Data Quality Reports for Session: 146893 User: ql_data Completed: 01/09/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D120918.2TWP/IRT/C2 - Data acquisition problemtwpirt200msC2.a1, twpirtC2.b1


DQRID : D120918.2
Start DateStart TimeEnd DateEnd Time
06/20/2012134007/20/20120410
Subject:
TWP/IRT/C2 - Data acquisition problem
DataStreams:twpirt200msC2.a1, twpirtC2.b1
Description:
Serial communication errors resulted in incomplete raw data resulting in incorrect 
processed data records. Response strings from the IRT were occasionally not parsed correctly 
causing a much smaller value of temperature (for example, '29' instead of '294.6') to be 
recorded.

The acquistion program was updated to improve communication with the IRT.
Measurements:twpirtC2.b1:
  • sky_ir_temp_min
  • IR Brightness Temperature(sky_ir_temp)
  • sky_ir_temp_max
  • sky_ir_temp_std

twpirt200msC2.a1:
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents