DataStreams:gangndrad20sM1.a0, gangndrad60sM1.b1
Description:
The up_long_hemisp values from the pyrgeometer were much too low, hovering near or just 
above 100 W/m^2.  This was particularly noticeable when comparing these values to those 
derived from the surface IRT.   

On 10/05/11, at 0348Z, tower was lowered for removal of PIR SN35836 and installation of 
PIR SN35839; tower raised at 0444Z. Following installation of PIR SN35839, data quality 
plots indicate proper up_long_hemisp values.
Measurements:gangndrad20sM1.a0:
  • inst_up_long_hemisp_dome_resist
  • inst_up_long_hemisp_case_resist
  • inst_up_long_hemisp_tp

gangndrad60sM1.b1:
  • up_long_hemisp_std
  • Upwelling Longwave Hemispheric Irradiance, Unventilated Pyrgeometer, Maxima(up_long_hemisp_max)
  • Upwelling Longwave Hemispheric Irradiance, Unventilated Pyrgeometer, Minima(up_long_hemisp_min)
  • Radiation, longwave, at 10-m height, upwelling hemispheric irrad., 1-min avg(up_long_hemisp)
  • inst_up_long_hemisp_tp
  • inst_up_long_hemisp_dome_temp
  • inst_up_long_hemisp_case_temp


Back To Table of Contents