Data Quality Reports for Session: 146225 User: wlj Completed: 11/26/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D120913.1PVC/SKYRAD/M1 - Intake Fan Failurepvcskyrad20sM1.a0, pvcskyrad60sM1.b1
D121008.1PVC/SKYRAD/M1 - Reprocess: Incorrect calibration coefficientspvcskyrad20sM1.a0, pvcskyrad60sM1.b1
D121018.10PVC/SKYRAD/M1 - Reprocess: Incorrect IRT Configurationpvcskyrad60sM1.b1
D121024.3PVC/SKYRAD/M1 - IRT failurepvcskyrad20sM1.a0, pvcskyrad60sM1.b1
D121024.5PVC/SKYRAD/M1 - IRT scale factorpvcskyrad20sM1.a0, pvcskyrad60sM1.b1


DQRID : D120913.1
Start DateStart TimeEnd DateEnd Time
06/25/2012000008/28/20122045
Subject:
PVC/SKYRAD/M1 - Intake Fan Failure
DataStreams:pvcskyrad20sM1.a0, pvcskyrad60sM1.b1
Description:
PSP (down_short_hemisp) ventilator fan failed from start of PVC deployment.  Site ops 
repaired/replaced on 8/28. Instrument data are not recoverable.

The down_short_hemisp data are not recoverable but can be re-constructed using the 
short_direct_normal and down_short_diffuse_hemispheric measurements and the following formula: 

Global = Direct * cos(zenith angle) + Diffuse
Measurements:pvcskyrad20sM1.a0:
  • inst_global

pvcskyrad60sM1.b1:
  • Radiation, shortwave, downwelling total hemispheric irradiance, 1-min avg(down_short_hemisp)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyranometer, Minima(down_short_hemisp_min)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyranometer, Maxima(down_short_hemisp_max)
  • down_short_hemisp_std


Back To Table of Contents

DQRID : D121008.1
Start DateStart TimeEnd DateEnd Time
06/25/2012000009/05/20122000
Subject:
PVC/SKYRAD/M1 - Reprocess: Incorrect calibration coefficients
DataStreams:pvcskyrad20sM1.a0, pvcskyrad60sM1.b1
Description:
Incorrect programs were used in Skyrad since PVC deployment period began.  The Skyrad 
program was corrected 09/05/2012 approx 20:00GMT. Reprocessing will be required of all 
instruments at this station except the IRT. This will require applying the following 
corrections for the time range specified. 
Serial numbers(incorrectly applied): 
PIR1-DIR: 34306
PIR2-DIR: 34303
Diffuse PSP: 34284
NIP: 34097
PSP-DS: 34293

Serial numbers(correct): 
PIR1-DIR: 34303
PIR2-DIR: 34306
Diffuse PSP: 35751
NIP: 37285
PSP-DS: 37302
    
Data: 
Longwave requires reprocessing using the 20second datastream and applying the following k1 
coefficients (note: the longwave instruments were swapped according to datalogger).
down_long_hemisp1 (old cal) = 0.23041
down_long_hemisp1 (corr cal) = 0.24750
down_long_hemisp2 (old cal) = 0.24750
down_long_hemisp2 (corr cal) = 0.23041

Shortwave only requires application of the following corrections to 1min and 20 second 
data.
down_short_diffuse_hemisp(corr) = down_short_diffuse_hemisp(old value)*(104.45(corr 
cal)/120.91(incorr cal))
short_direct_normal(corr) = short_direct_normal(old value)*(117.25(new cal)/124.97(incorr 
cal))
up_short_hemip(corr) = up_short_hemip(old value)*(119.21(new cal)/108.48(incorr cal))

Contact radiometer mentor with any comments/questions
Measurements:pvcskyrad20sM1.a0:
  • inst_down_long_hemisp1_case_resist
  • inst_global
  • inst_diffuse
  • inst_direct_normal
  • inst_down_long_hemisp2_case_resist
  • inst_down_long_hemisp2_dome_resist
  • inst_down_long_hemisp1_dome_resist
  • inst_down_long_hemisp1_tp
  • inst_down_long_hemisp2_tp

pvcskyrad60sM1.b1:
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • down_long_hemisp2_std
  • Radiation, shortwave, downwelling total hemispheric irradiance, 1-min avg(down_short_hemisp)
  • inst_down_long_hemisp2_case_temp
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp2)
  • Radiation, shortwave, downwelling diffuse hemispheric irradiance, 1-min avg(down_short_diffuse_hemisp)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyranometer, Maxima(down_short_hemisp_max)
  • inst_down_long_hemisp2_tp
  • down_long_hemisp1_max
  • down_long_hemisp2_min
  • Downwelling Shortwave Diffuse Hemisp. Irrad., Ventilated Pyranometer, Maxima(down_short_diffuse_hemisp_max)
  • down_long_hemisp1_std
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
  • Downwelling Shortwave Diffuse Hemisp. Irrad., Ventilated Pyranometer, Minima(down_short_diffuse_hemisp_min)
  • down_short_diffuse_hemisp_std
  • short_direct_normal_std
  • down_long_hemisp2_max
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyranometer, Minima(down_short_hemisp_min)
  • down_short_hemisp_std
  • inst_down_long_hemisp2_dome_temp
  • inst_down_long_hemisp1_case_temp
  • inst_down_long_hemisp1_dome_temp
  • down_long_hemisp1_min
  • inst_down_long_hemisp1_tp


Back To Table of Contents

DQRID : D121018.10
Start DateStart TimeEnd DateEnd Time
06/29/2012224010/18/20121259
Subject:
PVC/SKYRAD/M1 - Reprocess: Incorrect IRT Configuration
DataStreams:pvcskyrad60sM1.b1
Description:
The SKYRAD IRT was installed temporarily (waiting for the new enclosure) in
a vertical orientation, without the use of the gold mirror. When irt_comm
was started, the emissivity parameter in the startup file was not changed to 1.000 
compensate for the absence of the mirror, resulting in a positive bias of about 5 K.

These data may be corrected by adjusting the radiance values for the difference in 
emmissivity.
Measurements:pvcskyrad60sM1.b1:
  • sky_ir_temp_min
  • IR Brightness Temperature(sky_ir_temp)
  • sky_ir_temp_max


Back To Table of Contents

DQRID : D121024.3
Start DateStart TimeEnd DateEnd Time
10/01/2012055010/12/20121459
Subject:
PVC/SKYRAD/M1 - IRT failure
DataStreams:pvcskyrad20sM1.a0, pvcskyrad60sM1.b1
Description:
The IRT failed because of water intrusion.
Measurements:pvcskyrad20sM1.a0:
  • inst_sky_ir_signal

pvcskyrad60sM1.b1:
  • sky_ir_temp_std
  • sky_ir_temp_min
  • IR Brightness Temperature(sky_ir_temp)
  • sky_ir_temp_max


Back To Table of Contents

DQRID : D121024.5
Start DateStart TimeEnd DateEnd Time
10/18/2012130010/22/20121259
Subject:
PVC/SKYRAD/M1 - IRT scale factor
DataStreams:pvcskyrad20sM1.a0, pvcskyrad60sM1.b1
Description:
The IRT was replaced with a unit with the wrong analog output configuration, such that the 
sky temperature was below the measurable range, resulting in a constant output of 0 
volts or 173.2 K.
Measurements:pvcskyrad20sM1.a0:
  • inst_sky_ir_signal

pvcskyrad60sM1.b1:
  • sky_ir_temp_std
  • sky_ir_temp_min
  • IR Brightness Temperature(sky_ir_temp)
  • sky_ir_temp_max


Back To Table of Contents



END OF DATA