Data Quality Reports for Session: 143451 User: lovejoon82 Completed: 07/02/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D000203.3SGP/AOS/C1 - CN counter removedsgpaosC1.a0
D980528.1SGP/AOS/C1 - Optical Particle Counter (PCASP-X) downsgpaosC1.a0, sgpaosauxC1.a0
D990809.1SGP/AOS/C1 - Ozone instrument errorsgpaosauxC1.a0, sgpaosC1.a0


DQRID : D000203.3
Start DateStart TimeEnd DateEnd Time
11/16/1999044801/15/20000000
07/08/2002113108/05/20021424
Subject:
SGP/AOS/C1 - CN counter removed
DataStreams:sgpaosC1.a0
Description:
The AOS condensation particle counter performance degraded to unacceptable
levels and was removed for repair.  

For the first time period: upon inspection of the internal condition of the instrument, it 
was found that oil was present in the intake line.  It is unknown how this oil could 
have arrived in this position.  The instrument was thoroughly cleaned and back-flushed, and 
returned to service during the maintenance visit.

For the second time period: The degradation of began July 8th.  Data collected during this 
time period are either incorrect or missing and should not be used.
Measurements:sgpaosC1.a0:
  • Particle concentration from Condensation Particle Counter(CPCPartConc)


Back To Table of Contents

DQRID : D980528.1
Start DateStart TimeEnd DateEnd Time
06/09/1999082407/26/19992205
09/08/2000042610/13/20001630
07/01/1997093607/03/19971556
08/01/1997132908/08/19971348
05/05/1998092606/16/19981505
05/17/2002060205/20/20051847
05/23/2002201105/31/20021258
09/09/2002000010/07/20021905
Subject:
SGP/AOS/C1 - Optical Particle Counter (PCASP-X) down
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The AOS Optical Particle Spectrometer (PCASP-X sizing probe) was down during these 
periods.  This lowers the laser reference voltage and results in inaccurate sizing.
Measurements:sgpaosC1.a0:
  • Particle concentration (0.50 um < Dp < 0.60 um) from PMS PCASP optical particle
    counter, channel 13(pa50_p60conc)
  • Particle concentration (1.80 um < Dp < 2.00 um) from PMS PCASP optical particle
    counter, channel 22(pa180_p200conc)
  • Particle concentration (6.50 um < Dp < 8.00 um) from PMS PCASP optical particle
    counter, channel 30(pa650_p800conc)
  • Particle concentration (2.00 um < Dp < 2.30 um) from PMS PCASP optical particle
    counter, channel 23(pa200_p230conc)
  • Particle concentration (0.30 um < Dp < 0.35 um) from PMS PCASP optical particle
    counter, channel 9(pa30_p35conc)
  • Particle concentration (0.16 um < Dp < 0.18 um) from PMS PCASP optical particle
    counter, channel 4(pa16_p18conc)
  • Particle concentration (2.30 um < Dp < 2.60 um) from PMS PCASP optical particle
    counter, channel 24(pa230_p260conc)
  • Particle concentration (3.00 um < Dp < 3.50 um) from PMS PCASP optical particle
    counter, channel 26(pa300_p350conc)
  • Particle concentration (0.70 um < Dp < 0.80 um) from PMS PCASP optical particle
    counter, channel 15(pa70_p80conc)
  • Particle concentration (1.00 um < Dp < 1.30 um) from PMS PCASP optical particle
    counter, channel 18(pa100_p130conc)
  • Particle concentration (0.20 um < Dp < 0.23 um) from PMS PCASP optical particle
    counter, channel 6(pa20_p23conc)
  • Particle concentration (Dp > 10 um) from PMS PCASP optical particle counter,
    channel 0(Pa1000Conc)
  • Particle concentration (0.45 um < Dp < 0.50 um) from PMS PCASP optical particle
    counter, channel 12(pa45_p50conc)
  • Particle concentration (5.00 um < Dp < 6.50 um) from PMS PCASP optical particle
    counter, channel 29(pa500_p650conc)
  • Particle concentration (0.10 um < Dp < 0.12 um) from PMS PCASP optical particle
    counter, channel 1(pa10_p12conc)
  • Particle concentration (2.60 um < Dp < 3.00 um) from PMS PCASP optical particle
    counter, channel 25(pa260_p300conc)
  • Particle concentration (1.40 um < Dp < 1.60 um) from PMS PCASP optical particle
    counter, channel 20(pa140_p160conc)
  • Particle concentration (0.14 um < Dp < 0.16 um) from PMS PCASP optical particle
    counter, channel 3(pa14_p16conc)
  • Particle concentration (0.40 um < Dp < 0.45 um) from PMS PCASP optical particle
    counter, channel 11(pa40_p45conc)
  • Particle concentration (0.90 um < Dp < 1.00 um) from PMS PCASP optical particle
    counter, channel 17(pa90_p100conc)
  • Particle concentration (1.60 um < Dp < 1.80 um) from PMS PCASP optical particle
    counter, channel 21(pa160_p180conc)
  • Particle concentration (3.50 um < Dp < 4.00 um) from PMS PCASP optical particle
    counter, channel 27(pa350_p400conc)
  • Particle concentration (0.12 um < Dp < 0.14 um) from PMS PCASP optical particle
    counter, channel 2(pa12_p14conc)
  • Particle concentration (0.23 um < Dp < 0.26 um) from PMS PCASP optical particle
    counter, channel 7(pa23_p26conc)
  • Particle concentration (8.00 um < Dp < 10.00 um) from PMS PCASP optical particle
    counter, channel 31(pa800_p1000conc)
  • Particle concentration (1.30 um < Dp < 1.40 um) from PMS PCASP optical particle
    counter, channel 19(pa130_p140conc)
  • Particle concentration (4.00 um < Dp < 5.00 um) from PMS PCASP optical particle
    counter, channel 28(pa400_p500conc)
  • Particle concentration (0.18 um < Dp < 0.20 um) from PMS PCASP optical particle
    counter, channel 5(pa18_p20conc)
  • Particle concentration (0.60 um < Dp < 0.70 um) from PMS PCASP optical particle
    counter, channel 14(pa60_p70conc)
  • Particle concentration (0.26 um < Dp < 0.30 um) from PMS PCASP optical particle
    counter, channel 8(pa26_p30conc)
  • Particle concentration (0.80 um < Dp < 0.90 um) from PMS PCASP optical particle
    counter, channel 16(pa80_p90conc)
  • Particle concentration (0.35 um < Dp < 0.40 um) from PMS PCASP optical particle
    counter, channel 10(pa35_p40conc)

sgpaosauxC1.a0:
  • OPCFlow
  • OPCSheath
  • OPCLaserVolt


Back To Table of Contents

DQRID : D990809.1
Start DateStart TimeEnd DateEnd Time
06/25/1999163006/28/19991343
06/18/1999162806/22/19991537
Subject:
SGP/AOS/C1 - Ozone instrument error
DataStreams:sgpaosauxC1.a0, sgpaosC1.a0
Description:
A switch on the front panel of the ozone instrument was left in the wrong position.  This 
resulted in a bad value for ozone concentration being logged for these periods.
Measurements:sgpaosC1.a0:
  • Ozone concentration(Ozone)

sgpaosauxC1.a0:
  • O3_SetPt
  • O3_Span
  • O3_Cont
  • O3_Press
  • O3_LinHeater
  • O3_Offset
  • O3_LinTemp
  • O3_Mode
  • O3_Temp
  • O3_Spn
  • O3_Samp
  • O3_Zro


Back To Table of Contents