Data Quality Reports for Session: 143252 User: renata Completed: 06/21/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D120420.8SGP/QCRAD/C1 - NIP failuresgpqcrad1longC1.c1, sgpqcrad1longC1.c2, sgpqcrad1longC1.s1, sgpqcrad1longC1.s2


DQRID : D120420.8
Start DateStart TimeEnd DateEnd Time
06/12/2011000006/24/20111740
Subject:
SGP/QCRAD/C1 - NIP failure
DataStreams:sgpqcrad1longC1.c1, sgpqcrad1longC1.c2, sgpqcrad1longC1.s1, sgpqcrad1longC1.s2
Description:
The short_direct_normal (NIP instrument) cable failed during this time period.	Data are 
not recoverable.

Users can derive an estimate of direct normal values using the following formula:
short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle)

Documented in instrument DQR D110628.2
Measurements:sgpqcrad1longC1.c1:
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)

sgpqcrad1longC1.c2:
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)

sgpqcrad1longC1.s2:
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)

sgpqcrad1longC1.s1:
  • Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)


Back To Table of Contents



END OF DATA