Data Quality Reports for Session: 142486 User: zhkjianchi Completed: 05/18/2012


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D090520.3SGP/MFR10M/C1 - Heater failuresgpmfr10mC1.b1, sgpmfr10mC1.a0
D091112.3SGP/MFR10M/C1 - Bad connectionsgpmfr10mC1.a0, sgpmfr10mC1.b1
D100722.1SGP/MFR10M/C1 - Poor connectionsgpmfr10mC1.a0, sgpmfr10mC1.b1
D120229.2SGP/MFR10M/C1 - Noise in ch415, sensor degradationsgpmfr10mC1.a0, sgpmfr10mC1.b1


DQRID : D090520.3
Start DateStart TimeEnd DateEnd Time
05/17/2009052805/20/20092300
05/24/2009030006/10/20091600
Subject:
SGP/MFR10M/C1 - Heater failure
DataStreams:sgpmfr10mC1.b1, sgpmfr10mC1.a0
Description:
The heater stopped working on 20090517.  Maintenance on 20090520 seemed to resolve the 
problem, but the heater failed again on 5/24.  Additional maintenance on 20090610 seems to 
have resolved the issue.

The instrument is designed to operate with a head temp of around 40 deg C.  As the head 
was at ambient temp during the period covered by this DQR, the mentor suggests looking at 
the 25m MFR if one is interested in these data.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D091112.3
Start DateStart TimeEnd DateEnd Time
11/05/2009180911/25/20092000
Subject:
SGP/MFR10M/C1 - Bad connection
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
The instrument was pulled for quarterly calibration on 20091105.  When the instrument was 
reinstalled, the head cable was not making a good connection to the head.  Cleaning the 
connector and reinstalling resolved the issue.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D100722.1
Start DateStart TimeEnd DateEnd Time
05/11/2010201405/13/20101505
Subject:
SGP/MFR10M/C1 - Poor connection
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
The instrument was pulled for calibration on 5/11.  Upon re-installation, a good 
connection was not made.  Problem resolved on 5/13.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)

sgpmfr10mC1.a0:
  • head_temp
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 4(up_hemisp_narrowband_filter4)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 5(up_hemisp_narrowband_filter5)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 6(up_hemisp_narrowband_filter6)
  • 25 meter Broadband Upwelling Hemispheric Irradiance(up_hemisp_broadband)
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 3(up_hemisp_narrowband_filter3)
  • 25 meter Upwelling Narrowband Hemispheric Irradiance, Filter 2(up_hemisp_narrowband_filter2)


Back To Table of Contents

DQRID : D120229.2
Start DateStart TimeEnd DateEnd Time
09/09/2010000002/06/20122359
Subject:
SGP/MFR10M/C1 - Noise in ch415, sensor degradation
DataStreams:sgpmfr10mC1.a0, sgpmfr10mC1.b1
Description:
This instrument experienced general degradation over a period of more than a year.  Please 
see D120229.3 (DQPR-2693) and D120229.4 (DQPR-2763) for more information.

The 415nm channel was noisy.  The noise wasn't extreme, so initially at least, the mentor 
would not expect much impact in using the 415 data.  However, toward the end of the 
period covered by this DQR, we start seeing a general degradation of this instrument.  On 
20110720, after a calibration, channel 615 is dead.  On 20111101 the open channel stops 
functioning.  

This instrument is calibrated every three months.  The past couple/few calibrations showed 
a steady decline in the sensitivity of 415.  The declining sensitivity should be 
generally accounted for by each new calibration, but as time moved one, one should become less 
and less confident in those (415) data.

In general, the mentor would suggest using 20110720 as a cut-off date for reliable data 
(all channels).  Caution is given for all channels as the instrument is failing, though the 
channels not specifically mentioned may be OK.
Measurements:sgpmfr10mC1.b1:
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)

sgpmfr10mC1.a0:
  • 25 meter Upwelling Hemispheric Irradiance, Filter 1(up_hemisp_narrowband_filter1)


Back To Table of Contents



END OF DATA