Data Quality Reports for Session: 144917 User: perez Completed: 09/18/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D061214.3SGP/IRT/E19 - Reprocess: Calibration errorsgpirt200msE19.a1, sgpirtE19.b1
D071217.6SGP/IRT/E15 - Instrument noise problemsgpirt200msE15.a1, sgpirtE15.00, sgpirtE15.b1
D080131.1SGP/IRT/E13 - Reprocess: Sky temperature biassgpirt200msE13.a1, sgpirtE13.b1
D080424.3SGP/IRT/E16 - reference temperature incorrectsgpirt200msE16.a1, sgpirtE16.b1
D091211.1SGP/IRT/E6 - Response timesgpirt200msE6.a1, sgpirtE6.b1
D091211.2SGP/IRT/E7 - Response timesgpirt200msE7.a1, sgpirtE7.b1
D091211.3SGP/IRT/E8 - Response timesgpirt200msE8.a1, sgpirtE8.b1
D091211.4SGP/IRT/E9 - Response timesgpirt200msE9.a1, sgpirtE9.b1
D091215.1SGP/IRT/E11 - Response timesgpirt200msE11.a1, sgpirtE11.b1
D091215.2SGP/IRT/E16 - Response timesgpirt200msE16.a1, sgpirtE16.b1
D091215.3SGP/IRT/E19 - Response timesgpirt200msE19.a1, sgpirtE19.b1
D091215.4SGP/IRT/E20 - Response timesgpirt200msE20.a1, sgpirtE20.b1
D091215.5SGP/IRT/E13 - Response timesgpirt200msE13.a1, sgpirtE13.b1
D091217.3SGP/IRT/E10 - Response timesgpirt200msE10.a1, sgpirtE10.b1
D091218.1SGP/IRT/E5 - Response timesgpirt200msE5.a1, sgpirtE5.b1
D110630.1SGP/IRT/E12 - Reprocess: Calibration errorsgpirtE12.b1
D120717.4SGP/IRT/E21 - Instrument failuresgpirtE21.b1
D120810.3NSA/IRT/C1 - Serial communication problemnsairt200msC1.a1, nsairtC1.b1


DQRID : D061214.3
Start DateStart TimeEnd DateEnd Time
12/01/2005000004/27/20071618
Subject:
SGP/IRT/E19 - Reprocess: Calibration error
DataStreams:sgpirt200msE19.a1, sgpirtE19.b1
Description:
The IRT was measuring ~20K higher sky temperatures than at other nearby extended facilies. 
The calibration of the instrument was incorrect. The problem was corrected by replacing 
the IRT.

The offset in the calibration will be determined and the data will be corrected.
Measurements:sgpirtE19.b1:
  • IR Brightness Temperature(sky_ir_temp)
  • sky_ir_temp_max
  • sky_ir_temp_min

sgpirt200msE19.a1:
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D071217.6
Start DateStart TimeEnd DateEnd Time
07/12/2005153003/28/20081722
Subject:
SGP/IRT/E15 - Instrument noise problem
DataStreams:sgpirt200msE15.a1, sgpirtE15.00, sgpirtE15.b1
Description:
The data was much noisier than other nearby sites (a factor of 4-5 times more). The 
connections and communication and other settings were checked. The IRT was replaced on 
1/29/2008 at 17:37 but the response time was incorrectly set to 0.01 s. This problem was 
corrected when the setting was changed to 0.10 s.
Measurements:sgpirtE15.b1:
  • IR Brightness Temperature(sky_ir_temp)
  • sky_ir_temp_min

sgpirt200msE15.a1:
  • IR Brightness Temperature(sky_ir_temp)

sgpirtE15.00:
  • AVHRR satellite images for Darwin from BOM (RAW)(raw)


Back To Table of Contents

DQRID : D080131.1
Start DateStart TimeEnd DateEnd Time
01/11/2006000007/25/20072315
Subject:
SGP/IRT/E13 - Reprocess: Sky temperature bias
DataStreams:sgpirt200msE13.a1, sgpirtE13.b1
Description:
Measurements of sky temperature, especially those <250K,, were positively biased compared 
with AERI because there was no compensation for the effect of the emissivity and 
temperature of the gold mirror. The problem was corrected by changing the IRT emissivity from 
1.000 to 0.987, adding an RTD and transmitter electronics to measure the temperature of the 
mirror, and changing the ambient temperature measurement for internal reference to 
external mirror temperature.

Use sky temperature measurements less than 230K with caution. Biases as large as 30K have 
been observed when the actual temperature, as measured by AERI, is less than 210 K.
Measurements:sgpirtE13.b1:
  • IR Brightness Temperature(sky_ir_temp)
  • sky_ir_temp_max
  • sky_ir_temp_min

sgpirt200msE13.a1:
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D080424.3
Start DateStart TimeEnd DateEnd Time
04/11/2007160006/19/20082035
Subject:
SGP/IRT/E16 - reference temperature incorrect
DataStreams:sgpirt200msE16.a1, sgpirtE16.b1
Description:
The internal reference temperature was unrealistically high, ~360K (90C). The ambient 
temperature setting was checked. The problem was corrected by replacing the instrument. The 
sky temperature measurement appears to have been unaffected.
Measurements:sgpirtE16.b1:
  • ref_ir_temp_std
  • ref_ir_temp

sgpirt200msE16.a1:
  • ref_ir_temp


Back To Table of Contents

DQRID : D091211.1
Start DateStart TimeEnd DateEnd Time
08/30/2005000004/06/20062359
Subject:
SGP/IRT/E6 - Response time
DataStreams:sgpirt200msE6.a1, sgpirtE6.b1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This produced a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Measurements:sgpirtE6.b1:
  • ref_ir_temp_std
  • sky_ir_temp_std

sgpirt200msE6.a1:
  • ref_ir_temp
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D091211.2
Start DateStart TimeEnd DateEnd Time
11/30/2005000004/04/20062359
Subject:
SGP/IRT/E7 - Response time
DataStreams:sgpirt200msE7.a1, sgpirtE7.b1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This produced a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Measurements:sgpirtE7.b1:
  • ref_ir_temp_std
  • sky_ir_temp_std

sgpirt200msE7.a1:
  • ref_ir_temp
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D091211.3
Start DateStart TimeEnd DateEnd Time
11/30/2005000003/22/20062359
Subject:
SGP/IRT/E8 - Response time
DataStreams:sgpirt200msE8.a1, sgpirtE8.b1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This produced a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Measurements:sgpirt200msE8.a1:
  • ref_ir_temp
  • IR Brightness Temperature(sky_ir_temp)

sgpirtE8.b1:
  • sky_ir_temp_std
  • ref_ir_temp_std


Back To Table of Contents

DQRID : D091211.4
Start DateStart TimeEnd DateEnd Time
11/30/2005000004/04/20062359
Subject:
SGP/IRT/E9 - Response time
DataStreams:sgpirt200msE9.a1, sgpirtE9.b1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This produced a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Measurements:sgpirt200msE9.a1:
  • IR Brightness Temperature(sky_ir_temp)
  • ref_ir_temp

sgpirtE9.b1:
  • sky_ir_temp_std
  • ref_ir_temp_std


Back To Table of Contents

DQRID : D091215.1
Start DateStart TimeEnd DateEnd Time
12/01/2005000003/28/20062359
Subject:
SGP/IRT/E11 - Response time
DataStreams:sgpirt200msE11.a1, sgpirtE11.b1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This caused a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Measurements:sgpirtE11.b1:
  • sky_ir_temp_std
  • ref_ir_temp_std

sgpirt200msE11.a1:
  • IR Brightness Temperature(sky_ir_temp)
  • ref_ir_temp


Back To Table of Contents

DQRID : D091215.2
Start DateStart TimeEnd DateEnd Time
12/01/2005000003/29/20062359
Subject:
SGP/IRT/E16 - Response time
DataStreams:sgpirt200msE16.a1, sgpirtE16.b1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This caused a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Measurements:sgpirtE16.b1:
  • ref_ir_temp_std
  • sky_ir_temp_std

sgpirt200msE16.a1:
  • IR Brightness Temperature(sky_ir_temp)
  • ref_ir_temp


Back To Table of Contents

DQRID : D091215.3
Start DateStart TimeEnd DateEnd Time
12/01/2005000003/30/20062359
Subject:
SGP/IRT/E19 - Response time
DataStreams:sgpirt200msE19.a1, sgpirtE19.b1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This caused a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Measurements:sgpirtE19.b1:
  • sky_ir_temp_std
  • ref_ir_temp_std

sgpirt200msE19.a1:
  • ref_ir_temp
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D091215.4
Start DateStart TimeEnd DateEnd Time
12/07/2005000003/29/20062359
Subject:
SGP/IRT/E20 - Response time
DataStreams:sgpirt200msE20.a1, sgpirtE20.b1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This caused a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Measurements:sgpirtE20.b1:
  • ref_ir_temp_std
  • sky_ir_temp_std

sgpirt200msE20.a1:
  • ref_ir_temp
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D091215.5
Start DateStart TimeEnd DateEnd Time
01/11/2006000003/17/20062359
Subject:
SGP/IRT/E13 - Response time
DataStreams:sgpirt200msE13.a1, sgpirtE13.b1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This caused a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Measurements:sgpirtE13.b1:
  • ref_ir_temp_std
  • sky_ir_temp_std

sgpirt200msE13.a1:
  • ref_ir_temp
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D091217.3
Start DateStart TimeEnd DateEnd Time
11/30/2005000004/05/20061659
Subject:
SGP/IRT/E10 - Response time
DataStreams:sgpirt200msE10.a1, sgpirtE10.b1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This caused a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Measurements:sgpirt200msE10.a1:
  • IR Brightness Temperature(sky_ir_temp)
  • ref_ir_temp

sgpirtE10.b1:
  • sky_ir_temp_std
  • ref_ir_temp_std


Back To Table of Contents

DQRID : D091218.1
Start DateStart TimeEnd DateEnd Time
08/30/2005000011/02/20090800
Subject:
SGP/IRT/E5 - Response time
DataStreams:sgpirt200msE5.a1, sgpirtE5.b1
Description:
The time response of the temperature measurement was 1.0 sec instead of 0.1 sec (which it 
should have been to accommodate the 5Hz sample rate).
Measurements:sgpirt200msE5.a1:
  • ref_ir_temp
  • IR Brightness Temperature(sky_ir_temp)

sgpirtE5.b1:
  • ref_ir_temp_std
  • sky_ir_temp_std


Back To Table of Contents

DQRID : D110630.1
Start DateStart TimeEnd DateEnd Time
02/12/2009000005/11/20111520
Subject:
SGP/IRT/E12 - Reprocess: Calibration error
DataStreams:sgpirtE12.b1
Description:
The measured sky temperature was lower than other SGP facilities because the original 
calibration, as set be the vendor, was wrong. The IRT was replaced and returned to the vendor.
Measurements:sgpirtE12.b1:
  • sky_ir_temp_min
  • sky_ir_temp_max
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D120717.4
Start DateStart TimeEnd DateEnd Time
12/07/2011173305/30/20121800
Subject:
SGP/IRT/E21 - Instrument failure
DataStreams:sgpirtE21.b1
Description:
Instrument stopped recording data and was replaced on 2/23/2012 but the calibration was 
wrong, such that the sky temperatures were about 40 K less than at the other EFs. The 
problem was corrected when the IRT was replces again on 5/30/2012.

The data can be corrected if the calibration is determined.
Measurements:sgpirtE21.b1:
  • sky_ir_temp_max
  • sky_ir_temp_min
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D120810.3
Start DateStart TimeEnd DateEnd Time
06/18/2010230005/26/20121800
Subject:
NSA/IRT/C1 - Serial communication problem
DataStreams:nsairt200msC1.a1, nsairtC1.b1
Description:
Data acquisition was intermittent due to poor serial communication, causing many gaps in 
the datastream.
Measurements:nsairt200msC1.a1:
  • lon
  • time
  • alt
  • IR Brightness Temperature(sky_ir_temp)
  • ref_ir_temp
  • time_offset
  • lat
  • base_time

nsairtC1.b1:
  • sky_ir_temp_std
  • base_time
  • alt
  • lat
  • time_offset
  • time
  • logger_temp
  • sky_ir_temp_max
  • sky_ir_temp_min
  • IR Brightness Temperature(sky_ir_temp)
  • lon
  • internal_voltage


Back To Table of Contents



END OF DATA