Data Quality Reports for Session: 144930 User: lovejoon82 Completed: 09/19/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D000203.2SGP/AOS/C1 - Humidograph water circulation system failuresgpaosC1.a0
D000203.3SGP/AOS/C1 - CN counter removedsgpaosC1.a0
D000801.3SGP/AOS/C1 - Nephelometer failuresgpaosauxC1.a0, sgpaosC1.a0
D000801.4SGP/AOS/C1 - Nephelometer failuresgpaosC1.a0, sgpaosauxC1.a0
D000801.5SGP/AOS/C1 - Nephelometer failuresgpaosC1.a0, sgpaosauxC1.a0
D980505.1SGP/AOS/C1 - Annual MaintenancesgpaosC1.a0
D980528.1SGP/AOS/C1 - Optical Particle Counter (PCASP-X) downsgpaosC1.a0, sgpaosauxC1.a0


DQRID : D000203.2
Start DateStart TimeEnd DateEnd Time
12/02/1999230001/15/20000000
Subject:
SGP/AOS/C1 - Humidograph water circulation system failure
DataStreams:sgpaosC1.a0
Description:
The water circulation system for the AOS humidograph failed.  Consequently, the humidified 
nephelometer was reporting aerosol light scattering values during this period that were 
close to what the reference nephelometer was observing. The difference between the two 
sets of measurements reflect particle losses in the humidifier and different temperature 
environments for the two instruments. Humidified nephelometer data during this period are 
technically valid, they just do not reflect relative humidity scans (the normal operating 
mode).
Measurements:sgpaosC1.a0:
  • Backscattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_10um_HRH)
  • Backscattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_1um_HRH)
  • Total scattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_10um_HRH)
  • Backscattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_10um_HRH)
  • Total scattering coefficient at 10 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_10um_HRH)
  • Backscattering coefficient at 10 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_10um_HRH)
  • Total scattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_10um_HRH)
  • Backscattering coefficient at 1 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_1um_HRH)
  • Total scattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_1um_HRH)
  • Total scattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_1um_HRH)
  • Total scattering coefficient at 1 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_1um_HRH)
  • Backscattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_1um_HRH)


Back To Table of Contents

DQRID : D000203.3
Start DateStart TimeEnd DateEnd Time
11/16/1999044801/15/20000000
07/08/2002113108/05/20021424
Subject:
SGP/AOS/C1 - CN counter removed
DataStreams:sgpaosC1.a0
Description:
The AOS condensation particle counter performance degraded to unacceptable
levels and was removed for repair.  

For the first time period: upon inspection of the internal condition of the instrument, it 
was found that oil was present in the intake line.  It is unknown how this oil could 
have arrived in this position.  The instrument was thoroughly cleaned and back-flushed, and 
returned to service during the maintenance visit.

For the second time period: The degradation of began July 8th.  Data collected during this 
time period are either incorrect or missing and should not be used.
Measurements:sgpaosC1.a0:
  • Particle concentration from Condensation Particle Counter(CPCPartConc)


Back To Table of Contents

DQRID : D000801.3
Start DateStart TimeEnd DateEnd Time
06/02/2000184206/16/20002234
Subject:
SGP/AOS/C1 - Nephelometer failure
DataStreams:sgpaosauxC1.a0, sgpaosC1.a0
Description:
An inexperienced AOS technician inadvertently closed off the flow to the dual-nephelometer 
humidograph system during a Friday Preventative Maintenance.  This resulted in liquid 
water getting into the humidified nephelometer.  Flow was restored to the system on Monday, 
but the nephelometers were producing strange data.  A number of tests were performed 
over the next few days to remotely troubleshoot the situation, to no avail.  Eventually, the 
decision was made to pull both nephelometers and return them to Boulder for repair.  
This DQR refers only to the reference nephelometer, which was out of service for 2 periods 
due to 2 separate problems.  This DQR is for the first period, from 000602 (1842 GMT) to 
000616 (2234 GMT).  Light absorption are invalid whenever the reference nephelometer is 
not operational, because those data are required for correction of the absorption data.
Measurements:sgpaosC1.a0:
  • Total scattering coefficient at 1 um from 700 nm TSI Low RH Nephelometer(RedTScatCoef_1um_LRH)
  • Backscattering coefficient at 10 um, from 450 nm TSI Low RH Nephelometer(BluBScatCoef_10um_LRH)
  • Backscattering coefficient at 1 um, from 450 nm TSI Low RH Nephelometer(BluBScatCoef_1um_LRH)
  • Backscattering coefficient at 1 um from 550 nm TSI Low RH Nephelometer(GrnBScatCoef_1um_LRH)
  • Absorption coefficient at 10 um(Bap_I_10um)
  • Total scattering coefficient at 1 um from 550 nm TSI Low RH Nephelometer(GrnTScatCoef_1um_LRH)
  • Total scattering coefficient at 10 um from 700 nm TSI Low RH Nephelometer(RedTScatCoef_10um_LRH)
  • Total scattering coefficient at 10 um from 550 nm TSI Low RH Nephelometer(GrnTScatCoef_10um_LRH)
  • Backscattering coefficient at 1 um from 700 nm TSI Low RH Nephelometer(RedBScatCoef_1um_LRH)
  • Total scattering coefficient at 10 um from 450 nm TSI Low RH Nephelometer(BluTScatCoef_10um_LRH)
  • Absorption coefficient at 1 um(Bap_I_1um)
  • Backscattering coefficient at 10 um from 550 nm TSI Low RH Nephelometer(GrnBScatCoef_10um_LRH)
  • Total scattering coefficient at 1 um from 450 nm TSI Low RH Nephelometer(BluTScatCoef_1um_LRH)
  • Backscattering coefficient at 10 um from 700 nm TSI Low RH Nephelometer(RedBScatCoef_10um_LRH)

sgpaosauxC1.a0:
  • BluBScatZero_LRH
  • GrnTScatZero_LRH
  • BapRef
  • GrnRScatZero_LRH
  • GrnTScatSpn_LRH
  • RedTScatSpn_LRH
  • RedTScatZero_LRH
  • BluRScatZero_LRH
  • RedRScatZero_LRH
  • BapSignal
  • RedBScatZero_LRH
  • RedBScatSpn_LRH
  • GrnBScatSpn_LRH
  • BluBScatSpn_LRH
  • BluTScatZero_LRH
  • GrnBScatZero_LRH
  • BluTScatSpn_LRH


Back To Table of Contents

DQRID : D000801.4
Start DateStart TimeEnd DateEnd Time
06/20/2000134007/03/20001833
Subject:
SGP/AOS/C1 - Nephelometer failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
An inexperienced AOS technician inadvertently closed off the flow to the dual-nephelometer 
humidograph system during a Friday Preventative Maintenance.  This resulted in liquid 
water getting into the humidified nephelometer.  Flow was restored to the system on Monday, 
but the nephelometers were producing strange data.  A number of tests were performed 
over the next few days to remotely troubleshoot the situation, to no avail.  Eventually, the 
decision was made to pull both nephelometers and return them to Boulder for repair.  
This DQR refers only to the reference nephelometer, which was out of service for 2 periods 
due to 2 separate problems.  This DQR is for the second period, from 000620 (1340 GMT) to 
000703 (1833 GMT).  Light absorption measurements are invalid whenever the reference 
nephelometer is not operational, because those data are required for correction of the 
absorption data.
Measurements:sgpaosC1.a0:
  • Total scattering coefficient at 1 um from 700 nm TSI Low RH Nephelometer(RedTScatCoef_1um_LRH)
  • Backscattering coefficient at 10 um, from 450 nm TSI Low RH Nephelometer(BluBScatCoef_10um_LRH)
  • Backscattering coefficient at 1 um, from 450 nm TSI Low RH Nephelometer(BluBScatCoef_1um_LRH)
  • Backscattering coefficient at 1 um from 550 nm TSI Low RH Nephelometer(GrnBScatCoef_1um_LRH)
  • Absorption coefficient at 10 um(Bap_I_10um)
  • Total scattering coefficient at 1 um from 550 nm TSI Low RH Nephelometer(GrnTScatCoef_1um_LRH)
  • Total scattering coefficient at 10 um from 700 nm TSI Low RH Nephelometer(RedTScatCoef_10um_LRH)
  • Total scattering coefficient at 10 um from 550 nm TSI Low RH Nephelometer(GrnTScatCoef_10um_LRH)
  • Backscattering coefficient at 1 um from 700 nm TSI Low RH Nephelometer(RedBScatCoef_1um_LRH)
  • Total scattering coefficient at 10 um from 450 nm TSI Low RH Nephelometer(BluTScatCoef_10um_LRH)
  • Absorption coefficient at 1 um(Bap_I_1um)
  • Backscattering coefficient at 10 um from 550 nm TSI Low RH Nephelometer(GrnBScatCoef_10um_LRH)
  • Total scattering coefficient at 1 um from 450 nm TSI Low RH Nephelometer(BluTScatCoef_1um_LRH)
  • Backscattering coefficient at 10 um from 700 nm TSI Low RH Nephelometer(RedBScatCoef_10um_LRH)

sgpaosauxC1.a0:
  • BluBScatZero_LRH
  • GrnTScatZero_LRH
  • BapRef
  • GrnRScatZero_LRH
  • GrnTScatSpn_LRH
  • RedTScatSpn_LRH
  • RedTScatZero_LRH
  • BluRScatZero_LRH
  • RedRScatZero_LRH
  • BapSignal
  • RedBScatZero_LRH
  • RedBScatSpn_LRH
  • GrnBScatSpn_LRH
  • BluBScatSpn_LRH
  • BluTScatZero_LRH
  • GrnBScatZero_LRH
  • BluTScatSpn_LRH


Back To Table of Contents

DQRID : D000801.5
Start DateStart TimeEnd DateEnd Time
06/02/2000184207/03/20001833
Subject:
SGP/AOS/C1 - Nephelometer failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
An inexperienced AOS technician inadvertently closed off the flow to the dual-nephelometer 
humidograph system during a Friday Preventative Maintenance.  This resulted in liquid 
water getting into the humidified nephelometer.  Flow was restored to the system on Monday, 
but the nephelometers were producing strange data.  A number of tests were performed 
over the next few days to remotely troubleshoot the situation, to no avail.  Eventually, the 
decision was made to pull both nephelometers and return them to Boulder for repair.  
This DQR refers only to the humidified nephelometer, which was out of service for the entire 
period from 000602 through 000703.
Measurements:sgpaosC1.a0:
  • Backscattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_10um_HRH)
  • Total scattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_1um_HRH)
  • Total scattering coefficient at 1 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_1um_HRH)
  • Absorption coefficient at 10 um(Bap_I_10um)
  • Backscattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_1um_HRH)
  • Backscattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_1um_HRH)
  • Total scattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_10um_HRH)
  • Total scattering coefficient at 10 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_10um_HRH)
  • Total scattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_1um_HRH)
  • Absorption coefficient at 1 um(Bap_I_1um)
  • Backscattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_10um_HRH)
  • Backscattering coefficient at 10 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_10um_HRH)
  • Total scattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_10um_HRH)
  • Backscattering coefficient at 1 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_1um_HRH)

sgpaosauxC1.a0:
  • RedTScatZero_HRH
  • GrnRScatZero_HRH
  • BluRScatZero_HRH
  • BluTScatZero_HRH
  • GrnBScatSpn_HRH
  • BapRef
  • RedRScatZero_HRH
  • BluBScatZero_HRH
  • GrnTScatSpn_HRH
  • RedBScatZero_HRH
  • BapSignal
  • BluBScatSpn_HRH
  • GrnTScatZero_HRH
  • BluTScatSpn_HRH
  • RedTScatSpn_HRH
  • GrnBScatZero_HRH
  • RedBScatSpn_HRH


Back To Table of Contents

DQRID : D980505.1
Start DateStart TimeEnd DateEnd Time
07/19/1999145207/30/19991829
01/10/2000140001/15/20000000
04/20/1998144904/23/19982136
08/04/1997145908/08/19971348
09/30/2002130010/04/20021900
03/17/2003154703/21/20032248
Subject:
SGP/AOS/C1  - Annual Maintenance
DataStreams:sgpaosC1.a0
Description:
The AOS Annual maintenance/calibration visit occurred during these periods.  During these 
times, instruments were cleaned, calibrated, and when necessary, repaired.  Instruments 
were online collecting data for some of the time; however, periods of contamination are 
probable because aerosol sampling lines were opened for inspection and cleaning and pumps 
were switched off for periods of time. Any data fields during this period that contain 
data should be considered incorrect.
Measurements:sgpaosC1.a0:
  • Particle concentration (0.50 um < Dp < 0.60 um) from PMS PCASP optical particle
    counter, channel 13(pa50_p60conc)
  • Total scattering coefficient at 1 um from 700 nm TSI Low RH Nephelometer(RedTScatCoef_1um_LRH)
  • HGupstream_RH
  • TSINephTSamp_HRH
  • Backscattering coefficient at 10 um, from 450 nm TSI Low RH Nephelometer(BluBScatCoef_10um_LRH)
  • Pres530nm
  • Particle concentration (2.00 um < Dp < 2.30 um) from PMS PCASP optical particle
    counter, channel 23(pa200_p230conc)
  • HGdownstream_T
  • RH530nm
  • Backscattering coefficient at 1 um from 550 nm TSI Low RH Nephelometer(GrnBScatCoef_1um_LRH)
  • Total scattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_1um_HRH)
  • Particle concentration (2.30 um < Dp < 2.60 um) from PMS PCASP optical particle
    counter, channel 24(pa230_p260conc)
  • TSINephRHSamp_LRH
  • Total scattering coefficient at 1 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_1um_HRH)
  • Absorption coefficient at 10 um(Bap_I_10um)
  • Particle concentration (3.00 um < Dp < 3.50 um) from PMS PCASP optical particle
    counter, channel 26(pa300_p350conc)
  • Particle concentration (0.70 um < Dp < 0.80 um) from PMS PCASP optical particle
    counter, channel 15(pa70_p80conc)
  • Backscattering coefficient at 1 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_1um_HRH)
  • Total scattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnTScatCoef_10um_HRH)
  • Particle concentration (1.00 um < Dp < 1.30 um) from PMS PCASP optical particle
    counter, channel 18(pa100_p130conc)
  • Particle concentration (0.20 um < Dp < 0.23 um) from PMS PCASP optical particle
    counter, channel 6(pa20_p23conc)
  • Total scattering coefficient at 1 um from 550 nm TSI Low RH Nephelometer(GrnTScatCoef_1um_LRH)
  • TSINephRHSamp_HRH
  • Total scattering coefficient at 10 um from 450 nm TSI High RH Nephelometer(BluTScatCoef_10um_HRH)
  • Particle concentration (0.45 um < Dp < 0.50 um) from PMS PCASP optical particle
    counter, channel 12(pa45_p50conc)
  • Total scattering coefficient at 10 um from 700 nm TSI Low RH Nephelometer(RedTScatCoef_10um_LRH)
  • Backscattering coefficient at 530 nm from single wave nephelometer(Bscat530nm)
  • Particle concentration (1.40 um < Dp < 1.60 um) from PMS PCASP optical particle
    counter, channel 20(pa140_p160conc)
  • Particle concentration (0.14 um < Dp < 0.16 um) from PMS PCASP optical particle
    counter, channel 3(pa14_p16conc)
  • Particle concentration from Condensation Particle Counter(CPCPartConc)
  • TSINephTin_LRH
  • Total scattering coefficient at 10 um from 450 nm TSI Low RH Nephelometer(BluTScatCoef_10um_LRH)
  • Particle concentration (1.60 um < Dp < 1.80 um) from PMS PCASP optical particle
    counter, channel 21(pa160_p180conc)
  • Particle concentration (3.50 um < Dp < 4.00 um) from PMS PCASP optical particle
    counter, channel 27(pa350_p400conc)
  • Particle concentration (0.23 um < Dp < 0.26 um) from PMS PCASP optical particle
    counter, channel 7(pa23_p26conc)
  • Absorption coefficient at 1 um(Bap_I_1um)
  • TSINephTSamp_LRH
  • Particle concentration (4.00 um < Dp < 5.00 um) from PMS PCASP optical particle
    counter, channel 28(pa400_p500conc)
  • Particle concentration (0.60 um < Dp < 0.70 um) from PMS PCASP optical particle
    counter, channel 14(pa60_p70conc)
  • Total scattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_10um_HRH)
  • Total scattering coefficient at 1 um from 450 nm TSI Low RH Nephelometer(BluTScatCoef_1um_LRH)
  • Particle concentration (0.26 um < Dp < 0.30 um) from PMS PCASP optical particle
    counter, channel 8(pa26_p30conc)
  • TSINephPres_LRH
  • Particle concentration (1.80 um < Dp < 2.00 um) from PMS PCASP optical particle
    counter, channel 22(pa180_p200conc)
  • Backscattering coefficient at 10 um from 550 nm TSI High RH Nephelometer(GrnBScatCoef_10um_HRH)
  • Backscattering coefficient at 1 um, from 450 nm TSI Low RH Nephelometer(BluBScatCoef_1um_LRH)
  • Particle concentration (6.50 um < Dp < 8.00 um) from PMS PCASP optical particle
    counter, channel 30(pa650_p800conc)
  • Particle concentration (0.30 um < Dp < 0.35 um) from PMS PCASP optical particle
    counter, channel 9(pa30_p35conc)
  • Particle concentration (0.16 um < Dp < 0.18 um) from PMS PCASP optical particle
    counter, channel 4(pa16_p18conc)
  • TSINephPres_HRH
  • Backscattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_1um_HRH)
  • SampTemp
  • HGdownstream_RH
  • SampRH
  • Ozone concentration(Ozone)
  • Neph530Temp
  • Particle concentration (Dp > 10 um) from PMS PCASP optical particle counter,
    channel 0(Pa1000Conc)
  • Total scattering coefficient at 10 um from 550 nm TSI Low RH Nephelometer(GrnTScatCoef_10um_LRH)
  • Backscattering coefficient at 1 um from 700 nm TSI Low RH Nephelometer(RedBScatCoef_1um_LRH)
  • Total scattering coefficient at 1 um from 700 nm TSI High RH Nephelometer(RedTScatCoef_1um_HRH)
  • Particle concentration (0.10 um < Dp < 0.12 um) from PMS PCASP optical particle
    counter, channel 1(pa10_p12conc)
  • Particle concentration (5.00 um < Dp < 6.50 um) from PMS PCASP optical particle
    counter, channel 29(pa500_p650conc)
  • Particle concentration (2.60 um < Dp < 3.00 um) from PMS PCASP optical particle
    counter, channel 25(pa260_p300conc)
  • AmbTemp
  • HGupstream_T
  • Particle concentration (0.40 um < Dp < 0.45 um) from PMS PCASP optical particle
    counter, channel 11(pa40_p45conc)
  • Particle concentration (0.90 um < Dp < 1.00 um) from PMS PCASP optical particle
    counter, channel 17(pa90_p100conc)
  • Particle concentration (0.12 um < Dp < 0.14 um) from PMS PCASP optical particle
    counter, channel 2(pa12_p14conc)
  • Backscattering coefficient at 10 um from 700 nm TSI High RH Nephelometer(RedBScatCoef_10um_HRH)
  • Particle concentration (8.00 um < Dp < 10.00 um) from PMS PCASP optical particle
    counter, channel 31(pa800_p1000conc)
  • Backscattering coefficient at 10 um from 550 nm TSI Low RH Nephelometer(GrnBScatCoef_10um_LRH)
  • Particle concentration (1.30 um < Dp < 1.40 um) from PMS PCASP optical particle
    counter, channel 19(pa130_p140conc)
  • Backscattering coefficient at 10 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_10um_HRH)
  • Particle concentration (0.18 um < Dp < 0.20 um) from PMS PCASP optical particle
    counter, channel 5(pa18_p20conc)
  • Backscattering coefficient at 1 um, from 450 nm TSI High RH Nephelometer(BluBScatCoef_1um_HRH)
  • Backscattering coefficient at 10 um from 700 nm TSI Low RH Nephelometer(RedBScatCoef_10um_LRH)
  • TSINephTin_HRH
  • Particle concentration (0.80 um < Dp < 0.90 um) from PMS PCASP optical particle
    counter, channel 16(pa80_p90conc)
  • Particle concentration (0.35 um < Dp < 0.40 um) from PMS PCASP optical particle
    counter, channel 10(pa35_p40conc)


Back To Table of Contents

DQRID : D980528.1
Start DateStart TimeEnd DateEnd Time
06/09/1999082407/26/19992205
09/08/2000042610/13/20001630
07/01/1997093607/03/19971556
08/01/1997132908/08/19971348
05/05/1998092606/16/19981505
05/17/2002060205/20/20051847
05/23/2002201105/31/20021258
09/09/2002000010/07/20021905
Subject:
SGP/AOS/C1 - Optical Particle Counter (PCASP-X) down
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The AOS Optical Particle Spectrometer (PCASP-X sizing probe) was down during these 
periods.  This lowers the laser reference voltage and results in inaccurate sizing.
Measurements:sgpaosC1.a0:
  • Particle concentration (0.50 um < Dp < 0.60 um) from PMS PCASP optical particle
    counter, channel 13(pa50_p60conc)
  • Particle concentration (1.80 um < Dp < 2.00 um) from PMS PCASP optical particle
    counter, channel 22(pa180_p200conc)
  • Particle concentration (6.50 um < Dp < 8.00 um) from PMS PCASP optical particle
    counter, channel 30(pa650_p800conc)
  • Particle concentration (2.00 um < Dp < 2.30 um) from PMS PCASP optical particle
    counter, channel 23(pa200_p230conc)
  • Particle concentration (0.30 um < Dp < 0.35 um) from PMS PCASP optical particle
    counter, channel 9(pa30_p35conc)
  • Particle concentration (0.16 um < Dp < 0.18 um) from PMS PCASP optical particle
    counter, channel 4(pa16_p18conc)
  • Particle concentration (2.30 um < Dp < 2.60 um) from PMS PCASP optical particle
    counter, channel 24(pa230_p260conc)
  • Particle concentration (3.00 um < Dp < 3.50 um) from PMS PCASP optical particle
    counter, channel 26(pa300_p350conc)
  • Particle concentration (0.70 um < Dp < 0.80 um) from PMS PCASP optical particle
    counter, channel 15(pa70_p80conc)
  • Particle concentration (1.00 um < Dp < 1.30 um) from PMS PCASP optical particle
    counter, channel 18(pa100_p130conc)
  • Particle concentration (0.20 um < Dp < 0.23 um) from PMS PCASP optical particle
    counter, channel 6(pa20_p23conc)
  • Particle concentration (Dp > 10 um) from PMS PCASP optical particle counter,
    channel 0(Pa1000Conc)
  • Particle concentration (0.45 um < Dp < 0.50 um) from PMS PCASP optical particle
    counter, channel 12(pa45_p50conc)
  • Particle concentration (5.00 um < Dp < 6.50 um) from PMS PCASP optical particle
    counter, channel 29(pa500_p650conc)
  • Particle concentration (0.10 um < Dp < 0.12 um) from PMS PCASP optical particle
    counter, channel 1(pa10_p12conc)
  • Particle concentration (2.60 um < Dp < 3.00 um) from PMS PCASP optical particle
    counter, channel 25(pa260_p300conc)
  • Particle concentration (1.40 um < Dp < 1.60 um) from PMS PCASP optical particle
    counter, channel 20(pa140_p160conc)
  • Particle concentration (0.14 um < Dp < 0.16 um) from PMS PCASP optical particle
    counter, channel 3(pa14_p16conc)
  • Particle concentration (0.40 um < Dp < 0.45 um) from PMS PCASP optical particle
    counter, channel 11(pa40_p45conc)
  • Particle concentration (0.90 um < Dp < 1.00 um) from PMS PCASP optical particle
    counter, channel 17(pa90_p100conc)
  • Particle concentration (1.60 um < Dp < 1.80 um) from PMS PCASP optical particle
    counter, channel 21(pa160_p180conc)
  • Particle concentration (3.50 um < Dp < 4.00 um) from PMS PCASP optical particle
    counter, channel 27(pa350_p400conc)
  • Particle concentration (0.12 um < Dp < 0.14 um) from PMS PCASP optical particle
    counter, channel 2(pa12_p14conc)
  • Particle concentration (0.23 um < Dp < 0.26 um) from PMS PCASP optical particle
    counter, channel 7(pa23_p26conc)
  • Particle concentration (8.00 um < Dp < 10.00 um) from PMS PCASP optical particle
    counter, channel 31(pa800_p1000conc)
  • Particle concentration (1.30 um < Dp < 1.40 um) from PMS PCASP optical particle
    counter, channel 19(pa130_p140conc)
  • Particle concentration (4.00 um < Dp < 5.00 um) from PMS PCASP optical particle
    counter, channel 28(pa400_p500conc)
  • Particle concentration (0.18 um < Dp < 0.20 um) from PMS PCASP optical particle
    counter, channel 5(pa18_p20conc)
  • Particle concentration (0.60 um < Dp < 0.70 um) from PMS PCASP optical particle
    counter, channel 14(pa60_p70conc)
  • Particle concentration (0.26 um < Dp < 0.30 um) from PMS PCASP optical particle
    counter, channel 8(pa26_p30conc)
  • Particle concentration (0.80 um < Dp < 0.90 um) from PMS PCASP optical particle
    counter, channel 16(pa80_p90conc)
  • Particle concentration (0.35 um < Dp < 0.40 um) from PMS PCASP optical particle
    counter, channel 10(pa35_p40conc)

sgpaosauxC1.a0:
  • OPCFlow
  • OPCSheath
  • OPCLaserVolt


Back To Table of Contents