Data Quality Reports for Session: 143013 User: ql_data Completed: 06/12/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D070912.1SGP/IRT/E13 - Missing datasgpirt200msE13.a1, sgpirtE13.b1
D080131.1SGP/IRT/E13 - Reprocess: Sky temperature biassgpirt200msE13.a1, sgpirtE13.b1
D081216.3SGP/IRT/E13 - Reprocess: Sensor Calibrationsgpirt200msE13.a1, sgpirtE13.b1
D081217.1SGP/IRT/E13 - Reprocess: wrong unitssgpirt200msE13.a1, sgpirtE13.b1
D091215.5SGP/IRT/E13 - Response timesgpirt200msE13.a1, sgpirtE13.b1
D100907.2SGP/IRT/E21 - Missing datasgpirtE21.b1
D101215.3SGP/IRT/E13 - Instrument output invalidsgpirt200msE13.a1, sgpirtE13.b1
D110412.1SGP/IRT/E21 - TimeCheck ErrorsgpirtE21.b1
D110630.1SGP/IRT/E12 - Reprocess: Calibration errorsgpirtE12.b1


DQRID : D070912.1
Start DateStart TimeEnd DateEnd Time
09/11/2007000109/19/20071410
09/20/2007015110/10/20072118
02/13/2008204002/22/20081815
03/02/2008043003/07/20081736
03/09/2008220005/02/20082209
05/03/2008175605/27/20081148
05/27/2008173208/04/20080708
08/04/2008140010/16/20082018
10/11/2007015610/31/20071852
12/22/2007184812/26/20072006
01/07/2008000001/10/20081430
01/28/2008185701/30/20081631
02/09/2008200202/13/20081414
07/25/2007220007/27/20071457
08/29/2007132809/06/20072000
09/08/2007070009/10/20071937
Subject:
SGP/IRT/E13 - Missing data
DataStreams:sgpirt200msE13.a1, sgpirtE13.b1
Description:
Data are missing and unrecoverable.
Measurements:sgpirtE13.b1:
  • IR Brightness Temperature(sky_ir_temp)
  • alt
  • sky_ir_temp_max
  • samples_in_average
  • base_time
  • lon
  • sky_ir_temp_std
  • lat
  • sky_ir_temp_min
  • time_offset
  • time
  • ref_ir_temp
  • ref_ir_temp_std

sgpirt200msE13.a1:
  • alt
  • IR Brightness Temperature(sky_ir_temp)
  • time
  • lat
  • time_offset
  • ref_ir_temp
  • lon
  • base_time


Back To Table of Contents

DQRID : D080131.1
Start DateStart TimeEnd DateEnd Time
01/11/2006000007/25/20072315
Subject:
SGP/IRT/E13 - Reprocess: Sky temperature bias
DataStreams:sgpirt200msE13.a1, sgpirtE13.b1
Description:
Measurements of sky temperature, especially those <250K,, were positively biased compared 
with AERI because there was no compensation for the effect of the emissivity and 
temperature of the gold mirror. The problem was corrected by changing the IRT emissivity from 
1.000 to 0.987, adding an RTD and transmitter electronics to measure the temperature of the 
mirror, and changing the ambient temperature measurement for internal reference to 
external mirror temperature.

Use sky temperature measurements less than 230K with caution. Biases as large as 30K have 
been observed when the actual temperature, as measured by AERI, is less than 210 K.
Measurements:sgpirtE13.b1:
  • IR Brightness Temperature(sky_ir_temp)
  • sky_ir_temp_max
  • sky_ir_temp_min

sgpirt200msE13.a1:
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D081216.3
Start DateStart TimeEnd DateEnd Time
06/20/2008160702/12/20092140
Subject:
SGP/IRT/E13 - Reprocess: Sensor Calibration
DataStreams:sgpirt200msE13.a1, sgpirtE13.b1
Description:
The measurements are greater than AERI, especially when sky temperature is less than -40C. 
The lens and mirror was cleaned on 11/21/2008 and the mirror was replaced on 1/28/2009 
but these actions did not correct the problem. It was learned that the internal 
calibration factor was not changed by the manufacturer after the unit was repaired and recalibrated 
on 9/6/2007. The calibration factor was corrected from 1.2172 to 1.4100 on 2/12/2009.
Measurements:sgpirtE13.b1:
  • IR Brightness Temperature(sky_ir_temp)
  • sky_ir_temp_max
  • sky_ir_temp_std
  • sky_ir_temp_min

sgpirt200msE13.a1:
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D081217.1
Start DateStart TimeEnd DateEnd Time
09/12/2008211010/27/20082030
Subject:
SGP/IRT/E13 - Reprocess: wrong units
DataStreams:sgpirt200msE13.a1, sgpirtE13.b1
Description:
While diagnosing serial communication problems, the IRT configuration was reset to factory 
defaults so the temperature unit parameter was set to C instead of K.
Measurements:sgpirtE13.b1:
  • IR Brightness Temperature(sky_ir_temp)
  • ref_ir_temp
  • sky_ir_temp_max
  • sky_ir_temp_min

sgpirt200msE13.a1:
  • ref_ir_temp
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D091215.5
Start DateStart TimeEnd DateEnd Time
01/11/2006000003/17/20062359
Subject:
SGP/IRT/E13 - Response time
DataStreams:sgpirt200msE13.a1, sgpirtE13.b1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This caused a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Measurements:sgpirtE13.b1:
  • ref_ir_temp_std
  • sky_ir_temp_std

sgpirt200msE13.a1:
  • ref_ir_temp
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D100907.2
Start DateStart TimeEnd DateEnd Time
11/05/2009170011/09/20092217
12/16/2009165901/04/20101801
05/24/2010050006/07/20102306
06/09/2010170007/14/20101600
07/17/2010130007/27/20101300
05/18/2009210605/22/20091551
Subject:
SGP/IRT/E21 - Missing data
DataStreams:sgpirtE21.b1
Description:
Data are missing and unrecoverable.
Measurements:sgpirtE21.b1:
  • samples_in_average
  • ref_ir_temp
  • alt
  • time_offset
  • base_time
  • IR Brightness Temperature(sky_ir_temp)
  • lat
  • lon
  • sky_ir_temp_max
  • time
  • sky_ir_temp_min
  • ref_ir_temp_std
  • sky_ir_temp_std


Back To Table of Contents

DQRID : D101215.3
Start DateStart TimeEnd DateEnd Time
08/09/2010153410/04/20101445
Subject:
SGP/IRT/E13 - Instrument output invalid
DataStreams:sgpirt200msE13.a1, sgpirtE13.b1
Description:
The signal was missing for 25 min and then changed to a constant values of 
sky_ir_temp=262.22K and ref_ir_temp=313.10K. The problem was corrected when the intrument was 
power-cycled.
Measurements:sgpirtE13.b1:
  • IR Brightness Temperature(sky_ir_temp)
  • ref_ir_temp
  • sky_ir_temp_max
  • ref_ir_temp_std
  • sky_ir_temp_std
  • sky_ir_temp_min

sgpirt200msE13.a1:
  • ref_ir_temp
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D110412.1
Start DateStart TimeEnd DateEnd Time
06/07/2010000006/29/20102359
Subject:
SGP/IRT/E21 - TimeCheck Error
DataStreams:sgpirtE21.b1
Description:
The EF data system lost communication with the GPS likely starting on or around June 7th; 
the max time offset for E21 was -5.04 sec. Then it began increasing steadily:

Jun 8  Offset:	  -267.449039
Jun 9  Offset:	  -541.038624
Jun 10 Offset:	  -814.469141
Jun 11 Offset:	  -1076.862114
Jun 12 Offset:	  -1350.330223
Jun 13 Offset:	  -1623.639335
Jun 14 Offset:	  -1897.047754
Jun 15 Offset:	  -2170.552312
Jun 16 Offset:	  -2422.130746
Jun 17 Offset:	  -2684.607495
Jun 18 - Jun 22 (unknown)
Jun 23 Offset:	  -4204.730579

If the system clock drifts the irt_comm program will get the same incorrect time. IRT 
temperature data will be correct but the corresponding time stamp cannot be accurately known 
or simply corrected for.
Measurements:sgpirtE21.b1:
  • time
  • time_offset
  • base_time


Back To Table of Contents

DQRID : D110630.1
Start DateStart TimeEnd DateEnd Time
02/12/2009000005/11/20111520
Subject:
SGP/IRT/E12 - Reprocess: Calibration error
DataStreams:sgpirtE12.b1
Description:
The measured sky temperature was lower than other SGP facilities because the original 
calibration, as set be the vendor, was wrong. The IRT was replaced and returned to the vendor.
Measurements:sgpirtE12.b1:
  • sky_ir_temp_min
  • sky_ir_temp_max
  • IR Brightness Temperature(sky_ir_temp)


Back To Table of Contents