Data Quality Reports for Session: 145107 User: hs32 Completed: 09/28/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050614.1SGP/VCEIL/C1 - poor signalsgpvceil25kC1.b1
D051219.1NSA/VCEIL/C1 - poor sensitivity at low altitudensavceil25kC1.b1
D060113.1TWP/VCEIL/C3 - low receiver sensitivitytwpvceil25kC3.b1
D060517.3TWP/VCEIL/C2 - Vceil computer disconnectedtwpvceil25kC2.b1
D060717.5TWP/VCEIL/C2 - Time offsettwpvceil25kC2.b1
D061220.34TWP/VCEIL/C3 - standard instrument installedtwpvceil25kC3.b1
D070201.1TWP/VCEIL/C3 - Reduced sensitivity to high cloudstwpvceil25kC3.b1
D070201.2TWP/VCEIL/C2 - optical cross-talk compensationtwpvceil25kC2.b1
D070219.1SGP/VCEIL/B6 - contaminated windowsgpvceil25kB6.b1
D070329.1TWP/VCEIL/C2 - Receiver failuretwpvceil25kC2.b1
D070503.3TWP/VCEIL/C3 - Reduced sensitivity to high cloudstwpvceil25kC3.b1
D070522.1SGP/VCEIL/B6 - contaminated windowsgpvceil25kB6.b1
D071101.2SGP/VCEIL/B5 - Degraded hardwaresgpvceil25kB5.b1
D080924.4SGP/VCEIL/B5 - Missing datasgpvceil25kB5.b1
D090108.1SGP/VCEIL/B4 - Window contaminationsgpvceil25kB4.b1
D090507.2GRW/VCEIL/M1 - Receiver failuregrwvceil25kM1.b1
D090527.1TWP/VCEIL/C1 - Compensation fibertwpvceil25kC1.b1
D091026.1SGP/VCEIL/B1 - Warning flagsgpvceil25kB1.b1
D091210.1NSA/VCEIL/C1 - Degraded windownsavceil25kC1.b1
D100215.1SGP/VCEIL/C1/B4/B5/B6 - Reprocess: Incorrect serial_number and sum_backscatter fieldssgpvceil25kB4.b1, sgpvceil25kB5.b1, sgpvceil25kB6.b1, sgpvceil25kC1.b1
D100215.2SGP/VCEIL/B1 - Reprocess: Incorrect serial_number and sum_backscatter fieldssgpvceil25kB1.b1
D100215.3NSA/VCEIL/C1/C2 - Reprocess: Incorrect serial_number and sum_backscatter fieldsnsavceil25kC1.b1, nsavceil25kC2.b1
D100215.4TWP/VCEIL/C1 - Reprocess: Incorrect serial_number and sum_backscatter fieldstwpvceil25kC1.b1
D100215.5TWP/VCEIL/C2 - Reprocess: Incorrect serial_number and sum_backscatter fieldstwpvceil25kC2.b1
D100215.6TWP/VCEIL/C3 - Reprocess: Incorrect serial_number and sum_backscatter fieldstwpvceil25kC3.b1
D100216.3NIM/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fieldsnimvceil25kM1.b1
D100216.4FKB/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fieldsfkbvceil25kM1.b1
D100216.5HFE/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fieldshfevceil25kM1.b1
D100216.6GRW/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fieldsgrwvceil25kM1.b1
D100225.2HFE/VCEIL/S1 - Incorrect serial_number and sum_backscatter fieldshfevceil25kS1.b1
D100616.2NSA/VCEIL/C1 - Small gaps in datansavceil25kC1.b1


DQRID : D050614.1
Start DateStart TimeEnd DateEnd Time
05/05/2005184204/19/20061713
Subject:
SGP/VCEIL/C1 - poor signal
DataStreams:sgpvceil25kC1.b1
Description:
When parameters were changed (set noise_h2_compensation to "ON" and units to "METERS") to 
correct a low amplitude problem, the backscatter plots began displaying horizontal 
stripes indicating detector "ringing". However, the cloud base heights are not affected. The 
problem was corrected when the instrument was replaced.
Measurements:sgpvceil25kC1.b1:
  • detection_status
  • Backscatter(backscatter)


Back To Table of Contents

DQRID : D051219.1
Start DateStart TimeEnd DateEnd Time
12/27/2005193005/02/20060000
Subject:
NSA/VCEIL/C1 - poor sensitivity at low altitude
DataStreams:nsavceil25kC1.b1
Description:
The ceilometer has an internal amplifier that it sets to low or high depending on 
prevalant conditions. Normally, the gain is automatically set high during clear skies. It appears 
that during these times, a slight degradation in either the transmitter or receiver is 
being amplified and produces a "ripple" at low altitudes in the backscatter plot. Cloud 
base height detection is not affected.
Measurements:nsavceil25kC1.b1:
  • receiver_sensitivity
  • Backscatter(backscatter)


Back To Table of Contents

DQRID : D060113.1
Start DateStart TimeEnd DateEnd Time
12/12/2005055505/18/20062300
Subject:
TWP/VCEIL/C3 - low receiver sensitivity
DataStreams:twpvceil25kC3.b1
Description:
The ceilometer reported frequent receiver alarms. There are periods when the receiver 
sensitivity dropped to about 50%. The problem was corrected by replacing the instrument with 
a spare.
Measurements:twpvceil25kC3.b1:
  • receiver_sensitivity
  • Backscatter(backscatter)
  • sum_backscatter
  • Second lowest cloud base height(second_cbh)
  • Third cloud base height(third_cbh)
  • Lowest cloud base height detected.(first_cbh)
  • background_light
  • alt_highest_signal
  • vertical_visibility


Back To Table of Contents

DQRID : D060517.3
Start DateStart TimeEnd DateEnd Time
05/13/2006224905/15/20060410
Subject:
TWP/VCEIL/C2 - Vceil computer disconnected
DataStreams:twpvceil25kC2.b1
Description:
VCEIL computer was disconnected during this time period.
Measurements:twpvceil25kC2.b1:
  • status_string
  • detection_status
  • sum_backscatter
  • laser_pulse_energy
  • tilt_angle
  • alt_highest_signal
  • filename
  • laser_temperature
  • vertical_visibility
  • range
  • Third cloud base height(third_cbh)
  • measurement_parameters
  • status_flag
  • Second lowest cloud base height(second_cbh)
  • Lowest cloud base height detected.(first_cbh)
  • background_light
  • Backscatter(backscatter)
  • window_contamination
  • base_time
  • receiver_sensitivity
  • time_offset


Back To Table of Contents

DQRID : D060717.5
Start DateStart TimeEnd DateEnd Time
05/13/2006222407/17/20061912
Subject:
TWP/VCEIL/C2 - Time offset
DataStreams:twpvceil25kC2.b1
Description:
On 7/14/2006, I found that the ceilometer time was 8m 36s slow, the PC time display (lower 
right taskbar) was frozen at "22:24", and AboutTime was not running. Corrected it on 
7/17 by running AboutTime so that 19:03:29 became 19:12:05. This also corrected the frozen 
display on the Windows98 taskbar. 

I am not sure of the start time of this problem but there were several other Windows 
related issues that had to be corrected on 5/14 with indications of inappropriate access via 
Remote Administrator on 5/13. I do not know if the time offset was a sudden or gradual 
change.
Measurements:twpvceil25kC2.b1:
  • status_string
  • detection_status
  • sum_backscatter
  • laser_pulse_energy
  • tilt_angle
  • alt_highest_signal
  • laser_temperature
  • vertical_visibility
  • range
  • Third cloud base height(third_cbh)
  • measurement_parameters
  • Second lowest cloud base height(second_cbh)
  • status_flag
  • Lowest cloud base height detected.(first_cbh)
  • background_light
  • Backscatter(backscatter)
  • window_contamination
  • base_time
  • receiver_sensitivity
  • time_offset


Back To Table of Contents

DQRID : D061220.34
Start DateStart TimeEnd DateEnd Time
05/18/2006230012/20/20060400
Subject:
TWP/VCEIL/C3 - standard instrument installed
DataStreams:twpvceil25kC3.b1
Description:
A spare ceilometer without the tropics window option was installed. The manufacturer 
recommends the tropics window be used at +/- 25 deg lattitude to prevent direct solar 
radiation into the laser optics when the instrument pointed at zenith. No significant signal 
degradation is apparent during the autumnal equinox but, as a precaution, the unit was 
replaced with the original with a tropics window.
Measurements:twpvceil25kC3.b1:
  • Backscatter(backscatter)
  • laser_pulse_energy
  • laser_temperature
  • sum_backscatter
  • background_light
  • vertical_visibility


Back To Table of Contents

DQRID : D070201.1
Start DateStart TimeEnd DateEnd Time
12/20/2006040101/12/20070120
Subject:
TWP/VCEIL/C3 - Reduced sensitivity to high clouds
DataStreams:twpvceil25kC3.b1
Description:
The instrument produced a significant number of optical cross-talk warnings and displayed 
a reduced capacity to detect high clouds.  The internal cables were inspected, the 
compensation fiber and optics monitor were replaced, the crosstalk compensation was adjusted, 
and the system tested with an optical termination hood, all with no improvement.  While 
waiting for a new receiver board, receiver ribbon cable, and compensation fiber cable from 
Vaisala, a spare ceilometer with a standard, "non-tropics", window was installed.
Measurements:twpvceil25kC3.b1:
  • receiver_sensitivity
  • Second lowest cloud base height(second_cbh)
  • Third cloud base height(third_cbh)
  • detection_status
  • alt_highest_signal


Back To Table of Contents

DQRID : D070201.2
Start DateStart TimeEnd DateEnd Time
12/13/2006000002/05/20070028
Subject:
TWP/VCEIL/C2 - optical cross-talk compensation
DataStreams:twpvceil25kC2.b1
Description:
The ceilometer gave frequent optical cross-talk compensation warnings, was insensitive to 
high clouds, and the tropics window was deteriorated. The instrument was replaced with a 
spare.
Measurements:twpvceil25kC2.b1:
  • Second lowest cloud base height(second_cbh)
  • alt_highest_signal
  • Backscatter(backscatter)
  • window_contamination
  • vertical_visibility
  • Third cloud base height(third_cbh)


Back To Table of Contents

DQRID : D070219.1
Start DateStart TimeEnd DateEnd Time
02/08/2007150002/27/20072035
Subject:
SGP/VCEIL/B6 - contaminated window
DataStreams:sgpvceil25kB6.b1
Description:
Backscatter has consistently high amplitude because the window was dirty. The problem was 
corrected when the window was cleaned.
Measurements:sgpvceil25kB6.b1:
  • Third cloud base height(third_cbh)
  • Lowest cloud base height detected.(first_cbh)
  • Second lowest cloud base height(second_cbh)
  • Backscatter(backscatter)


Back To Table of Contents

DQRID : D070329.1
Start DateStart TimeEnd DateEnd Time
03/11/2007212004/17/20072325
Subject:
TWP/VCEIL/C2 - Receiver failure
DataStreams:twpvceil25kC2.b1
Description:
The receiver failed and the instrument was replaced.  Data during the specified period are 
not valid.
Measurements:twpvceil25kC2.b1:
  • sum_backscatter
  • Second lowest cloud base height(second_cbh)
  • Lowest cloud base height detected.(first_cbh)
  • alt_highest_signal
  • Backscatter(backscatter)
  • background_light
  • vertical_visibility
  • Third cloud base height(third_cbh)


Back To Table of Contents

DQRID : D070503.3
Start DateStart TimeEnd DateEnd Time
04/18/2007163007/11/20070550
Subject:
TWP/VCEIL/C3 - Reduced sensitivity to high clouds
DataStreams:twpvceil25kC3.b1
Description:
The instrument had reduced sensitivity to high clouds.

The instrument produced a significant number of optical cross-talk warnings and displayed 
a reduced capacity to detect high clouds. The window was cleaned and the instrument was 
power-cycled but with no improvement. Diagnostics indicated the problem was with the 
compensation fiber cable. 

The problem was corrected by replacing this standard-window instrument (that was tilted at 
16 degrees) with a tropics unit that became available.
Measurements:twpvceil25kC3.b1:
  • receiver_sensitivity
  • Second lowest cloud base height(second_cbh)
  • Third cloud base height(third_cbh)
  • alt_highest_signal
  • vertical_visibility


Back To Table of Contents

DQRID : D070522.1
Start DateStart TimeEnd DateEnd Time
05/10/2007233005/22/20071930
Subject:
SGP/VCEIL/B6 - contaminated window
DataStreams:sgpvceil25kB6.b1
Description:
The window was contamined with bird droppings producing a status warning and causing the 
backscatter plot to have a higher than normal amplitude. The problem was corrected when 
the window was cleaned during regular preventative maintenance.
Measurements:sgpvceil25kB6.b1:
  • Third cloud base height(third_cbh)
  • Lowest cloud base height detected.(first_cbh)
  • Second lowest cloud base height(second_cbh)
  • sum_backscatter
  • Backscatter(backscatter)
  • alt_highest_signal
  • vertical_visibility


Back To Table of Contents

DQRID : D071101.2
Start DateStart TimeEnd DateEnd Time
10/25/2007063011/08/20071729
Subject:
SGP/VCEIL/B5 - Degraded hardware
DataStreams:sgpvceil25kB5.b1
Description:
Following repair at Vaisala and re-installation on 10/23/2007, the ceilometer experiencing 
intermittent receiver alarms and discontinuities in the backscatter plot. Vaisala 
recommended the old firmware card be replaced but that did not resolve the problems. The 
instrument was replaced with a spare and returned to the vendor for repair.
Measurements:sgpvceil25kB5.b1:
  • detection_status
  • laser_temperature
  • Third cloud base height(third_cbh)
  • status_flag
  • background_light
  • vertical_visibility
  • Backscatter(backscatter)
  • range
  • sum_backscatter
  • Second lowest cloud base height(second_cbh)
  • receiver_sensitivity
  • status_string
  • measurement_parameters
  • alt_highest_signal
  • Lowest cloud base height detected.(first_cbh)
  • laser_pulse_energy
  • window_contamination
  • tilt_angle


Back To Table of Contents

DQRID : D080924.4
Start DateStart TimeEnd DateEnd Time
03/08/2002185903/11/20022013
09/23/2008203309/25/20081945
Subject:
SGP/VCEIL/B5 - Missing data
DataStreams:sgpvceil25kB5.b1
Description:
Data are missing and unrecoverable.
Measurements:sgpvceil25kB5.b1:
  • detection_status
  • laser_temperature
  • Third cloud base height(third_cbh)
  • base_time
  • status_flag
  • lat
  • background_light
  • vertical_visibility
  • Backscatter(backscatter)
  • range
  • sum_backscatter
  • Second lowest cloud base height(second_cbh)
  • lon
  • receiver_sensitivity
  • status_string
  • measurement_parameters
  • alt_highest_signal
  • Lowest cloud base height detected.(first_cbh)
  • laser_pulse_energy
  • window_contamination
  • time_offset
  • tilt_angle
  • alt


Back To Table of Contents

DQRID : D090108.1
Start DateStart TimeEnd DateEnd Time
12/10/2008153004/07/20091530
Subject:
SGP/VCEIL/B4 - Window contamination
DataStreams:sgpvceil25kB4.b1
Description:
During this period, the VCEIL window repeatedly became contaminated by bird droppings 
shortly after each cleaning. Detection of clouds was intermittently limited. The window 
warnings are contained in the status_flag data.

The cleaning frequency was increased and then bird deterrent spikes were installed on 
3/24/2009. 

Cloud height measurements are suspect when the status_string field contains the "Window 
contaminated" warning.
Measurements:sgpvceil25kB4.b1:
  • Backscatter(backscatter)
  • Lowest cloud base height detected.(first_cbh)
  • Third cloud base height(third_cbh)
  • Second lowest cloud base height(second_cbh)
  • vertical_visibility


Back To Table of Contents

DQRID : D090507.2
Start DateStart TimeEnd DateEnd Time
04/22/2009092505/16/20091010
Subject:
GRW/VCEIL/M1 - Receiver failure
DataStreams:grwvceil25kM1.b1
Description:
The AMF VCEIL had lower-than-normal laser transmitter power, ~84%, so the instrument was 
replaced with a spare on 4/22. The spare VCEIL had just been repaired but began to display 
receiver failure warnings.

Diagnostic information was sent to Vaisala and we discussed either replacing the receiver 
with the one from the original VCEIL (with the degraded transmitter) or replacing the 
transmitter in the original VCEIL with the one from the spare. Vaisala recommended the 
latter because the original VCEIL is a much newer unit.

On 5/16, the site operator replaced the transmitter in the original VCEIL with the one 
from the spare, realigned the optics, recalibrated the laser, and put the instrument back 
into service, successfully correcting the problem.

The data during this period were intermittently incorrect, containing warning flags and 
alarms and unreasonable cloud base heights.
Measurements:grwvceil25kM1.b1:
  • vertical_visibility
  • Lowest cloud base height detected.(first_cbh)
  • background_light
  • Backscatter(backscatter)
  • Third cloud base height(third_cbh)
  • Second lowest cloud base height(second_cbh)
  • alt_highest_signal


Back To Table of Contents

DQRID : D090527.1
Start DateStart TimeEnd DateEnd Time
05/25/2009154006/29/20090000
Subject:
TWP/VCEIL/C1 - Compensation fiber
DataStreams:twpvceil25kC1.b1
Description:
The VCEIL was reporting a "receiver optical cross-talk compensation is poor" warning. The 
window was cleaned and the instrument was power-cycled. The problem was corrected when 
the compensation fiber was replaced and ants were removed from communication connector.  
Cloud detection was impaired.
Measurements:twpvceil25kC1.b1:
  • Second lowest cloud base height(second_cbh)
  • alt_highest_signal
  • sum_backscatter
  • receiver_sensitivity
  • laser_pulse_energy
  • Lowest cloud base height detected.(first_cbh)
  • vertical_visibility
  • Third cloud base height(third_cbh)
  • Backscatter(backscatter)
  • background_light


Back To Table of Contents

DQRID : D091026.1
Start DateStart TimeEnd DateEnd Time
04/01/2001000010/20/20091227
Subject:
SGP/VCEIL/B1 - Warning flag
DataStreams:sgpvceil25kB1.b1
Description:
Beginning in 2000 (when this datastream was named sgpvceil25kB1.a1), the ceilometer was 
continuously giving warnings for low battery or charge voltage. Either the battery switch 
may not have been "ON" or the battery needed to be replaced. 

Attempts to correct the problem were unsuccessful until the time of site decommissioning. 
Maintenance efforts ceased 10/1/2009 and the battery was not checked prior to the site 
being decommissioned on 10/20/2009.
Measurements:sgpvceil25kB1.b1:
  • status_flag


Back To Table of Contents

DQRID : D091210.1
Start DateStart TimeEnd DateEnd Time
05/02/2006000012/03/20092107
Subject:
NSA/VCEIL/C1 - Degraded window
DataStreams:nsavceil25kC1.b1
Description:
The ceilometer continuously gave warnings for window contamination because the window had 
deteriorated. Cloud base height detection did not appear to be significantly affected. 
The window conditioner was replaced on 8/11/09 but this did not clear the warning. The 
instrument was replaced.
Measurements:nsavceil25kC1.b1:
  • Third cloud base height(third_cbh)
  • alt_highest_signal
  • Second lowest cloud base height(second_cbh)
  • status_flag
  • sum_backscatter
  • Lowest cloud base height detected.(first_cbh)
  • Backscatter(backscatter)
  • status_string
  • background_light
  • vertical_visibility
  • window_contamination


Back To Table of Contents

DQRID : D100215.1
Start DateStart TimeEnd DateEnd Time
04/01/2001000010/31/20092359
Subject:
SGP/VCEIL/C1/B4/B5/B6 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:sgpvceil25kB4.b1, sgpvceil25kB5.b1, sgpvceil25kB6.b1, sgpvceil25kC1.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.

Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.

This DQR also applies to a1 level data collected prior to b1 level.
Measurements:sgpvceil25kC1.b1:
  • sum_backscatter

sgpvceil25kB4.b1:
  • sum_backscatter

sgpvceil25kB6.b1:
  • sum_backscatter

sgpvceil25kB5.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100215.2
Start DateStart TimeEnd DateEnd Time
04/01/2001000010/21/20092359
Subject:
SGP/VCEIL/B1 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:sgpvceil25kB1.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
   
This DQR also applies to a1 level data collected prior to b1 level.
Measurements:sgpvceil25kB1.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100215.3
Start DateStart TimeEnd DateEnd Time
08/22/1999000010/31/20092359
Subject:
NSA/VCEIL/C1/C2 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:nsavceil25kC1.b1, nsavceil25kC2.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
   
This DQR also applies to a1 and b1 level data collected prior to the begin date.
Measurements:nsavceil25kC2.b1:
  • sum_backscatter

nsavceil25kC1.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100215.4
Start DateStart TimeEnd DateEnd Time
10/11/1996000010/31/20092359
Subject:
TWP/VCEIL/C1 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:twpvceil25kC1.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
   
This DQR also applies to a1 level data.
Measurements:twpvceil25kC1.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100215.5
Start DateStart TimeEnd DateEnd Time
10/28/1998000010/31/20092359
Subject:
TWP/VCEIL/C2 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:twpvceil25kC2.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
   
This DQR also applies to a1 level data.
Measurements:twpvceil25kC2.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100215.6
Start DateStart TimeEnd DateEnd Time
02/27/2002000010/31/20092359
Subject:
TWP/VCEIL/C3 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:twpvceil25kC3.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
   
This DQR also applies to a1 level data.
Measurements:twpvceil25kC3.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100216.3
Start DateStart TimeEnd DateEnd Time
11/23/2005000001/07/20072359
Subject:
NIM/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:nimvceil25kM1.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
Measurements:nimvceil25kM1.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100216.4
Start DateStart TimeEnd DateEnd Time
03/12/2007000001/01/20082359
Subject:
FKB/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:fkbvceil25kM1.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
Measurements:fkbvceil25kM1.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100216.5
Start DateStart TimeEnd DateEnd Time
05/08/2008000012/28/20082359
Subject:
HFE/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:hfevceil25kM1.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
Measurements:hfevceil25kM1.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100216.6
Start DateStart TimeEnd DateEnd Time
04/13/2009000010/31/20092359
Subject:
GRW/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:grwvceil25kM1.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
Measurements:grwvceil25kM1.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100225.2
Start DateStart TimeEnd DateEnd Time
10/06/2008000012/13/20082359
Subject:
HFE/VCEIL/S1 - Incorrect serial_number and sum_backscatter fields
DataStreams:hfevceil25kS1.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
Measurements:hfevceil25kS1.b1:
  • sum_backscatter


Back To Table of Contents

DQRID : D100616.2
Start DateStart TimeEnd DateEnd Time
12/03/2009210706/15/20102143
Subject:
NSA/VCEIL/C1 - Small gaps in data
DataStreams:nsavceil25kC1.b1
Description:
When the ceilometer was replaced, small gaps in the netCDF files occurred due to the 
presence of '?' characters in the backscatter and header records of the raw data that could 
not be ingested. It is unknown whether this was caused by a poor CPU board or 
incompatibilities with the computer configuration (hyper-threading). The problem was corrected when 
the instrument and computer were both replaced. 

During this period, approximately 3% of data are missing each hour due to this problem.
Measurements:nsavceil25kC1.b1:
  • Third cloud base height(third_cbh)
  • alt_highest_signal
  • laser_temperature
  • receiver_sensitivity
  • Second lowest cloud base height(second_cbh)
  • Lowest cloud base height detected.(first_cbh)
  • vertical_visibility
  • tilt_angle
  • time_offset
  • measurement_parameters
  • status_flag
  • sum_backscatter
  • range
  • base_time
  • Backscatter(backscatter)
  • status_string
  • background_light
  • window_contamination
  • detection_status
  • laser_pulse_energy


Back To Table of Contents