Data Quality Reports for Session: 146793 User: sws0982 Completed: 01/07/2013


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040130.1SGP/AOS/C1 - Ozone Instrument removed for repairsgpaosC1.a0, sgpaosauxC1.a0
D041026.3SGP/AOS/C1 - PCASP instrument problemsgpaosC1.a0, sgpaosauxC1.a0
D041026.4SGP/AOS/C1 - Ozone instrument removed for repairsgpaosauxC1.a0, sgpaosC1.a0
D041026.5SGP/AOS/C1 - Condensation Particle Counter failuresgpaosC1.a0, sgpaosauxC1.a0
D041026.6SGP/AOS/C1 - Wet nephelometer problemssgpaosC1.a0, sgpaosauxC1.a0
D050120.2SGP/AOS/C1 - Failed RH sensorsgpaosC1.a0
D980528.1SGP/AOS/C1 - Optical Particle Counter (PCASP-X) downsgpaosC1.a0, sgpaosauxC1.a0


DQRID : D040130.1
Start DateStart TimeEnd DateEnd Time
11/21/2003163601/27/20042208
Subject:
SGP/AOS/C1 - Ozone Instrument removed for repair
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The ozone instrument was removed for calibration on 2003 11/21.  It was shipped to Boulder 
and was cleaned and calibrated.  It was sent back to SGP on 2003 12/02, but was damaged 
in transit by FedEx.  The instrument was reinstalled on 2003 12/03 but was found to be 
defective.  Large spikes were observed in the data.  Afetr a few days the instrument was 
removed and shipped to a vendor for repair.  The ozone monitor was returned to service 
after repair at 2004 01/27 (2208 GMT).
Measurements:sgpaosC1.a0:
  • Ozone concentration(Ozone)

sgpaosauxC1.a0:
  • O3_SetPt
  • O3_Span
  • O3_Cont
  • O3_Press
  • O3_LinHeater
  • O3_Offset
  • O3_LinTemp
  • O3_Mode
  • O3_Temp
  • O3_Spn
  • O3_Samp
  • O3_Zro


Back To Table of Contents

DQRID : D041026.3
Start DateStart TimeEnd DateEnd Time
01/10/2004000010/26/20041200
Subject:
SGP/AOS/C1 - PCASP instrument problem
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
PCASP (optical particle counter) gave unusually low aerosol signal counts despite numerous 
attempts to fix the instrument by the mentor and manufacturer.
Measurements:sgpaosC1.a0:
  • Particle concentration (0.50 um < Dp < 0.60 um) from PMS PCASP optical particle
    counter, channel 13(pa50_p60conc)
  • Particle concentration (1.80 um < Dp < 2.00 um) from PMS PCASP optical particle
    counter, channel 22(pa180_p200conc)
  • Particle concentration (6.50 um < Dp < 8.00 um) from PMS PCASP optical particle
    counter, channel 30(pa650_p800conc)
  • Particle concentration (2.00 um < Dp < 2.30 um) from PMS PCASP optical particle
    counter, channel 23(pa200_p230conc)
  • Particle concentration (0.30 um < Dp < 0.35 um) from PMS PCASP optical particle
    counter, channel 9(pa30_p35conc)
  • Particle concentration (0.16 um < Dp < 0.18 um) from PMS PCASP optical particle
    counter, channel 4(pa16_p18conc)
  • Particle concentration (2.30 um < Dp < 2.60 um) from PMS PCASP optical particle
    counter, channel 24(pa230_p260conc)
  • Particle concentration (0.70 um < Dp < 0.80 um) from PMS PCASP optical particle
    counter, channel 15(pa70_p80conc)
  • Particle concentration (3.00 um < Dp < 3.50 um) from PMS PCASP optical particle
    counter, channel 26(pa300_p350conc)
  • Particle concentration (1.00 um < Dp < 1.30 um) from PMS PCASP optical particle
    counter, channel 18(pa100_p130conc)
  • Particle concentration (0.20 um < Dp < 0.23 um) from PMS PCASP optical particle
    counter, channel 6(pa20_p23conc)
  • Particle concentration (Dp > 10 um) from PMS PCASP optical particle counter,
    channel 0(Pa1000Conc)
  • Particle concentration (0.45 um < Dp < 0.50 um) from PMS PCASP optical particle
    counter, channel 12(pa45_p50conc)
  • Particle concentration (5.00 um < Dp < 6.50 um) from PMS PCASP optical particle
    counter, channel 29(pa500_p650conc)
  • Particle concentration (0.10 um < Dp < 0.12 um) from PMS PCASP optical particle
    counter, channel 1(pa10_p12conc)
  • Particle concentration (2.60 um < Dp < 3.00 um) from PMS PCASP optical particle
    counter, channel 25(pa260_p300conc)
  • Particle concentration (1.40 um < Dp < 1.60 um) from PMS PCASP optical particle
    counter, channel 20(pa140_p160conc)
  • Particle concentration (0.14 um < Dp < 0.16 um) from PMS PCASP optical particle
    counter, channel 3(pa14_p16conc)
  • Particle concentration (0.40 um < Dp < 0.45 um) from PMS PCASP optical particle
    counter, channel 11(pa40_p45conc)
  • Particle concentration (0.90 um < Dp < 1.00 um) from PMS PCASP optical particle
    counter, channel 17(pa90_p100conc)
  • Particle concentration (1.60 um < Dp < 1.80 um) from PMS PCASP optical particle
    counter, channel 21(pa160_p180conc)
  • Particle concentration (3.50 um < Dp < 4.00 um) from PMS PCASP optical particle
    counter, channel 27(pa350_p400conc)
  • Particle concentration (0.12 um < Dp < 0.14 um) from PMS PCASP optical particle
    counter, channel 2(pa12_p14conc)
  • Particle concentration (0.23 um < Dp < 0.26 um) from PMS PCASP optical particle
    counter, channel 7(pa23_p26conc)
  • Particle concentration (8.00 um < Dp < 10.00 um) from PMS PCASP optical particle
    counter, channel 31(pa800_p1000conc)
  • Particle concentration (1.30 um < Dp < 1.40 um) from PMS PCASP optical particle
    counter, channel 19(pa130_p140conc)
  • Particle concentration (4.00 um < Dp < 5.00 um) from PMS PCASP optical particle
    counter, channel 28(pa400_p500conc)
  • Particle concentration (0.18 um < Dp < 0.20 um) from PMS PCASP optical particle
    counter, channel 5(pa18_p20conc)
  • Particle concentration (0.60 um < Dp < 0.70 um) from PMS PCASP optical particle
    counter, channel 14(pa60_p70conc)
  • Particle concentration (0.26 um < Dp < 0.30 um) from PMS PCASP optical particle
    counter, channel 8(pa26_p30conc)
  • Particle concentration (0.80 um < Dp < 0.90 um) from PMS PCASP optical particle
    counter, channel 16(pa80_p90conc)
  • Particle concentration (0.35 um < Dp < 0.40 um) from PMS PCASP optical particle
    counter, channel 10(pa35_p40conc)

sgpaosauxC1.a0:
  • OPCFlow
  • OPCSheath
  • OPCLaserVolt


Back To Table of Contents

DQRID : D041026.4
Start DateStart TimeEnd DateEnd Time
11/21/2003160001/28/20040000
Subject:
SGP/AOS/C1 - Ozone instrument removed for repair
DataStreams:sgpaosauxC1.a0, sgpaosC1.a0
Description:
The ozone monitor was removed and sent back to the manufacturer for repair.
Measurements:sgpaosC1.a0:
  • Ozone concentration(Ozone)

sgpaosauxC1.a0:
  • O3_Span
  • O3_SetPt
  • O3_Cont
  • O3_Press
  • O3_LinHeater
  • O3_Offset
  • O3_LinTemp
  • O3_Mode
  • O3_Temp
  • O3_Spn
  • O3_Samp
  • O3_Zro


Back To Table of Contents

DQRID : D041026.5
Start DateStart TimeEnd DateEnd Time
07/28/2004140008/27/20041540
Subject:
SGP/AOS/C1 - Condensation Particle Counter failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The CPC signal declined to a low value. The instrument was shipped back to the mentor, 
cleaned and repaired. The instrument inlet was clogged.
Measurements:sgpaosC1.a0:
  • Particle concentration from Condensation Particle Counter(CPCPartConc)

sgpaosauxC1.a0:
  • CPCSaturT
  • CPCDeltT
  • CPCCondT


Back To Table of Contents

DQRID : D041026.6
Start DateStart TimeEnd DateEnd Time
01/24/2004000002/18/20041800
Subject:
SGP/AOS/C1 - Wet nephelometer problems
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The humidifier tube was coated with mold and had to be replaced. This affects the wet 
nepheleometer measurements and the calculations of the aerosol hygroscopic growth.

The humidifer was cleaned and replaced. A filter was added to the water circulation to 
minimize mold.
Measurements:sgpaosC1.a0:
  • HGupstream_RH
  • TSINephTSamp_HRH
  • Backscattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_10um_HRH)
  • HGdownstream_T
  • Total scattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_1um_HRH)
  • Total scattering coefficient at 1 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_1um_HRH)
  • TSINephPres_HRH
  • Backscattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_1um_HRH)
  • HGdownstream_RH
  • Backscattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_1um_HRH)
  • Total scattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_10um_HRH)
  • TSINephRHSamp_HRH
  • Total scattering coefficient at 10 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_10um_HRH)
  • Total scattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_1um_HRH)
  • HGupstream_T
  • Backscattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_10um_HRH)
  • Backscattering coefficient at 10 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_10um_HRH)
  • Backscattering coefficient at 1 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_1um_HRH)
  • Total scattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_10um_HRH)
  • TSINephTin_HRH

sgpaosauxC1.a0:
  • RedTScatZero_HRH
  • GrnRScatZero_HRH
  • BluRScatZero_HRH
  • BluTScatZero_HRH
  • TSINephLV_HRH
  • GrnBScatSpn_HRH
  • RedRScatZero_HRH
  • BluBScatZero_HRH
  • GrnTScatSpn_HRH
  • RedBScatZero_HRH
  • BluBScatSpn_HRH
  • GrnTScatZero_HRH
  • TSISensGrn_HRH
  • BluTScatSpn_HRH
  • RedTScatSpn_HRH
  • GrnBScatZero_HRH
  • RedBScatSpn_HRH
  • TSINephLC_HRH


Back To Table of Contents

DQRID : D050120.2
Start DateStart TimeEnd DateEnd Time
01/01/2004000005/01/20052359
Subject:
SGP/AOS/C1 - Failed RH sensor
DataStreams:sgpaosC1.a0
Description:
The RH sensor inside the humidified nephelometer began to slowly decline in value in the 
upper RH range beginning in early 2004 and becoming more noticable poor on 20041117. The 
sensor currently reaches a high RH value of about 60% when the actual RH is closer to 80%.

The sensor was replaced in the Spring of 2005, but the problem remained.  Effective 
1/1/2004, this sensor is not used to calculate scientific data and is therefore essentially 
abandoned in the field.  An end-date of 20050501 is being used as a placeholder.
Measurements:sgpaosC1.a0:
  • TSINephRHSamp_HRH


Back To Table of Contents

DQRID : D980528.1
Start DateStart TimeEnd DateEnd Time
06/09/1999082407/26/19992205
09/08/2000042610/13/20001630
07/01/1997093607/03/19971556
08/01/1997132908/08/19971348
05/05/1998092606/16/19981505
05/17/2002060205/20/20051847
05/23/2002201105/31/20021258
09/09/2002000010/07/20021905
Subject:
SGP/AOS/C1 - Optical Particle Counter (PCASP-X) down
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The AOS Optical Particle Spectrometer (PCASP-X sizing probe) was down during these 
periods.  This lowers the laser reference voltage and results in inaccurate sizing.
Measurements:sgpaosC1.a0:
  • Particle concentration (0.50 um < Dp < 0.60 um) from PMS PCASP optical particle
    counter, channel 13(pa50_p60conc)
  • Particle concentration (1.80 um < Dp < 2.00 um) from PMS PCASP optical particle
    counter, channel 22(pa180_p200conc)
  • Particle concentration (6.50 um < Dp < 8.00 um) from PMS PCASP optical particle
    counter, channel 30(pa650_p800conc)
  • Particle concentration (2.00 um < Dp < 2.30 um) from PMS PCASP optical particle
    counter, channel 23(pa200_p230conc)
  • Particle concentration (0.30 um < Dp < 0.35 um) from PMS PCASP optical particle
    counter, channel 9(pa30_p35conc)
  • Particle concentration (0.16 um < Dp < 0.18 um) from PMS PCASP optical particle
    counter, channel 4(pa16_p18conc)
  • Particle concentration (2.30 um < Dp < 2.60 um) from PMS PCASP optical particle
    counter, channel 24(pa230_p260conc)
  • Particle concentration (3.00 um < Dp < 3.50 um) from PMS PCASP optical particle
    counter, channel 26(pa300_p350conc)
  • Particle concentration (0.70 um < Dp < 0.80 um) from PMS PCASP optical particle
    counter, channel 15(pa70_p80conc)
  • Particle concentration (1.00 um < Dp < 1.30 um) from PMS PCASP optical particle
    counter, channel 18(pa100_p130conc)
  • Particle concentration (0.20 um < Dp < 0.23 um) from PMS PCASP optical particle
    counter, channel 6(pa20_p23conc)
  • Particle concentration (Dp > 10 um) from PMS PCASP optical particle counter,
    channel 0(Pa1000Conc)
  • Particle concentration (0.45 um < Dp < 0.50 um) from PMS PCASP optical particle
    counter, channel 12(pa45_p50conc)
  • Particle concentration (5.00 um < Dp < 6.50 um) from PMS PCASP optical particle
    counter, channel 29(pa500_p650conc)
  • Particle concentration (0.10 um < Dp < 0.12 um) from PMS PCASP optical particle
    counter, channel 1(pa10_p12conc)
  • Particle concentration (2.60 um < Dp < 3.00 um) from PMS PCASP optical particle
    counter, channel 25(pa260_p300conc)
  • Particle concentration (1.40 um < Dp < 1.60 um) from PMS PCASP optical particle
    counter, channel 20(pa140_p160conc)
  • Particle concentration (0.14 um < Dp < 0.16 um) from PMS PCASP optical particle
    counter, channel 3(pa14_p16conc)
  • Particle concentration (0.40 um < Dp < 0.45 um) from PMS PCASP optical particle
    counter, channel 11(pa40_p45conc)
  • Particle concentration (0.90 um < Dp < 1.00 um) from PMS PCASP optical particle
    counter, channel 17(pa90_p100conc)
  • Particle concentration (1.60 um < Dp < 1.80 um) from PMS PCASP optical particle
    counter, channel 21(pa160_p180conc)
  • Particle concentration (3.50 um < Dp < 4.00 um) from PMS PCASP optical particle
    counter, channel 27(pa350_p400conc)
  • Particle concentration (0.12 um < Dp < 0.14 um) from PMS PCASP optical particle
    counter, channel 2(pa12_p14conc)
  • Particle concentration (0.23 um < Dp < 0.26 um) from PMS PCASP optical particle
    counter, channel 7(pa23_p26conc)
  • Particle concentration (8.00 um < Dp < 10.00 um) from PMS PCASP optical particle
    counter, channel 31(pa800_p1000conc)
  • Particle concentration (1.30 um < Dp < 1.40 um) from PMS PCASP optical particle
    counter, channel 19(pa130_p140conc)
  • Particle concentration (4.00 um < Dp < 5.00 um) from PMS PCASP optical particle
    counter, channel 28(pa400_p500conc)
  • Particle concentration (0.18 um < Dp < 0.20 um) from PMS PCASP optical particle
    counter, channel 5(pa18_p20conc)
  • Particle concentration (0.60 um < Dp < 0.70 um) from PMS PCASP optical particle
    counter, channel 14(pa60_p70conc)
  • Particle concentration (0.26 um < Dp < 0.30 um) from PMS PCASP optical particle
    counter, channel 8(pa26_p30conc)
  • Particle concentration (0.80 um < Dp < 0.90 um) from PMS PCASP optical particle
    counter, channel 16(pa80_p90conc)
  • Particle concentration (0.35 um < Dp < 0.40 um) from PMS PCASP optical particle
    counter, channel 10(pa35_p40conc)

sgpaosauxC1.a0:
  • OPCFlow
  • OPCSheath
  • OPCLaserVolt


Back To Table of Contents



END OF DATA